Endoscopic Endonasal Occipitocervical Fixation with a Customized Three-Dimensional Printed Titanium Plate-Screw Construct: A Cadaveric Feasibility Study

被引:0
|
作者
Kaye, Joel [1 ]
Na, John [1 ]
Stephan, Katarina [1 ]
Bal, Ella [1 ]
Nebor, Ivanna [1 ]
Bucherl, Sean [2 ]
Nauman, Eric [2 ]
Nasser, Rani [1 ]
Virojanapa, Justin [1 ]
Andaluz, Norberto [1 ]
Forbes, Jonathan A. [1 ]
机构
[1] Univ Cincinnati, Coll Med, Dept Neurol Surg, Cincinnati, OH 45221 USA
[2] Univ Cincinnati, Coll Engn & Appl Sci, Dept Biomed Engn, Cincinnati, OH USA
关键词
Craniocervical junction; Endonasal; Occipitocervical fusion; Odontoidectomy; VENTRAL CRANIOCERVICAL JUNCTION; CRANIOVERTEBRAL JUNCTION; BIOMECHANICAL EVALUATION; RESECTION; PLACEMENT; STABILITY; EXPOSURE; C1-C2;
D O I
10.1016/j.wneu.2024.07.055
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To evaluate the feasibility of a novel method for occipitocervical fixation (OCF) through the endonasal corridor. METHODS: Thin-cut computed tomography scans were obtained for 5 cadaveric specimens. Image segmentation was used to reconstruct 3D models of each O-C1 joint complex. Using computer-aided design software, plates were custom-designed to span each O-C1 joint, sit flush onto the bony surface, and accommodate screws. The final models were 3D-printed in titanium. For implantation, specimens were held in pin-fixation and registered to neuronavigation. A rigid 0o endoscope was used for endonasal visualization. An inverted U-shaped nasopharyngeal flap was raised to expose the occipital condyles and C1. The plates were introduced and fixed with bone screws. Computed tomography scans were obtained to assess screw accuracy and proximity to critical neurovascular structures. Screw entry points and trajectories were recorded. RESULTS: Endonasal OCF was performed on 5 cadaveric specimens. The mean starting point for occipital condyle screws was 6.17 mm lateral and 5.38 mm rostral to the medial O-C1 joint. Mean axial and sagittal trajectories were 7.98 degrees degrees and 6.71 degrees, degrees , respectively. The mean starting point for C1 screws was 16.11 mm lateral to the C1 anterior tubercle and 6.39 mm caudal to the medial O-C1 joint. Mean axial and sagittal trajectories were 10.97 degrees degrees and L 9.91 degrees , respectively. CONCLUSIONS: Endonasal OCF is technically and anatomically feasible. The application of this technique may allow for same-stage endonasal decompression and fixation, offering a minimally invasive alternative to current methods of fixation and advancing surgeons' ability to treat pathology of the craniovertebral junction. Next steps will focus on biomechanical testing.
引用
收藏
页码:E959 / E969
页数:11
相关论文
共 50 条
  • [1] Feasibility of Occipital Condyle Screw Placement for Occipitocervical Fixation A Cadaveric Study and Description of a Novel Technique
    Uribe, Juan S.
    Ramos, Edivin
    Vale, Fernando
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2008, 21 (08): : 540 - 546
  • [2] Three-dimensional printed cast assisted screw fixation of calcaneal fractures: a prospective study
    Song, Qizhi
    Li, Tao
    Xia, Huan
    Li, Yan
    Feng, Chengbin
    Lin, Yajun
    Wang, Huahong
    Hu, Jinbiao
    Jiang, Qilong
    BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
  • [3] Three-dimensional printed cast assisted screw fixation of calcaneal fractures: a prospective study
    Qizhi Song
    Tao Li
    Huan Xia
    Yan Li
    Chengbin Feng
    Yajun Lin
    Huahong Wang
    Jinbiao Hu
    Qilong Jiang
    BMC Musculoskeletal Disorders, 24
  • [4] Three-dimensional endoscopic endonasal study of skull base anatomy
    Abarca-Olivas, Javier
    Monjas-Canovas, Irene
    Lopez-Alvarez, Beatriz
    Lloret-Garcia, Jaime
    Sanchez-del Campo, Jose
    Ramon Gras-Albert, Juan
    Moreno-Lopez, Pedro
    NEUROCIRUGIA, 2014, 25 (01): : 1 - 7
  • [5] Analysis of Complications After Cranioplasty with a Customized Three-Dimensional Titanium Mesh Plate
    Roh, Haewon
    Kim, Junwon
    Kim, Jang Hun
    Chong, Kyuha
    Yoon, Won Ki
    Kwon, Taek-Hyun
    Kim, Jong Hyun
    WORLD NEUROSURGERY, 2019, 123 : E39 - E44
  • [6] A three-dimensional printed patient-specific scaphoid replacement: a cadaveric study
    Honigmann, Philipp
    Schumacher, Ralf
    Marek, Romy
    Buttner, Franz
    Thieringer, Florian
    Haefeli, Mathias
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2018, 43 (04) : 407 - 412
  • [7] Three-dimensional finite element analysis of occipitocervical fixation using an anterior occiput-to-axis locking plate system: a pilot study
    Cai, Xianhua
    Yu, Yang
    Liu, Zhichao
    Zhang, Meichao
    Huang, Weibing
    SPINE JOURNAL, 2014, 14 (08): : 1399 - 1409
  • [8] Three-Dimensional Printed Model-Assisted Screw Installation in Treating Posterior Atlantoaxial Internal Fixation
    Minyi Yang
    Nannan Zhang
    Haodong Shi
    Hui Li
    Shichang Liu
    Zongrang Song
    Lequn Shan
    Qining Wu
    Dingjun Hao
    Scientific Reports, 8
  • [9] Three-Dimensional Printed Model-Assisted Screw Installation in Treating Posterior Atlantoaxial Internal Fixation
    Yang, Minyi
    Zhang, Nannan
    Shi, Haodong
    Li, Hui
    Liu, Shichang
    Song, Zongrang
    Shan, Lequn
    Wu, Qining
    Hao, Dingjun
    SCIENTIFIC REPORTS, 2018, 8
  • [10] Three-Dimensional Radiologic Study on Index Measurement of Endonasal Endoscopic Optic Nerve Decompression
    Zhao, Yao
    Duan, Haobo
    Liu, Jianming
    Cheng, Kailiang
    Han, Yingying
    Li, Youqiong
    JOURNAL OF CRANIOFACIAL SURGERY, 2017, 28 (06) : 1598 - 1602