共 50 条
Endoscopic Endonasal Occipitocervical Fixation with a Customized Three-Dimensional Printed Titanium Plate-Screw Construct: A Cadaveric Feasibility Study
被引:0
|作者:
Kaye, Joel
[1
]
Na, John
[1
]
Stephan, Katarina
[1
]
Bal, Ella
[1
]
Nebor, Ivanna
[1
]
Bucherl, Sean
[2
]
Nauman, Eric
[2
]
Nasser, Rani
[1
]
Virojanapa, Justin
[1
]
Andaluz, Norberto
[1
]
Forbes, Jonathan A.
[1
]
机构:
[1] Univ Cincinnati, Coll Med, Dept Neurol Surg, Cincinnati, OH 45221 USA
[2] Univ Cincinnati, Coll Engn & Appl Sci, Dept Biomed Engn, Cincinnati, OH USA
关键词:
Craniocervical junction;
Endonasal;
Occipitocervical fusion;
Odontoidectomy;
VENTRAL CRANIOCERVICAL JUNCTION;
CRANIOVERTEBRAL JUNCTION;
BIOMECHANICAL EVALUATION;
RESECTION;
PLACEMENT;
STABILITY;
EXPOSURE;
C1-C2;
D O I:
10.1016/j.wneu.2024.07.055
中图分类号:
R74 [神经病学与精神病学];
学科分类号:
摘要:
OBJECTIVE: To evaluate the feasibility of a novel method for occipitocervical fixation (OCF) through the endonasal corridor. METHODS: Thin-cut computed tomography scans were obtained for 5 cadaveric specimens. Image segmentation was used to reconstruct 3D models of each O-C1 joint complex. Using computer-aided design software, plates were custom-designed to span each O-C1 joint, sit flush onto the bony surface, and accommodate screws. The final models were 3D-printed in titanium. For implantation, specimens were held in pin-fixation and registered to neuronavigation. A rigid 0o endoscope was used for endonasal visualization. An inverted U-shaped nasopharyngeal flap was raised to expose the occipital condyles and C1. The plates were introduced and fixed with bone screws. Computed tomography scans were obtained to assess screw accuracy and proximity to critical neurovascular structures. Screw entry points and trajectories were recorded. RESULTS: Endonasal OCF was performed on 5 cadaveric specimens. The mean starting point for occipital condyle screws was 6.17 mm lateral and 5.38 mm rostral to the medial O-C1 joint. Mean axial and sagittal trajectories were 7.98 degrees degrees and 6.71 degrees, degrees , respectively. The mean starting point for C1 screws was 16.11 mm lateral to the C1 anterior tubercle and 6.39 mm caudal to the medial O-C1 joint. Mean axial and sagittal trajectories were 10.97 degrees degrees and L 9.91 degrees , respectively. CONCLUSIONS: Endonasal OCF is technically and anatomically feasible. The application of this technique may allow for same-stage endonasal decompression and fixation, offering a minimally invasive alternative to current methods of fixation and advancing surgeons' ability to treat pathology of the craniovertebral junction. Next steps will focus on biomechanical testing.
引用
收藏
页码:E959 / E969
页数:11
相关论文