A simple method for partial volume effect correction in PET images

被引:0
|
作者
Department of Nuclear Medicine, General Hospital of PLA, Beijing 100853, China [1 ]
不详 [2 ]
机构
来源
Hedianzixue Yu Tance Jishue | 2007年 / 6卷 / 1231-1238期
关键词
D O I
暂无
中图分类号
学科分类号
摘要
The partial volume effects (PVE) make the lesion distorted in PET images, which influences clinical diagnosis. A PVE correction method was set up in this study. According to the principle causing PVE, the relationships between the characteristic profiles of the transversal slice and the image of one dimension lesion were investigated based on a large of simulate experiment results for cylindrical lesions. The PVE correction for two dimension lesion image was simplified to perform with one dimension lesion image. It was named one dimension correction. Then the correction method was evaluated with the phantom study. Results: (1) One dimension correction method: The procedure of PVE correction consisted of three steps and the corrected value of lesion' size and radioactivity concentration were formulated. (2) Results of evaluation with the phantom study: The lesions' sizes and radioactivity concentrations were corrected satisfactorily, and the differences between the corrected and the true values were within 6% and 7% for the lesions larger than the system resolution (FWHM). For the smaller lesion, because of the heavier PVE there were much difference between the uncorrected and the true values of lesion' sizes and radioactivity concentrations, it reached more than 80%; it was reduced to about half after correction. A PVE correction approach for PET images of two dimension lesions, the one dimension correction, was proposed. It was simple and fast, and the correction results were satisfactory.
引用
收藏
页码:1231 / 1238
相关论文
共 50 条
  • [41] Simple method for correction of distortion in images
    Gaxiola, Leopoldo N.
    Juarez-Salazar, Rigoberto
    Diaz-Ramirez, Victor H.
    OPTICS AND PHOTONICS FOR INFORMATION PROCESSING X, 2016, 9970
  • [42] A Correction of Partial Volume Effect for Using PET to Define Metabolically Active Tumor Volume for Pancreatic Cancer
    Dalah, E.
    Cungeng, Y.
    Moraru, I.
    Erickson, B.
    Li, X.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [43] Imaging the choroid plexus for partial volume correction of (R)-[C-11]verapamil PET images
    Wang, Shaonan
    Feldmann, Maria
    Hinz, Rainer
    Koepp, Matthias
    Asselin, Marie-Claude
    Jackson, Alan
    NEUROIMAGE, 2010, 52 : S188 - S188
  • [44] Partial volume effect correction in PET image using iterative deconvolution and shearlet transform
    Jomaa, Hajer
    Mabrouk, Rostom
    Khlifa, Nawres
    2018 4TH INTERNATIONAL CONFERENCE ON ADVANCED TECHNOLOGIES FOR SIGNAL AND IMAGE PROCESSING (ATSIP), 2018,
  • [45] Effect of partial-volume correction on binding potential in high-resolution PET
    Funck, Thomas
    Paquette, Caroline
    Evans, Alan
    Thiel, Alexander
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [46] Improved diagnostic performance of florbetaben β-amyloid PET via partial volume effect correction
    Barthel, Henryk
    Rullmann, Michael
    Dukart, Juergen
    Herbert, Stefanie
    Luthardt, Julia
    Gertz, Hermann-Josef
    Reininger, Cornelia
    Sabri, Osama
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2012, 32 : S148 - S148
  • [47] Effect of partial-volume correction on binding potential in high-resolution PET
    Funck, Thomas
    Paquette, Caroline
    Evans, Alan
    Thiel, Alexander
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [48] A Simultaneous Method for MRI-based Partial Volume Correction and Image Registration in Brain PET
    Ibaraki, M.
    Matsubara, K.
    Kinoshita, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S424 - S425
  • [49] Partial-volume correction in PET: Validation of an iterative postreconstruction method with phantom and patient data
    Teo, Boon-Keng
    Seo, Youngho
    Bacharach, Stephen L.
    Carrasquillo, Jorge A.
    Libutti, Steven K.
    Shukla, Himanshu
    Hasegawa, Bruce H.
    Hawkins, Randall A.
    Franc, Benjamin L.
    JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (05) : 802 - 810
  • [50] Correction for partial volume averaging of noisy PET data.
    Slifstein, M
    Mawlawi, O
    Laruelle, M
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 117P - 117P