Longitudinal monitoring of reconstructed activity concentration on a clinical time-of-flight PET/CT scanner

被引:1
|
作者
MacDonald L.R. [1 ]
Perkins A.E. [2 ]
Tung C.-H. [2 ]
机构
[1] University of Washington, Radiology Department, 1715 Northeast Columbia Road, Box 357987, Seattle, 98195-7987, WA
[2] Philips Healthcare, 595 Miner Road, Highland Heights, 44143, OH
基金
美国国家卫生研究院;
关键词
PET/CT; quality control procedure; quantitative imaging biomarker; quantitative imaging biomarker alliance; SUV variance;
D O I
10.1117/1.JMI.4.1.011004
中图分类号
学科分类号
摘要
Positron emission tomography (PET) images are potential quantitative biomarkers. Understanding long-term (months/years) biomarker variability is important for establishing confidence intervals on studies using such biomarkers over these time frames. PET biomarkers are derived from activity concentration (ρ) extracted from PET images. Over 30 months, we measured the stability of decay-normalized counts (Nn) and ρ by scanning the same 4.5-cm-diameter Ge-68 cylinder weekly, the same Na-22 point source daily, and a refilled 20-cm F-18 cylinder phantom monthly on a clinical TOF-PET/CT scanner. Longitudinal and adjacent-measurement variability was characterized. We found no drift in ρ or Nn for properly calibrated images over 24 months. During this time, ρmean ranged ±5% to 6% for count-matched Ge-68 and F-18 images, with coefficient of variation (COV) across time of 2.3% (Ge-68, 81 scans) and 3.2% (F-18, 24 scans). At typical patient image count levels the Ge-68 ρmean (ρmax) COV across time was 6.9% (9.6%). Changes in ρmean between adjacent F-18 scans (Δdays=34) ranged between ±10%, with corresponding date-matched changes in Ge-68 ρmean ranging ±2%. We recommend (1) tracking trends in Nn with image ρ as a check of quantitative data corrections/calibrations and (2) tracking both mean and COV of ρ (single time point measures) to hundredths precision using standardized uptake values. © 2016 Society of Photo-Optical Instrumentation Engineers (SPIE).
引用
收藏
相关论文
共 50 条
  • [41] Noise reduction using a Bayesian penalized-likelihood reconstruction algorithm on a time-of-flight PET-CT scanner
    Paulo R. R. V. Caribé
    M. Koole
    Yves D’Asseler
    B. Van Den Broeck
    S. Vandenberghe
    EJNMMI Physics, 6
  • [42] Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
    Li Huo
    Nan Li
    Heyu Wu
    Wenjia Zhu
    Haiqun Xing
    Jiantao Ba
    Tong Wang
    Fang Li
    Hui Zhang
    EJNMMI Physics, 5
  • [43] Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
    Huo, Li
    Li, Nan
    Wu, Heyu
    Zhu, Wenjia
    Xing, Haiqun
    Ba, Jiantao
    Wang, Tong
    Li, Fang
    Zhang, Hui
    EJNMMI PHYSICS, 2018, 5
  • [44] LOW DOSE FDG BRAIN IMAGING ON A HIGH SENSITIVITY TIME-OF-FLIGHT PET SCANNER
    Bailey, D.
    Bailey, E.
    Willowson, K.
    INTERNAL MEDICINE JOURNAL, 2012, 42 : 5 - 5
  • [45] Waveform-Sampling Electronics for a Whole-Body Time-of-Flight PET Scanner
    Ashmanskas, W. J.
    LeGeyt, B. C.
    Newcomer, F. M.
    Panetta, J. V.
    Ryan, W. A.
    Van Berg, R.
    Wiener, R. I.
    Karp, J. S.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (03) : 1174 - 1181
  • [46] Evaluation of Joint Reconstruction of Activity and Attenuation in Time-of-Flight PET
    Wang, Jingnan
    Wang, Tong
    Zhu, Wenjia
    Zhang, Yue
    Li, Nan
    Zhang, Hui
    Huo, Li
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [47] The impact of time-of-flight acquisition on PET/CT standardized uptake values
    Binzel, Katherine
    Zhang, Jun
    Hall, Nathan
    Knopp, Michael
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [48] Evaluation of Fourier rebinning of time-of-flight PET data to non time-of-flight
    Li, Quanzheng
    Cho, S.
    Ahn, S.
    Ren, R.
    Leahy, R.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [49] Optimisation of penalized likelihood estimation reconstruction (Q.Clear) on a digital time-of-flight PET-CT scanner for four different PET tracers
    Lindstrom, Elin
    Lindsjo, Lars
    Ilan, Ezgi
    Sundin, Anders
    Sorensen, Jens
    Danfors, Torsten
    Antoni, Gunnar
    Lubberink, Mark
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [50] Improving time resolution in time-of-flight PET
    Conti, Maurizio
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S194 - S198