Comparison of biomechanical effects of polyetheretherketone (PEEK) rods and titanium rods in lumbar long-segment instrumentation: a finite element study

被引:3
|
作者
Li, Chao [1 ]
Zhao, Yao [1 ]
Qi, Longtao [1 ]
Xu, Beiyu [1 ]
Yue, Lei [1 ]
Zhu, Ranlyu [1 ]
Li, Chunde [1 ]
机构
[1] Peking Univ First Hosp, Dept Orthoped, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
lumbar degenerative disease; posterior lumbar instrumentation; PEEK; dynamic stabilization; adjacent segment degeneration; finite element analysis; PEDICLE SCREW FIXATION; INTERBODY FUSION; SPINE; ADJACENT; STABILIZATION; STIFFNESS; STRESS;
D O I
10.3389/fbioe.2024.1416046
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Introduction Polyetheretherketone (PEEK) lumbar fusion rods have been successfully used in short-segment posterior instrumentation to prevent adjacent segment degeneration. However, limited studies have reported their application in lumbar long-segment instrumentation. This study aimed to compare the biomechanical performances of PEEK rods and titanium rods in lumbar long-segment instrumentation using finite element (FE) models, with the expectation of providing clinical guidance.Methods A lumbar FE model (A) and four lumbar fixation FE models (BI, CI, BII, CII) of the L1-S1 vertebral body were developed using CT image segmentation (A: intact model; BI: intact model with L2-S1 PEEK rod internal fixation; CI: intact model with L2-S1 titanium rod internal fixation; BII: intact model with L3-S1 PEEK rod internal fixation; CII: intact model with L3-S1 titanium rod internal fixation). A 150-N preload was applied to the top surface of L1, similar to the intact model. The stresses on the lumbar intervertebral disc, facet joint, pedicle screws, and rods were calculated to evaluate the biomechanical effect of the different fixation procedures in lumbar long-segment instrumented surgery.Results Under the four physiological motion states, the average stresses on the adjacent segment intervertebral disc and facet joint in all fixation models were greater than those in the intact model. Furthermore, the average stresses on the adjacent segment intervertebral disc and facet joint were greater in models CI and CII than in models BI and BII, respectively. The average stresses on the pedicle screws and rods were decreased in models BI and BII compared with models CI and CII under the four physiological motion states, respectively.Discussion The PEEK rod internal fixation system may have better biomechanical properties than the titanium rod internal fixation system in delaying adjacent segment degeneration, improving the lumbar function of postoperative patients, and reducing the risk of screw loosening and breakage in lumbar long-segment instrumentation.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Assessment of the suitability of biodegradable rods for use in posterior lumbar fusion: An in-vitro biomechanical evaluation and finite element analysis
    Tsuang, Fon-Yih
    Hsieh, Yueh-Ying
    Kuo, Yi-Jie
    Chen, Chia-Hsien
    Lin, Feng-Huei
    Chen, Chen-Sheng
    Chiang, Chang-Jung
    PLOS ONE, 2017, 12 (11):
  • [22] The biomechanical effects of Ti versus PEEK used in the PLIF surgery on lumbar spine: a finite element analysis
    Wang, Hongwei
    Wan, Yi
    Liu, Xinyu
    Ren, Bing
    Xia, Yan
    Liu, Zhanqiang
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (10) : 1115 - 1124
  • [23] Biomechanical Effects of Proximal Polyetheretherketone Rod Extension on the Upper Instrumented and Adjacent Levels in a Human Long-Segment Construct: A Cadaveric Model
    Pereira, Bernardo de Andrada
    Lehrman, Jennifer N.
    Sawa, Anna G. U.
    Wangsawatwong, Piyanat
    Godzik, Jakub
    Xu, David S.
    Turner, Jay D.
    Kelly, Brian P.
    Uribe, Juan S.
    NEUROSPINE, 2022, 19 (03) : 828 - 837
  • [24] Biomechanical effects of laminectomies in the human lumbar spine: a finite element study
    Spina, Nicholas T.
    Moreno, Genesis S.
    Brodke, Darrel S.
    Finley, Sean M.
    Ellis, Benjamin J.
    SPINE JOURNAL, 2021, 21 (01): : 150 - 159
  • [25] Influence of Lumbar Lordosis on Posterior Rod Strain in Long-Segment Construct During Biomechanical Loading: A Cadaveric Study
    Pereira, Bernardo de Andrada
    Sawa, Anna G. U.
    Godzik, Jakub
    Lehrman, Jennifer N.
    Uribe, Juan S.
    Turner, Jay D.
    Kelly, Brian P.
    NEUROSPINE, 2021, 18 (03) : 635 - 643
  • [26] Biomechanical Comparison Between Posterior Long-Segment Fixation, Short-Segment Fixation, and Short-Segment Fixation With Intermediate Screws for the Treatment of Thoracolumbar Burst Fracture: A Finite Element Analysis
    Limthongkul, Worawat
    Wannaratsiri, Nutthee
    Sukjamsri, Chamaiporn
    Benyajati, Chi-Na
    Limthongkul, Pimpa
    Tanasansomboon, Teerachat
    Yingsakmongkol, Wicharn
    Singhatanadgige, Weerasak
    INTERNATIONAL JOURNAL OF SPINE SURGERY, 2023, 17 (03): : 442 - 448
  • [27] Biomechanical Effects of Titanium and Carbon Fiber- Reinforced PEEK as Dental Implant Materials: A Finite Element Analysis
    Ediboglu, Engin
    Akdeniz, Sidika Sinem
    Beyler, Esra
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2025, 40 (01)
  • [28] Biomechanical Effects of Multi-segment Fixation on Lumbar Spine and Sacroiliac Joints: A Finite Element Analysis
    Zhao, Geng
    Wang, Lianlei
    Wang, Hongwei
    Li, Chao
    Yuan, Suomao
    Sun, Junyuan
    Tian, Yonghao
    Liu, Xinyu
    ORTHOPAEDIC SURGERY, 2024, 16 (10) : 2499 - 2508
  • [29] Is Preventative Long-Segment Surgery for Multi-Level Spondylolysis Necessary? A Finite Element Analysis Study
    Mo, Jianqiang
    Zhang, Wen
    Zhong, Dongyan
    Xu, Hao
    Wang, Lan
    Yu, Jia
    Luo, Zongping
    PLOS ONE, 2016, 11 (02):
  • [30] Influence of double rods and interbody cages on range of motion and rod stress after spinopelvic instrumentation: a finite element study
    Leszczynski, Aleksander
    Meyer, Frank
    Charles, Yann-Philippe
    Deck, Caroline
    Bourdet, Nicolas
    Willinger, Remy
    EUROPEAN SPINE JOURNAL, 2022, 31 (06) : 1515 - 1524