Evaluation of 3D Printed Anthropomorphic Head Phantom for Calibration of TLD in Eye Lens Dosimeter

被引:0
|
作者
Endarko E. [1 ]
Pitaloka A.W. [1 ]
Inayah A.A. [1 ]
Ramadhan F.R. [1 ]
Kardianto K. [2 ]
Rahayuningsih B. [2 ]
机构
[1] Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS-Sukolilo, East Java, Surabaya
[2] Balai Pengamanan Fasilitas Kesehatan Gubeng, East Java, Surabaya
来源
关键词
Calibration; Dosimetry; Phantom; Radiation; Thermoluminescent Dosimetry;
D O I
10.31661/jbpe.v0i0.2109-1401
中图分类号
学科分类号
摘要
Background: Calibration of Thermo Luminescent Dosimetry (TLD) in eye lens dosimeter requires a standard phantom. The use of anthropomorphic phantoms in calibration needs evaluation. Objective: This study aimed to analyze the angular response of the TLD on the fabricated 3D anthropomorphic head phantom and Computerized Imaging Reference Systems (CIRS)-Computed Tomography (CT) dose phantom as a standard phantom irradiated with Cs-137 and to compare the absorbed dose and linear attenuation for both phantoms. Hp (3) analysis, conversion coefficient (hpK (3)), and calibration factor (CF) are also investigated. Material and Methods: In this experimental study, the fabricated 3D printed anthropomorphic head phantom was analyzed using polylactic acid (PLA) with the skull and then filled with the artificial brain and cerebrospinal fluid (CSF) as a test phantom. TLD-700H and TLD Reader Harshaw 6600 plus were used to analyze the angular response of Cs-137 radiation and to determine the absorbed dose and linear attenuation coefficient of test and standard phantoms. Results: The effect of the angle of radiation source towards TLD reading at the anthropomorphic head phantom has a similar value to the standard phantom with a calibration factor ranging from 0.82 to 1. The absorbed dose measurement and the linear attenuation coefficient of the anthropomorphic head phantom with the standard phantom have different values of 2.52 and 3.78%, respectively. Conclusion: The fabricated 3D printed anthropomorphic head phantom has good potential as an alternative to standard phantoms for TLD calibration in eye lens do-simeter. © 2023, Shiraz University of Medical Sciences. All rights reserved.
引用
收藏
页码:193 / 202
页数:9
相关论文
共 50 条
  • [21] Development of a physical 3D anthropomorphic breast phantom
    Carton, Ann-Katherine
    Bakic, Predrag
    Ullberg, Christer
    Derand, Helen
    Maidment, Andrew D. A.
    MEDICAL PHYSICS, 2011, 38 (02) : 891 - 896
  • [22] Investigation of 3D Dosimetry for An Anthropomorphic Spine Phantom
    Grant, R.
    Ibbott, G.
    Yang, J.
    Adamovics, J.
    Followill, D.
    MEDICAL PHYSICS, 2012, 39 (06) : 3724 - 3724
  • [23] Development of a 3D Anthropomorphic Phantom Generator for Microwave Imaging Applications of the Head and Neck Region
    Pelicano, Ana Catarina
    Conceicao, Raquel C.
    SENSORS, 2020, 20 (07)
  • [24] 3D Anthropomorphic Phantom Facilitates Accurate Dosimetry for Head and Neck Radiopharmaceutical Therapy Trials
    Adam, D.
    Grudzinski, J.
    Cox, B.
    Marsh, I. R.
    Bradshaw, T.
    Hill, P.
    Harari, P.
    Bednarz, B. P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (SUPPL 1) : S313 - S313
  • [25] Clinical evaluation of a transmission detector system and comparison with a homogeneous 3D phantom dosimeter
    Giglioli, Francesca Romana
    Gallio, Elena
    Franco, Pierfrancesco
    Badellino, Serena
    Ricardi, Umberto
    Fiandra, Christian
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 58 : 159 - 164
  • [26] Fast and Simple Automatic 3D Ultrasound Probe Calibration Based on 3D Printed Phantom and an Untracked Marker
    Shen, Jun
    Zemiti, Nabil
    Dillenseger, Jean-Louis
    Poignet, Philippe
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 878 - 882
  • [27] Development of an anthropomorphic phantom based on 3D printing for assessment of dose delivered to the eye and adjacent tissues
    Pereira, Dirceu D.
    Cardoso, Simone C.
    Batista, Delano V. S.
    de Souza, Felipe M. L.
    de Sousa, Juan V. M.
    Goncalves, Odair D.
    da Rosa, Luiz A. R.
    RADIATION PHYSICS AND CHEMISTRY, 2022, 199
  • [28] Validation on an anthropomorphic phantom of FORE optimization in 3D PET
    De Bernardi, Elisabetta
    Mazzoli, Marco
    Zito, Felicia
    Baselli, Giuseppe
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 247 - 250
  • [29] Using 3D Printing to Manufacture An Anthropomorphic Thorax Phantom
    Hazelaar, C.
    Dahele, M.
    van Eijnatten, M.
    Wolff, J.
    Forouzanfar, T.
    Slotman, B.
    Verbakel, W.
    MEDICAL PHYSICS, 2017, 44 (06) : 3113 - 3113
  • [30] A 3D semi-anthropomorphic photoacoustic breast phantom
    Dantuma, Maura
    van Dommelen, Rianne C.
    Manohar, Srirang
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2019, 2019, 10878