The rrimage pet data acquisition system: initial results

被引:21
|
作者
Sportelli G. [1 ,2 ]
Ahmad S. [3 ]
Belcari N. [1 ,2 ]
Bisogni M.G. [1 ,2 ]
Camarlinghi N. [1 ,2 ]
Di Pasquale A. [1 ,2 ]
Dussoni S. [1 ]
Fleury J. [3 ]
Morrocchi M. [1 ,2 ]
Zaccaro E. [1 ,2 ]
Del Guerra A. [1 ,2 ]
机构
[1] Dipartimento di Fisica 'E. Fermi', Università di Pisa, Pisa
[2] Istituto Nazionale di Fisica Nucleare, Sezione di Pisa, Pisa
[3] Weeroc SAS, Orsay
来源
| 1600年 / Institute of Electrical and Electronics Engineers Inc.卷 / 01期
关键词
Data acquisition systems; PET instrumentation; photodetectors;
D O I
10.1109/TNS.2016.2633237
中图分类号
学科分类号
摘要
We present the first results obtained with a prototype of the PET read-out electronics of the trimodal PET/MRI/EEG TRIMAGE scanner. The read-out is based on the 64-channel TRIROC ASIC and on an acquisition board that will control up to 12 ASICs. The output of each ASIC is processed in parallel and sent to a host system that in the final version will receive data from 18 acquisition boards. Blocks of 64 SiPMs are one-to-one coupled to a dual-layer staggered LYSO crystal matrix and read by a single ASIC. The FPGA reads the sparse output from the ASICs and reconstructs for each event a full image of the light pattern coming from the LYSO matrix. This pattern can be then processed on-line or sent to the host PC for post-processing. Early tests were conducted by using a prototype board with single LYSO crystals of 3.3 mm × 3.3 mm × 8 mm and dual layer staggered LYSO matrices. Results show that the ASIC can sustain input rates above 58 kHz on all its channels, with small variations depending on the discriminating thresholds, being this limit due its digital output stage. With the single crystals setup, we obtained an energy resolution of 10.7% at 511 keV and a coincidence time resolution of 420 ps FWHM. With the staggered matrix the obtained mean energy resolution was 16% on the top layer and 18% on the bottom layer. The flood maps obtained with the LYSO matrix setup show that the pixels on both the staggered levels are clearly identifiable. © 2017 IEEE.
引用
收藏
页码:168 / 177
页数:9
相关论文
共 50 条
  • [41] Design of a Data Acquisition System to Reduce Count Rate Losses in a PET Scanner
    Romanov, Leonid V.
    Holtermann, Paul F.
    McDaniel, David L.
    Widen, Jim I.
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [42] Firmware Upgrade for the Data Acquisition System of the LabPET Small Animal PET Scanner
    Tetrault, M. -A.
    Bergeron, M.
    Lecomte, R.
    Fontaine, R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (02) : 556 - 560
  • [43] Firmware Upgrade for the Data Acquisition System of the LabPET™ Small Animal PET Scanner
    Tetrault, M. -A.
    Bergeron, M.
    Lecomte, R.
    Fontaine, R.
    2009 16TH IEEE-NPSS REAL TIME CONFERENCE, 2009, : 244 - +
  • [44] Clinical evaluation of PET image quality as a function of acquisition time in a new TOF-PET/MR compared to TOF-PET/CT - initial results
    Konstantinos Zeimpekis
    Martin Huellner
    Felipe De Galiza Barbosa
    Edwin Ter Voert
    Helen Davison
    Gaspar Delso
    Patrick Veit-Haibach
    EJNMMI Physics, 2 (Suppl 1)
  • [45] Clinical Evaluation of PET Image Quality as a Function of Acquisition Time in a New TOF-PET/MRI Compared to TOF-PET/CT—Initial Results
    Konstantinos G. Zeimpekis
    Felipe Barbosa
    Martin Hüllner
    Edwin ter Voert
    Helen Davison
    Patrick Veit-Haibach
    Gaspar Delso
    Molecular Imaging and Biology, 2015, 17 : 735 - 744
  • [46] Experimental Results of Dynamic Load Scheduling in the CMS Data Acquisition System
    Simon, Michal
    Sakulin, Hannes
    Kozielski, Stanislaw
    COMPUTER NETWORKS, CN 2013, 2013, 370 : 11 - 20
  • [47] NOνA Data Acquisition System
    Kasahara, Susan M.
    PROCEEDINGS OF THE 2ND INTERNATIONAL CONFERENCE ON TECHNOLOGY AND INSTRUMENTATION IN PARTICLE PHYSICS (TIPP 2011), 2012, 37 : 1876 - 1883
  • [48] The NOνA Data Acquisition System
    Norman, A.
    INTERNATIONAL CONFERENCE ON COMPUTING IN HIGH ENERGY AND NUCLEAR PHYSICS 2012 (CHEP2012), PTS 1-6, 2012, 396
  • [49] DATA ACQUISITION SYSTEM
    MASTER, CA
    INSTRUMENTS & CONTROL SYSTEMS, 1969, 42 (03): : 137 - &
  • [50] ROCSTAR: Data Acquisition Electronics for TOF PET
    LeGeyt, B. C.
    Ashmanskas, W. J.
    Dormer, J.
    Krishnamoorthy, S.
    Karp, J. S.
    Surti, S.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,