Development of a chemiluminescent immunoassay based on magnetic solid phase for quantification of homocysteine in human serum

被引:0
|
作者
Liu, Yankui [1 ,2 ]
Li, Yaru [1 ]
Wang, Rong [1 ]
Chen, Shuning [1 ]
Sun, Ning [2 ]
Qi, Xiaowei [1 ]
机构
[1] Jiangnan Univ, Affiliated Hosp, Dept Pathol, 1000 Hefeng Rd, Wuxi 214122, Jiangsu, Peoples R China
[2] Jiangnan Univ, Wuxi Sch Med, 1800 Lihu Ave, Wuxi 214028, Jiangsu, Peoples R China
关键词
HCY; Cardiovascular diseases; Chemiluminescent immunoassay; ENZYME-IMMUNOASSAY; PLASMA; HYPERHOMOCYSTEINEMIA; FOLATE; LIVER;
D O I
10.1186/s12896-024-00903-z
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
BackgroundHomocysteine (HCY) is a sulfur-containing amino acid that is an independent or important risk factor for the occurrence of many chronic diseases and is one of the most important indicators for determining health risks. However, existing HCY detection methods do not meet the requirements of clinical diagnosis. Therefore, there is an urgent need to establish new detection methods to meet the needs of clinical detection.ResultsIn this study, we used the principle of competitive method to establish a new method for the determination of HCY in human serum using a chemiluminescent enzyme immunoassay in conjunction with a chemiluminescent assay instrument that uses magnetic microparticles as the solid phase of the immunoreaction. The established method achieved satisfactory results in terms of minimum detection limit, specificity, accuracy, and clinical application. The limit of detection was 0.03 ng/mL. The intra-assay coefficient of variation (CV) was 1.94-5.05%, the inter-assay CV was 2.29-6.88%, and the recovery rate was 88.60-93.27%. Cross-reactivity with L-cysteine ranged from 0.0100 to 0.0200 mu mol/L, and cross-reactivity with glutathione ranged from 0.0100 to 0.200 mu mol/L, all of which were less than the limit of detection (LoD) of this method. The linear factor R of this method was greater than 0.99.ConclusionsIn summary, the developed method showed a good correlation with the product from Abbott. A total of 996 clinical patients with cardiovascular diseases were evaluated using the method developed in this study.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Magnetic bead-based chemiluminescent metal immunoassay with a colloidal gold label
    Fan, AP
    Lau, CW
    Lu, JZ
    ANALYTICAL CHEMISTRY, 2005, 77 (10) : 3238 - 3242
  • [42] Development of a rapid and sensitivity magnetic chemiluminescence immunoassay for DNA methyltransferase 1 in human serum
    Sitian He
    Leiliang He
    Beibei Liu
    Songcheng Yu
    Li'e Liu
    Yongmei Tian
    Jia Wang
    Lihua Ding
    Yilin Wang
    Lingbo Qu
    Fei Yu
    Yongjun Wu
    ChineseChemicalLetters, 2019, 30 (05) : 1031 - 1034
  • [43] Development of a rapid and sensitivity magnetic chemiluminescence immunoassay for DNA methyltransferase 1 in human serum
    He, Sitian
    He, Leiliang
    Liu, Beibei
    Yu, Songcheng
    Liu, Li'e
    Tian, Yongmei
    Wang, Jia
    Ding, Lihua
    Wang, Yilin
    Qu, Lingbo
    Yu, Fei
    Wu, Yongjun
    CHINESE CHEMICAL LETTERS, 2019, 30 (05) : 1031 - 1034
  • [44] DEVELOPMENT OF A SOLID-PHASE ENZYME-IMMUNOASSAY FOR THE MEASUREMENT OF SERUM CA-125
    MARKESE, JJ
    BRYG, K
    JACKSON, CM
    SHAW, NY
    TUMOR BIOLOGY, 1987, 8 (06) : 320 - 321
  • [45] DEVELOPMENT OF A SOLID-PHASE ENZYME-IMMUNOASSAY FOR THE MEASUREMENT OF SERUM CA-125
    SHAW, N
    MARKESE, J
    SWANSON, P
    PATRICK, K
    HESSE, R
    SCHUMANN, W
    TUMOUR BIOLOGY, 1986, 7 (04): : 316 - 316
  • [46] SOLID-PHASE ENZYME-IMMUNOASSAY FOR THE QUANTITATION OF PANCREATIC LIPASE IN HUMAN-SERUM
    ROOD, FW
    COLEMAN, PF
    CLINICAL CHEMISTRY, 1986, 32 (06) : 1129 - 1130
  • [47] Magnetic chemiluminescent immunoassay for human C-reactive protein on the centrifugal microfluidics platform
    Czilwik, G.
    Vashist, S. K.
    Klein, V.
    Buderer, A.
    Roth, G.
    von Stetten, F.
    Zengerle, R.
    Mark, D.
    RSC ADVANCES, 2015, 5 (76): : 61906 - 61912
  • [48] Excess EDTA interferes with cortisol measurement using a solid-phase, chemiluminescent enzyme immunoassay
    Kemppainen, Robert J.
    Behrend, Ellen N.
    Carter, Stephanie F.
    Cole, Janeva E.
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2018, 30 (03) : 438 - 441
  • [49] A Highly Sensitive Immunoassay to Enable Quantification of a Nanobody-Based Imaging Agent in Human Serum
    Guelman, Sebastian
    Huang, Catherine
    Peng, Kun
    AAPS JOURNAL, 2025, 27 (01):
  • [50] SOLID-PHASE ENZYME-IMMUNOASSAY FOR SERUM THEOPHYLLINE
    DANILOVA, NP
    BEKMAN, NI
    SHUSTOVA, LV
    VASILOV, RG
    KHIMIKO-FARMATSEVTICHESKII ZHURNAL, 1989, 23 (07): : 871 - 875