Assessment of organ and effective dose when using region-of-interest attenuators in cone-beam CT and interventional fluoroscopy

被引:8
|
作者
Xiong Z. [1 ,2 ]
Vijayan S. [1 ,2 ]
Rudin S. [1 ,2 ,3 ]
Bednarek D.R. [1 ,2 ,3 ]
机构
[1] University at Buffalo, Medical Physics Program, Buffalo, NY
[2] University at Buffalo, Toshiba Stroke and Vascular Research Center, Buffalo, NY
[3] University at Buffalo, Department of Radiology, Buffalo, NY
基金
美国国家卫生研究院;
关键词
cone-beam CT; dose reduction; EGSnrc; interventional fluoroscopy procedure; Monte Carlo; region-of-interest imaging;
D O I
10.1117/1.JMI.4.3.031210
中图分类号
学科分类号
摘要
In some medical-imaging procedures using cone-beam CT (CBCT) and fluoroscopy, only the center of the field of view (FOV) may be needed to be visualized with optimal image quality. To reduce the dose to the patient while maintaining visualization of the entire FOV, a Cu attenuator with a circular aperture for the region of interest (ROI) is used. The potential organ and effective dose reductions of ROI imaging when applied to CBCT and interventional fluoroscopic procedures were determined using EGSnrc Monte Carlo code. The Monte Carlo model was first validated by comparing the surface dose distribution in a solid-water block phantom with measurement by Gafchromic film. The dependence of dose reduction on the ROI attenuator thickness, the opening size of the ROI, the axial beam position, and the location of the different organs for both neuro and thoracic imaging was evaluated. The results showed a reduction in most organ doses of 45% to 70% and in effective dose of 46% to 66% compared to the dose in a CBCT scan and in an interventional procedure without the ROI attenuator. This work provides evidence of a substantial reduction of organ and effective doses when using an ROI attenuator during CBCT and fluoroscopic procedures. © 2017 Society of Photo-Optical Instrumentation Engineers (SPIE).
引用
收藏
相关论文
共 50 条
  • [21] Region-of-interest image reconstruction with intensity weighting in circular cone-beam CT for image-guided radiation therapy
    Cho, Seungryong
    Pearson, Erik
    Pelizzari, Charles A.
    Pan, Xiaochuan
    MEDICAL PHYSICS, 2009, 36 (04) : 1184 - 1192
  • [22] Cone-Beam CT With Enhanced Needle Guidance and Augmented Fluoroscopy Overlay: Applications in Interventional Radiology
    Key, Brandon M.
    Tutton, Sean M.
    Scheidt, Matthew J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2023, 221 (01) : 92 - 101
  • [23] Region-of-interest (ROI) cone-beam computed tomography (CBCT) using rotational digital angiography (DA) acquisitions
    More, A.
    Patel, V.
    Hoffmann, K.
    Bednarek, D.
    Rudin, S.
    MEDICAL PHYSICS, 2007, 34 (06) : 2341 - 2341
  • [24] Dose calculation using megavoltage cone-beam CT
    Morin, Olivier
    Chen, Josephine
    Aubin, Michele
    Gillis, Amy
    Aubry, Jean-Francois
    Bose, Supratik
    Chen, Hong
    Descovich, Martina
    Xia, Ping
    Pouliot, Jean
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (04): : 1201 - 1210
  • [25] Automatic Region-Of-Interest Delineation of Cone-Beam CT Images for Adapting Treatment and Assessing Response in Head-And-Neck Cancer
    Hou, J.
    Chen, W.
    Guerrero, M.
    D'Souza, W.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [26] Estimating organ dose with optimized peak dose index in cone-beam CT scans
    Haba, Tomonobu
    Nishihara, Yusei
    Saito, Yasunori
    Tomimura, Takeshi
    Ogawa, Shuta
    Tanabe, Kaho
    Asada, Yasuki
    Kobayashi, Masanao
    Koyama, Shuji
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 118
  • [27] Personalized organ dose assessment of cone-beam CT imaging for image-guided thoracic radiotherapy
    Zhang, Y.
    Wu, H.
    Chen, Z.
    Knisely, J.
    Nath, R.
    Feng, Z.
    Bao, S.
    Deng, J.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S266 - S266
  • [28] Region-of-interest (ROI) cone-beam computed tomography (CBCT) using rotational digital subtraction angiograpby (DSA) acquisition
    Patel, V.
    More, A.
    Hoffmann, K.
    Bednarek, D.
    Rudin, S.
    MEDICAL PHYSICS, 2007, 34 (06) : 2343 - 2343
  • [29] Motion compensated iterative reconstruction of a region of interest in cardiac cone-beam CT
    Isola, A. A.
    Ziegler, A.
    Schaefer, D.
    Koehler, T.
    Niessen, W. J.
    Grass, M.
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2010, 34 (02) : 149 - 159
  • [30] Region of interest reconstruction from truncated data in circular cone-beam CT
    Yu, Lifeng
    Zou, Yu
    Sidky, Emil Y.
    Pelizzari, Charles A.
    Munro, Peter
    Pan, Xiaochuan
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (07) : 869 - 881