Simultaneous acquisition of multislice PET and MR images: Initial results with a MR-compatible PET scanner

被引:0
|
作者
Catana, Ciprian
Wu, Yibao
Judenhofer, Martin S.
Qi, Jinyi
Pichler, Berrid J.
Cherry, Simon R.
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[2] Univ Tubingen, Clin Radiol, Lab Preclin Imaging & Imaging Technol, Tubingen, Germany
关键词
instrumentation; PET; MRI; multimodality imaging; position-sensitive avalanche photodiode; molecular imaging; animal imaging;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PET and MRI are powerful imaging techniques that are largely complementary in the information they provide. We have designed and built a MR-compatible PET scanner based on avalanche photodiode technology that allows simultaneous acquisition of PET and MR images in small animals. Methods: The PET scanner insert uses magnetic field-insensitive, position-sensitive avalanche photodiode (PSAPD) detectors coupled, via short lengths of optical fibers, to arrays of lutetium oxyorthosilicate (LSO) scintillator crystals. The optical fibers are used to minimize electromagnetic interference between the radiofrequency and gradient coils and the PET detector system. The PET detector module components and the complete PET insert assembly are described. PET data were acquired with and without MR sequences running, and detector flood histograms were compared with the ones generated from the data acquired outside the magnet. A uniform MR phantom was also imaged to assess the effect of the PET detector on the MR data acquisition. Simultaneous PET and MRI studies of a mouse were performed ex vivo. Results: PSAPDs can be successfully used to read out large numbers of scintillator crystals coupled through optical fibers with acceptable performance in terms of energy and timing resolution and crystal identification. The PSAPD-LSO detector performs well in the 7-T magnet, and no visible artifacts are detected in the MR images using standard pulse sequences. Conclusion: The first images from the complete system have been successfully acquired and reconstructed, demonstrating that simultaneous PET and MRI studies are feasible and opening up interesting possibilities for dual-modality molecular imaging studies.
引用
收藏
页码:1968 / 1976
页数:9
相关论文
共 50 条
  • [1] Simultaneous PET/MR imaging with a MR-compatible SiPM PET scanner
    Yoon, Hyun Suk
    Ko, Guen Bae
    Kwon, Sun Il
    Lee, Chan Mi
    Ito, Mikiko
    Song, In Chan
    Lee, Dong Soo
    Hong, Seong Jong
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [2] A block detector for a multislice, depth-of-interaction MR-compatible PET
    Yamamoto, S
    Takamatsu, S
    Murayama, H
    Minato, K
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (01) : 33 - 37
  • [3] An MR-compatible singles detection and processing unit for simultaneous preclinical PET/MR
    Weissler, Bjoern
    Gebhardt, Pierre
    Zinke, Manfred
    Kiessling, Fabian
    Schulz, Volkmar
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2759 - 2761
  • [4] A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner
    Slates, RB
    Farahani, K
    Shao, YP
    Marsden, PK
    Taylor, J
    Summers, PE
    Williams, S
    Beech, J
    Cherry, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (08): : 2015 - 2027
  • [5] Design and development of an MR-compatible PET scanner for imaging small animals
    Mackewn, JE
    Strul, D
    Hallett, WA
    Halsted, P
    Page, RA
    Keevil, SF
    Williams, SCR
    Cherry, SR
    Marsden, PK
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) : 1376 - 1380
  • [6] MR-compatible blood sampler for PET
    Breuer, J.
    Grazioso, R.
    Corbeil, J.
    Zhang, N.
    Schmand, M.
    Wienhard, K.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3426 - +
  • [7] Design and development of an MR-Compatible PET scanner for imaging small animals
    Mackewn, JE
    Strul, D
    Hallett, WA
    Halsted, P
    Page, RA
    Keevil, SF
    Williams, SCR
    Cherry, SR
    Marsden, PK
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3271 - 3274
  • [8] PET Performance Evaluation of an MR-Compatible PET Insert
    Wu, Yibao
    Catana, Ciprian
    Farrell, Richard
    Dokhale, Purushottam A.
    Shah, Kanai S.
    Qi, Jinyi
    Cherry, Simon R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 574 - 580
  • [9] PET Performance Evaluation of a pre-clinical SiPM based MR-compatible PET scanner
    Mackewn, Jane E.
    Lerche, Christoph W.
    Sunassee, Kavitha
    de Rosales, Rafael T. M.
    Phinikaridou, Alkystis
    Salomon, Andre
    Ayres, Richard
    Tsoumpas, Charalampos
    Soultanidis, Georgios M.
    Schaeffter, Tobias
    Marsden, Paul K.
    Schulz, Volkmar
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2776 - 2779
  • [10] Verification of the MR components attenuation maps for an MR/PET scanner with simultaneous acquisition
    Eldib, Mootaz
    Faul, David
    Pawlak, John
    Doshi, Niraj
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53