Development of a rapid and sensitivity magnetic chemiluminescence immunoassay for DNA methyltransferase 1 in human serum

被引:1
|
作者
Sitian He [1 ]
Leiliang He [1 ]
Beibei Liu [2 ]
Songcheng Yu [1 ]
Li'e Liu [1 ]
Yongmei Tian [1 ]
Jia Wang [1 ]
Lihua Ding [1 ]
Yilin Wang [1 ]
Lingbo Qu [3 ]
Fei Yu [1 ]
Yongjun Wu [1 ]
机构
[1] College of Public Health, Zhengzhou University
[2] College of Chemistry and Molecular Engineering, Zhengzhou University
基金
中国国家自然科学基金;
关键词
DNA methyltransferase 1; Biomarker; Chemiluminescence immunoassay; Magnetic particles; Human serum;
D O I
暂无
中图分类号
O657.3 [光化学分析法(光谱分析法)]; R446.6 [免疫学检验];
学科分类号
070302 ; 081704 ; 100208 ;
摘要
DNA methyltransferase 1(DNMT1) is a useful biomarker for lung cancer in early clinical diagnosis. A rapid magnetic chemiluminescence immunoassay(MCLIA) for DNMT1 in human serum has been developed.Horseradish peroxidase(HRP)-second-Ab was used to labeled polyclonal antibodies of anti-DNMT1. DNMT1 in sample integrates with specific immunomagnetic beads and can constitute a supersandwiched immunoreaction. In magnetic field, nonspecific materials can be separated. After luminescent substrate luminol-H;O;-BIP was added, the relative light unit(RLU) of HRP was detected and was discovered to be directly proportional to the content of DNMT1 in sample. The correlative variables involved in the MCLIA value were optimized and the methodological evaluation was carried out. After optimization, in the range of0.5–128 ng/mL, the linear regression equation was y = 0.5014 x + 1.769(x was logCDNMT1, y was relative luminescence units(RLU)/RLU0), and the limit of detection was 0.01 ng/mL. The RSD of intra-and interassays were 15.8%–16.9% and 14.3%–18.1%, respectively. The recovery was from 70.0% to 106.2%.Furthermore, paralleled with purchasable enzyme-linked immunosorbent assay(ELISA) kits, MCLEIA had lower detection limit, wider linear range and shorter detection time. Therefore, the MCLEIA established in this study could be used for the sensitive detection of DNMT1 in serum sample.
引用
收藏
页码:1031 / 1034
页数:4
相关论文
共 50 条
  • [1] Development of a rapid and sensitivity magnetic chemiluminescence immunoassay for DNA methyltransferase 1 in human serum
    He, Sitian
    He, Leiliang
    Liu, Beibei
    Yu, Songcheng
    Liu, Li'e
    Tian, Yongmei
    Wang, Jia
    Ding, Lihua
    Wang, Yilin
    Qu, Lingbo
    Yu, Fei
    Wu, Yongjun
    CHINESE CHEMICAL LETTERS, 2019, 30 (05) : 1031 - 1034
  • [2] Enhanced chemiluminescence enzyme-linked immunoassay for the determination of DNA methyltransferase 1 in human serum
    Dong, Jiajia
    He, Leiliang
    Yu, Fei
    Yu, Songcheng
    Liu, Lie
    Tian, Yongmei
    Wang, Yilin
    Wang, Jia
    Qu, Lingbo
    Wu, Yongjun
    Han, Runping
    LUMINESCENCE, 2019, 34 (03) : 368 - 374
  • [3] Development of magnetic particle-based chemiluminescence immunoassay for measurement of human procalcitonin in serum
    Liao, Minjing
    Zheng, Jiao
    Xu, Ye
    Qiu, Yilan
    Xia, Chuan
    Zhong, Zhihong
    Liu, Lihui
    Liu, Hongrong
    Liu, Rushi
    Liang, Songyue
    JOURNAL OF IMMUNOLOGICAL METHODS, 2021, 488
  • [4] A rapid and sensitive chemiluminescence immunoassay based on magnetic particles for squamous cell carcinoma antigen in human serum
    Zhang, Huisheng
    Qi, Suwen
    CLINICA CHIMICA ACTA, 2011, 412 (17-18) : 1572 - 1577
  • [5] Development of high-performance magnetic chemiluminescence enzyme immunoassay for α-fetoprotein (AFP) in human serum
    Wang, Xu
    Zhang, Qian-Yun
    Li, Zhen-Jia
    Ying, Xi-Tang
    Lin, Jin-Ming
    CLINICA CHIMICA ACTA, 2008, 393 (02) : 90 - 94
  • [6] Development of a rapid and high-performance chemiluminescence immunoassay based on magnetic particles for protein S100B in human serum
    Zhang, Huisheng
    Qi, Suwen
    Rao, Jie
    Li, Qiaoliang
    Yin, Li
    Lu, Yuejun
    LUMINESCENCE, 2013, 28 (06) : 927 - 932
  • [7] A simple and rapid capillary chemiluminescence immunoassay for quantitatively detecting human serum HBsAg
    Xiang, A.
    Wei, F.
    Lei, X.
    Liu, Y.
    Liu, Y.
    Guo, Y.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2013, 32 (12) : 1557 - 1564
  • [8] A simple and rapid capillary chemiluminescence immunoassay for quantitatively detecting human serum HBsAg
    A. Xiang
    F. Wei
    X. Lei
    Y. Liu
    Y. Liu
    Y. Guo
    European Journal of Clinical Microbiology & Infectious Diseases, 2013, 32 : 1557 - 1564
  • [9] A Rapid Quantitative Chemiluminescence Immunoassay for Vitamin B12 in Human Serum
    Chen, Xiang
    Ren, Feng
    Xu, Jingjing
    Yu, Zheng
    Lin, Xuemei
    Bai, Zhonghu
    Gong, Fang
    CLINICAL LABORATORY, 2020, 66 (03) : 277 - 286
  • [10] Development of a Sensitive Chemiluminescence Immunoassay for the Quantification of Folic Acid in Human Serum
    Chen, Xiang
    Zhou, Qiyang
    Zhang, Ting
    Wang, ChunXin
    Yu, Zheng
    Ahamada, Hadji
    Bai, Zhonghu
    Huang, Xuan
    JOURNAL OF ANALYTICAL METHODS IN CHEMISTRY, 2019, 2019