A fluid temperature interpolation method for wide temperature range distributed flow calorimeter

被引:0
|
作者
Zhang, Tao [1 ]
Lv, Yiming [1 ]
Xie, Tiancheng [1 ]
Hu, Dongfang [1 ]
Xu, Jinxin [1 ]
Hou, Dexin [1 ]
Ding, Jiong [1 ]
机构
[1] China Jiliang Univ, Coll Metrol Measurement & Instrument, Xueyuan St 258, Hangzhou 310018, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2024年 / 95卷 / 10期
关键词
HEAT-TRANSFER COEFFICIENT; HYDROCARBON FUEL;
D O I
10.1063/5.0231588
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The distributed flow calorimeter is employed to measure reaction enthalpy, mixing enthalpy, and other thermodynamic parameters of hydrocarbon fuels. This information serves as a foundation for selecting and developing hydrocarbon fuels for hypersonic flight. The fluid temperature in the tube is a key factor in characterizing its thermodynamic behavior. Given the challenges of monitoring fluid temperature within a tube using current flow calorimetry, a distributed method for calculating fluid temperature in tubes under high-temperature conditions is proposed. This method realizes the interpolation of the enthalpy function of the experimental fluid through several sets of experiments with varying power levels. The fluid temperature in the tube is calculated by considering the microelement as the research object. First, the methodology for calculating fluid temperature in narrow pipes across a wide temperature range is presented. Second, the simulation model of the flow calorimeter is established, and the methodology is verified through numerical simulation. Finally, a flow calorimetric experimental device is setup. N-decane was used as the fluid in the experiment, and the temperature was calculated, and the calculated results were compared with the NIST data. In the temperature range of 295.50-609.38 K, the relative error range of the calculation of n-decane temperature is -0.61% to 1.24%. The experimental results show that the method can effectively estimate the fluid temperature of the distributed flow calorimeter.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Design of a low temperature coefficient bandgap reference with a wide temperature range
    An J.
    Wu C.
    Xi'an Dianzi Keji Daxue Xuebao/Journal of Xidian University, 2021, 48 (04): : 91 - 96
  • [32] Method for Predicting the Fracture Toughness of Pipeline Steels within a Wide Temperature Range
    Baron, A. A.
    RUSSIAN METALLURGY, 2015, 2015 (03): : 216 - 221
  • [33] A method for strength analysis of brittle tubular specimens over a wide temperature range
    Polyakov, NI
    Gritsyna, VT
    Brailovsikii, VB
    INDUSTRIAL LABORATORY, 1996, 62 (10): : 636 - 637
  • [34] A new method for achieving enhanced dielectric response over a wide temperature range
    Deepam Maurya
    Fu-Chang Sun
    S. Pamir Alpay
    Shashank Priya
    Scientific Reports, 5
  • [36] Novel method to prepare O/W microemulsions within a wide temperature range
    Wenping, LA
    Tadros, TF
    ACTA PHYSICO-CHIMICA SINICA, 2000, 16 (06) : 538 - 542
  • [37] Simplified Calibration Method for Radiation Temperature Measurement System with Wide Dynamic Range
    Ma Dongxiao
    Wang Jiachun
    Chen Zongsheng
    Wang Bing
    Liu Yang
    ACTA OPTICA SINICA, 2019, 39 (06)
  • [38] PREDICTION OF FRACTURE TOUGHNESS TEMPERATURE DEPENDENCE OVER A WIDE TEMPERATURE RANGE USING SIMPLIFIED AND DIRECT SCALING METHOD
    Inoue, Takashi
    Meshii, Toshiyuki
    PROCEEDINGS OF THE ASME PRESSURE VESSELS AND PIPING CONFERENCE, 2018, VOL 3B, 2019,
  • [39] A new method for achieving enhanced dielectric response over a wide temperature range
    Maurya, Deepam
    Sun, Fu-Chang
    Alpay, S. Pamir
    Priya, Shashank
    SCIENTIFIC REPORTS, 2015, 5
  • [40] A METHOD OF MEASURING THE CONTACT PRESSURE IN COMPRESSION OF RUBBER SPECIMENS IN A WIDE TEMPERATURE RANGE
    GORELIK, BM
    BUKHINA, MF
    SAZHENOV, AF
    INDUSTRIAL LABORATORY, 1959, 25 (11): : 1435 - 1437