The rrimage pet data acquisition system: initial results

被引:21
|
作者
Sportelli G. [1 ,2 ]
Ahmad S. [3 ]
Belcari N. [1 ,2 ]
Bisogni M.G. [1 ,2 ]
Camarlinghi N. [1 ,2 ]
Di Pasquale A. [1 ,2 ]
Dussoni S. [1 ]
Fleury J. [3 ]
Morrocchi M. [1 ,2 ]
Zaccaro E. [1 ,2 ]
Del Guerra A. [1 ,2 ]
机构
[1] Dipartimento di Fisica 'E. Fermi', Università di Pisa, Pisa
[2] Istituto Nazionale di Fisica Nucleare, Sezione di Pisa, Pisa
[3] Weeroc SAS, Orsay
来源
| 1600年 / Institute of Electrical and Electronics Engineers Inc.卷 / 01期
关键词
Data acquisition systems; PET instrumentation; photodetectors;
D O I
10.1109/TNS.2016.2633237
中图分类号
学科分类号
摘要
We present the first results obtained with a prototype of the PET read-out electronics of the trimodal PET/MRI/EEG TRIMAGE scanner. The read-out is based on the 64-channel TRIROC ASIC and on an acquisition board that will control up to 12 ASICs. The output of each ASIC is processed in parallel and sent to a host system that in the final version will receive data from 18 acquisition boards. Blocks of 64 SiPMs are one-to-one coupled to a dual-layer staggered LYSO crystal matrix and read by a single ASIC. The FPGA reads the sparse output from the ASICs and reconstructs for each event a full image of the light pattern coming from the LYSO matrix. This pattern can be then processed on-line or sent to the host PC for post-processing. Early tests were conducted by using a prototype board with single LYSO crystals of 3.3 mm × 3.3 mm × 8 mm and dual layer staggered LYSO matrices. Results show that the ASIC can sustain input rates above 58 kHz on all its channels, with small variations depending on the discriminating thresholds, being this limit due its digital output stage. With the single crystals setup, we obtained an energy resolution of 10.7% at 511 keV and a coincidence time resolution of 420 ps FWHM. With the staggered matrix the obtained mean energy resolution was 16% on the top layer and 18% on the bottom layer. The flood maps obtained with the LYSO matrix setup show that the pixels on both the staggered levels are clearly identifiable. © 2017 IEEE.
引用
收藏
页码:168 / 177
页数:9
相关论文
共 50 条
  • [1] SiPM based Detector Module and Digital Data Acquisition System for PET: Initial Results
    Safavi-Naeini, M.
    Franklin, D. R.
    Petasecca, M.
    Lerch, M. L. F.
    Kirkham, R.
    Moorhead, G.
    Dunn, P.
    De Geronimo, G.
    Rosenfeld, A. B.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2679 - +
  • [2] Characterization and test of a Data Acquisition System for PET
    Morrocchi, Matteo
    Marcatili, Sara
    Belcari, Nicola
    Bisogni, Maria Giuseppina
    Collazuol, Gianmaria
    Ambrosi, Giovanni
    Corsi, Francesco
    Foresta, Maurizio
    Marzocca, Cristoforo
    Matarrese, Gianvito
    Sportelli, Giancarlo
    Guerra, Pedro
    Santos, Andres
    Del Guerra, Alberto
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 621 - 625
  • [3] Evaluation of a data acquisition system for SPECT (PET)
    Stenstrom, P
    Rillbert, A
    Habte, F
    Bohm, C
    Larsson, SA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (04) : 1655 - 1659
  • [4] Evaluation of a data acquisition system for SPECT (PET)
    Stenstrom, P
    Rillbert, A
    Habte, F
    Bohm, C
    Larsson, SA
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1332 - 1336
  • [5] Time over Threshold Data Acquisition System for PET
    Sanchez-Tembleque, V.
    Fraile, L. M.
    Udias, J. M.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [6] Simultaneous acquisition of multislice PET and MR images: Initial results with a MR-compatible PET scanner
    Catana, Ciprian
    Wu, Yibao
    Judenhofer, Martin S.
    Qi, Jinyi
    Pichler, Berrid J.
    Cherry, Simon R.
    JOURNAL OF NUCLEAR MEDICINE, 2006, 47 (12) : 1968 - 1976
  • [7] DATA ACQUISITION ON A PET
    HARVEY, ED
    HILLS, DA
    WIRELESS WORLD, 1983, 89 (1564): : 65 - &
  • [8] Fast NaF PET/CT acquisition with digital PET/CT system; single initial experience
    Esmail, Abdulredha
    Aldowaisan, Yousef
    Mohamed, Izzalarab
    Devassia, Simson
    Alfeeli, Mahmoud
    Sadeq, Alyaa
    Marafi, Fahad
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [9] Initial Modal Results and Operating Data Acquisition of Shock/Vibration Fixture
    Larsen, William
    Blough, Jason R.
    DeClerck, James P.
    VanKarsen, Charles D.
    Soine, David E.
    Jones, Richard
    TOPICS IN MODAL ANALYSIS & TESTING, VOL 9, 2019, : 363 - 370
  • [10] Simplified FPGA-based data acquisition system for PET