Platelet activation following rotary pump speed increase

被引:0
|
作者
Snyder, T.A. [1 ]
Watach, M.J. [1 ]
Litwak, K.N. [1 ]
Wagner, W.R. [1 ]
机构
[1] McGowan Center for Artificial Organ Development, University of Pittsburgh, Pittsburgh, PA, United States
关键词
Axial flow ventricular assist devices - Blood damage - Flow cytometric assay - Hemolysis - Platelet activation - Standard femoral cannulation - Thromboembolic risk;
D O I
10.1097/00002480-200003000-00096
中图分类号
学科分类号
摘要
Axial flow ventricular assist devices represent the next generation of heart assist devices. Blood damage may result from the high shear field produced by these devices. Blood damage has previously been quantified in terms of hemolysis. Shear induced platelet activation remains a concern. This phenomenon was assessed using several flow cytometric assays we have developed to the quantify in vivo circulating activated platelets (ACT-PLT), circulating platelet microaggregates (AGG-PLT), platelet consumption and platelet life span. The assays were used to measure platelet activation before, during, and after a step increase in the pump revolutions per minute (RPMs) in three calves implanted with the UOP/Nimbus axial flow ventricular assist device. A marked increase in ACT-PLTs and AGG-PLTs immediately followed the RPM increase, with recovery in these parameters over time. Platelet life span decreased from 102 ± 10 hours before the RPM change to 83 ± 14 hours (p<0.05) after the step increase. This corresponded to a 10 ± 4% increase in average daily platelet consumption. These measurements could ultimately be used to define a tolerable RPM operating range as well as the effect of the rate of RPM change on platelet activation. Control algorithms might utilize such information to minimize platelet activation and reduce thromboembolic risk.
引用
收藏
相关论文
共 50 条
  • [21] A Microfabricated Rotary Pump
    Hou-Pu Chou
    Marc A. Unger
    Stephen R. Quake
    Biomedical Microdevices, 2001, 3 : 323 - 330
  • [22] A Microfabricated Rotary Pump
    Chou, Hou-Pu
    Unger, Marc A.
    Quake, Stephen R.
    BIOMEDICAL MICRODEVICES, 2001, 3 (04) : 323 - 330
  • [23] Rotary pump construction
    Siebrecht, W
    ZEITSCHRIFT DES VEREINES DEUTSCHER INGENIEURE, 1932, 76 : 377 - 380
  • [24] Rotary heat pump
    Juodvalkis, J.
    Blazevicius, E.
    Karbauskaite, J.
    Nenorta, V.
    MECHANIKA, 2011, (05): : 535 - 539
  • [25] Markers of platelet activation following DDAVP infusion
    Maclean, RM
    Phillips, SA
    Brown, SA
    Berryman, KM
    Giddings, JC
    Collins, PW
    THROMBOSIS AND HAEMOSTASIS, 1999, : 166 - 166
  • [26] PLATELET ACTIVATION AND RESTENOSIS FOLLOWING INTRACORONARY STENTING
    MARKOVITZ, JH
    ROUBIN, GS
    PARKS, JM
    CIRCULATION, 1995, 92 (08) : 414 - 414
  • [27] PLATELET ACTIVATION FOLLOWING CEREBRAL-ANGIOGRAPHY
    GAWEL, M
    BURKETT, M
    ROSE, FC
    ACTA NEUROLOGICA SCANDINAVICA, 1980, 61 (04): : 240 - 243
  • [28] Mechanical activation of talc in high-energy speed rotary mechanoactivator
    Andric, Ljubisa
    Acimovic-Pavlovic, Zagorka
    Pavlovic, Ninoslav
    Milosevic, Vladan
    Milicevic, Sonja
    CERAMICS INTERNATIONAL, 2012, 38 (04) : 2913 - 2920
  • [29] Influence Mechanism and Effect of Rotary Shell Speed on Roto-Jet Pump Performance
    Qi H.
    Wencui L.
    Xiaobing W.
    Jiancheng S.
    International Journal of Fluid Machinery and Systems, 2024, 17 (01) : 44 - 56
  • [30] Flow Field Analysis of a Centrifugal Rotary Blood Pump under Active Speed Modulations
    Huang F.
    Wang C.
    Ruan X.
    1600, Chinese Mechanical Engineering Society (32): : 2915 - 2923