Dose mapping of an irradiation chamber of a gamma cell unit using Fricke gel dosimeters and Monte Carlo simulation

被引:0
|
作者
Edalatkhah, Elham [1 ]
Badiei, Shahryar [1 ]
机构
[1] Nucl Sci & Technol Res Inst, Radiat Applicat Res Sch, Tehran, Iran
关键词
Radiation processing; Gamma cell unit; Monte Carlo simulation; Fricke gel dosimeters; MCNP CODE; VERIFICATION; FACILITY; GEANT4; FILMS; TOOL;
D O I
10.1016/j.apradiso.2024.111631
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A gamma cell unit is used for gamma irradiation of small volume samples. Due to the sample volume, there is a dose distribution, which is imperative to be known for appropriate irradiation. In this study, samples of agarose Fricke gel dosimeters were prepared for the dose mapping of the gamma cell unit available in our laboratory. To place the dosimeters at precise locations within the irradiation chamber of the gamma cell unit, a specific phantom was designed and fabricated. In this way, the dose distribution was estimated using optical absorbance measurement of the gel dosimeters. For verification, isodose curves were also obtained using Monte Carlo simulations. The results indicate that the central east-west and north-south points in the irradiation chamber have the highest and lowest dose values, respectively. The experimental and simulation ratios of the dose uniformity was calculated to be 1.53 and 1.48, respectively. The accuracy of the dosimeter performance and the system's calibration quality are confirmed by the good agreement between the simulation and experimental results.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Gamma/neutron dose evaluation using Fricke gel and alanine gel dosimeters to be applied in boron neutron capture therapy
    Mangueira, T. F.
    Silva, C. F.
    Coelho, P. R. P.
    Campos, L. L.
    APPLIED RADIATION AND ISOTOPES, 2010, 68 (4-5) : 791 - 794
  • [2] Determination of absorbed dose in CDTN gamma irradiator using Fricke dosimeters
    Centro de Desenvolvimento da Tecnologia Nuclear - CDTN, Comissão Nacional de Energia Nuclear - CNEN, Belo Horizonte, Brazil
    Int. J. Nucl. Energy Sci. Technol., 2 (105-115):
  • [3] Preparation of polymer gel dosimeters for low gamma irradiation dose
    El-Kelany, Moushera A.
    Khozemy, Ehab E.
    Goda, Heba
    Awad, Boshra
    RADIOCHIMICA ACTA, 2023, 111 (03) : 195 - 202
  • [4] Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation
    Vedelago, J.
    Chacon Obando, D.
    Malano, F.
    Conejeros, R.
    Figueroa, R.
    Garcia, D.
    Gonzalez, G.
    Romero, M.
    Santibanez, M.
    Strumia, M. C.
    Velasquez, J.
    Mattea, F.
    Valente, M.
    RADIATION MEASUREMENTS, 2016, 91 : 54 - 64
  • [5] Dose mapping of the 60Co gamma irradiation facility and a real irradiated product - Measurements and Monte Carlo simulation
    Majer, Marija
    Pasaricek, Luka
    Knezevic, Zeljka
    RADIATION PHYSICS AND CHEMISTRY, 2024, 214
  • [6] Monte Carlo validation of dose mapping for the Tunisian Gamma Irradiation Facility using the MCNP6 code
    Dridi, W.
    Daoudi, M.
    Farah, K.
    Hosni, F.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 173
  • [7] Monte Carlo Simulation of the Irradiation of Alanine Coated Film Dosimeters with Accelerated Electrons
    Uribe, R. M.
    Salvat, F.
    Cleland, M. R.
    Berejka, A.
    APPLICATION OF ACCELERATORS IN RESEARCH AND INDUSTRY, 2009, 1099 : 66 - 70
  • [8] Commissioning dosimetry and in situ dose mapping of a semi-industrial Cobalt-60 gamma-irradiation facility using Fricke and Ceric-cerous dosimetry system and comparison with Monte Carlo simulation data
    Mortuza, Md Firoz
    Lepore, Luigi
    Khedkar, Kalpana
    Thangam, Saravanan
    Nahar, Arifatun
    Jamil, Hossen Mohammad
    Bandi, Laxminarayan
    Alam, Md Khorshed
    RADIATION PHYSICS AND CHEMISTRY, 2018, 144 : 256 - 264
  • [9] Dosimetry of gamma chamber blood irradiator using PAGAT gel dosimeter and Monte Carlo simulations
    Mohammadyari, Parvin
    Zehtabian, Mehdi
    Sina, Sedigheh
    Tavasoli, Ali Reza
    Faghihi, Reza
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (01): : 317 - 330
  • [10] Radiotherapy dosimetry parameters intercomparison among eight gel dosimeters by Monte Carlo simulation
    Mariotti, V
    Gayol, A.
    Pianoschi, T.
    Mattea, F.
    Vedelago, J.
    Perez, P.
    Valente, M.
    Alva-Sanchez, M.
    RADIATION PHYSICS AND CHEMISTRY, 2022, 190