Detection of Micromolar Glucose Levels in Human Sweat Using an Organic Transistor-Based Enzymatic Sensor

被引:0
|
作者
Sasaki, Yui [1 ,2 ,3 ]
Ohshiro, Kohei [1 ]
Kato, Miyuki [1 ]
Haba, Daijiro [4 ,5 ]
Nakagami, Gojiro [4 ]
Minami, Tsuyoshi [1 ]
机构
[1] Univ Tokyo, Inst Ind Sci, 4-6-1 Komaba,Meguro Ku, Tokyo 1538505, Japan
[2] Univ Tokyo, Res Ctr Adv Sci & Technol, 4-6-1 Komaba,Meguro Ku, Tokyo 1538904, Japan
[3] PRESTO, JST, 4-1-8,Honcho, Kawaguchi, Saitama 3320012, Japan
[4] Univ Tokyo, Grad Sch Med, Dept Wound Care Management Gerontol Nursing, 7-3-1 Hongo,Bunkyo Ku, Tokyo 1130033, Japan
[5] Ishikawa Prefectural Nursing Univ, Grad Sch Nursing, Dept Wellbeing Nursing, 1-1 Gakuendai, Kahoku, Ishikawa 9291210, Japan
来源
CHEMELECTROCHEM | 2024年 / 11卷 / 19期
基金
日本学术振兴会;
关键词
Sweat analysis; Organic transistor; Glucose; Biosensors; Enzymes; FIELD-EFFECT TRANSISTOR; SELF-ASSEMBLED MONOLAYER; DIRECT ELECTRON-TRANSFER; CYTOCHROME-C; INHIBITION; KINETICS; OXIDASE;
D O I
10.1002/celc.202400292
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
Sweat glucose serves as a significant biomarker of health, necessitating accurate determination at the micromolar level for noninvasive monitoring. To address this need, we design an organic field-effect transistor (OFET)-based enzymatic sensor to quantify glucose levels in human sweat. The extended-gate structure of the OFET device ensures stable analyte detection in human sweat owing to its isolated configuration. The extended-gate-type OFET has been functionalized with glucose oxidase and an N-ethylphenazonium-based mediator-attached monolayer. This configuration facilitates electron relay, enabling accurate and reproducible glucose detection. Leveraging the amplification ability of the OFET, the enzymatic sensor exhibited highly sensitive glucose detection, achieving a low limit of detection (2.9 mu M) suitable for sweat analysis requirements. In addition, the sensor exhibited high discriminability in detecting glucose amidst interferents commonly found in sweat, indicating its practical feasibility for sweat analysis. Validation of glucose recovery rates (95-105 %) in human sweat, without pretreatment, was performed using established instrumental analysis methods. An enzymatic sensor based on an extended-gate-type organic transistor allows for highly sensitive and selective detection of sweat glucose at the micromolar level. The accuracy, validated by HPLC analysis, indicates the potential of the organic transistor-based sensor for reliable sweat analysis. image
引用
收藏
页数:5
相关论文
共 50 条
  • [21] Glucose Detection in Human Sweat Using an Electronic Nose
    Olarte, Oscar
    Chilo, Jose
    Pelegri-Sebastia, Jose
    Barbe, Kurt
    Van Moer, Wendy
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 1462 - 1465
  • [22] A Paper-based Enzymatic Sensor Array for Visual Detection of Glucose Levels in Urine
    Mohammadifar, Maedeh
    Choi, Seokheun
    2018 IEEE 12TH INTERNATIONAL CONFERENCE ON NANO/MOLECULAR MEDICINE AND ENGINEERING (IEEE - NANOMED), 2018, : 244 - 247
  • [23] An enzymatic glucose detection sensor using ZnO nanostructure
    Mohammed Marie
    Sanghamitra Mandal
    Omar Manasreh
    MRS Advances, 2016, 1 (13) : 847 - 853
  • [24] An enzymatic glucose detection sensor using ZnO nanostructure
    Marie, Mohammed
    Mandal, Sanghamitra
    Manasreh, Omar
    MRS ADVANCES, 2016, 1 (13): : 847 - 853
  • [25] Polar Organic Gate Dielectrics for Graphene Field-Effect Transistor-Based Sensor Technology
    Kam, Kevin A.
    Tengan, Brianne I. C.
    Hayashi, Cody K.
    Ordonez, Richard C.
    Garmire, David G.
    SENSORS, 2018, 18 (09)
  • [26] A non-enzymatic flexible and wearable sensor based on thermal transfer printing technology for continuous glucose detection in sweat
    Liang, Xinshuo
    Zhang, Shuai
    Meng, Shuo
    Tan, Renjie
    Zhang, Ke
    Hu, Jinlian
    MICROCHEMICAL JOURNAL, 2024, 207
  • [27] Pt/MXene-Based Flexible Wearable Non-Enzymatic Electrochemical Sensor for Continuous Glucose Detection in Sweat
    Li, Quan-Fu
    Chen, Xu
    Wang, Hai
    Liu, Ming
    Peng, Hui-Ling
    ACS APPLIED MATERIALS & INTERFACES, 2023, 15 (10) : 13290 - 13298
  • [28] Ab Initio Study of Carbon Nanotube Transistor-Based Gas Sensor for NO2 Detection
    Sivasathya, S.
    Thiruvadigal, D. John
    ASIAN JOURNAL OF CHEMISTRY, 2013, 25 : S411 - S413
  • [29] Development of ZnFe(PBA)@Mo3C2Tx Nanohybrid based Non-Enzymatic Sensor for the Detection of Glucose in Human Sweat
    Ravitchandiran, Arrthi
    Dhandapani, Preethi
    Rajendra, Saradh Prasad
    Alsalhi, Mohamad S.
    Rajamohan, Rajaram
    Angaiah, Subramania
    CHEMISTRYSELECT, 2024, 9 (31):
  • [30] Influence of Bioreceptor Layer Structure on Myelin Basic Protein Detection using Organic Field Effect Transistor-Based Biosensors
    Song, Jian
    Dailey, Jennifer
    Li, Hui
    Jang, Hyun-June
    Russell, Luisa
    Zhang, Pengfei
    Searson, Peter C.
    Wang, Jeff Tza-Huei
    Everett, Allen D.
    Katz, Howard E.
    ADVANCED FUNCTIONAL MATERIALS, 2018, 28 (37)