Assessment of organ and effective dose when using region-of-interest attenuators in cone-beam CT and interventional fluoroscopy

被引:8
|
作者
Xiong Z. [1 ,2 ]
Vijayan S. [1 ,2 ]
Rudin S. [1 ,2 ,3 ]
Bednarek D.R. [1 ,2 ,3 ]
机构
[1] University at Buffalo, Medical Physics Program, Buffalo, NY
[2] University at Buffalo, Toshiba Stroke and Vascular Research Center, Buffalo, NY
[3] University at Buffalo, Department of Radiology, Buffalo, NY
基金
美国国家卫生研究院;
关键词
cone-beam CT; dose reduction; EGSnrc; interventional fluoroscopy procedure; Monte Carlo; region-of-interest imaging;
D O I
10.1117/1.JMI.4.3.031210
中图分类号
学科分类号
摘要
In some medical-imaging procedures using cone-beam CT (CBCT) and fluoroscopy, only the center of the field of view (FOV) may be needed to be visualized with optimal image quality. To reduce the dose to the patient while maintaining visualization of the entire FOV, a Cu attenuator with a circular aperture for the region of interest (ROI) is used. The potential organ and effective dose reductions of ROI imaging when applied to CBCT and interventional fluoroscopic procedures were determined using EGSnrc Monte Carlo code. The Monte Carlo model was first validated by comparing the surface dose distribution in a solid-water block phantom with measurement by Gafchromic film. The dependence of dose reduction on the ROI attenuator thickness, the opening size of the ROI, the axial beam position, and the location of the different organs for both neuro and thoracic imaging was evaluated. The results showed a reduction in most organ doses of 45% to 70% and in effective dose of 46% to 66% compared to the dose in a CBCT scan and in an interventional procedure without the ROI attenuator. This work provides evidence of a substantial reduction of organ and effective doses when using an ROI attenuator during CBCT and fluoroscopic procedures. © 2017 Society of Photo-Optical Instrumentation Engineers (SPIE).
引用
收藏
相关论文
共 50 条
  • [1] Organ and effective dose reduction for region-of-interest (ROI) CBCT and fluoroscopy
    Xiong, Zhenyu
    Vijayan, Sarath
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [2] Region-of-interest reconstruction for a cone-beam dental CT with a circular trajectory
    Hu, Zhanli
    Zou, Jing
    Gui, Jianbao
    Zheng, Hairong
    Xia, Dan
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 708 : 39 - 45
  • [3] Improving image accuracy of region-of-interest in cone-beam CT using prior image
    Lee, Jiseoc
    Kim, Jin Sung
    Cho, Seungryong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 252 - 261
  • [4] Dynamic Region-of-Interest Cone-Beam CT for Image-Guided Postmastectomy Radiotherapy
    Pearson, Erik
    Al-Hallaq, Hania
    Pan, Xiaochuan
    Pelizzari, Charles
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3187 - 3190
  • [5] Region-of-interest cone beam CT with a spiral scan
    Tam, KC
    Samarasekera, S
    Sauer, F
    PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 274 - 283
  • [6] Region-of-interest image reconstruction in circular cone-beam microCT
    Cho, Seungryong
    Bian, Junguo
    Pelizzari, Charles A.
    Chen, Chin-Tu
    He, Tong-Chuan
    Pan, Xiaochuan
    MEDICAL PHYSICS, 2007, 34 (12) : 4923 - 4933
  • [7] Preliminary study of region-of-interest image reconstruction with intensity weighting in cone-beam CT using iterative algorithm
    Son, Kihong
    Lee, Jiseoc
    Lee, Younjeong
    Kim, Jin Sung
    Cho, Seungryong
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [8] Backprojection spiral scan region-of-interest cone beam CT
    Tam, KC
    Ladendorf, B
    Sauer, F
    Lauritsch, G
    Steinmetz, A
    MEDICAL IMAGING 1999: IMAGE PROCESSING, PTS 1 AND 2, 1999, 3661 : 433 - 441
  • [9] An organ and effective dose study of XVI and OBI cone-beam CT systems
    Hyer, Daniel E.
    Serago, Christopher F.
    Kim, Siyong
    Li, Jonathan G.
    Hintenlang, David E.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (02): : 181 - 197
  • [10] Half-Fan-Based Intensity-Weighted Region-of-Interest Imaging for Low-Dose Cone-Beam CT
    Yoo, B.
    Pua, R.
    Kim, J.
    Cho, S.
    MEDICAL PHYSICS, 2011, 38 (06)