Single antiplatelet therapy following Amplatzer left atrial appendage occlusion

被引:2
|
作者
Kramer, Anders [1 ,2 ]
Korsholm, Kasper [1 ,2 ]
Nielsen-Kudsk, Jens Erik [1 ,2 ]
机构
[1] Aarhus Univ Hosp, Dept Cardiol, Palle Juul Jensens Blvd 99, DK-8200 Aarhus, Denmark
[2] Aarhus Univ, Dept Clin Med, Aarhus, Denmark
关键词
anticoagulant therapy; ischaemic stroke; LAA closure; CARDIAC COMPUTED-TOMOGRAPHY; WATCHMAN DEVICE; CLINICAL IMPACT; CLOSURE; CONTRAINDICATION; ANTICOAGULATION;
D O I
10.4244/EIJ-D-23-00684
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: The optimal antithrombotic therapy following left atrial appendage occlusion (LAAO) remains debated. Ideally, this therapy should effectively prevent device-related thrombosis (DRT) while minimising the associated bleeding risk. AIMS: We aimed to evaluate the long-term safety and efficacy of a postprocedural single antiplatelet therapy (SAPT) strategy following Amplatzer LAAO in a large consecutive cohort. METHODS: This retrospective, single-centre, observational study included all patients discharged on SAPT after LAAO with the Amplatzer Cardiac Plug (ACP) or Amplatzer Amulet between March 2010 and December 2021 at Aarhus University Hospital, Denmark. Baseline, procedural, and imaging data were obtained locally, while clinical outcomes and medication data were extracted from the Danish national health registries. RESULTS: A total of 553 patients underwent Amplatzer LAAO during the specified time frame. Of these, 431 (77.9%) high bleeding risk patients were discharged on SAPT with either acetylsalicylic acid (n=403, 72.9%) or clopidogrel (n=28, 5.1%). At 6 months, 173 (41.7%) patients were not on any antithrombotic therapy. The mean CHA 2 DS 2-VASc and HAS-BLED scores were 3.9 +/- 1.5 and 3.4 +/- 1.1, respectively. DRT was detected in 6 (1.5%) patients on 8-week follow-up imaging using cardiac computed tomography (n=386, 89.6%) or transoesophageal echocardiography (n=27, 6.3%). The 1-year ischaemic stroke rate was 2.2% (95% confidence interval [CI]: 1.14.2). One-year rates for major bleeding and cardiovascular death were 5.9% (95% CI: 4.0-8.9) and 2.9% (95% CI: 1.6-5.1), respectively. CONCLUSIONS: SAPT following Amplatzer LAAO displayed rates of DRT and stroke comparable to those reported with more intensive antithrombotic regimens. Meanwhile, we observed low rates of major bleeding.
引用
收藏
页码:e301 / e311
页数:24
相关论文
共 50 条
  • [21] Early safety of the Amplatzer Cardiac Plug™ for left atrial appendage occlusion
    Meerkin, David
    Butnaru, Adi
    Dratva, Dmitry
    Bertrand, Olivier F.
    Tzivoni, Dan
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (04) : 3920 - 3925
  • [22] Early Safety of the Amplatzer Cardiac Plug™ for Left Atrial Appendage Occlusion
    Meerkin, David
    Butnaru, Adi
    Dratva, Dmitry
    Bertrand, Olivier
    Tzivoni, Dan
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (17) : B223 - B223
  • [23] Left atrial appendage occlusion with the Amplatzer Amulet: update on device sizing
    Freixa, Xavier
    Aminian, Adel
    Tzikas, Apostolos
    Saw, Jacqueline
    Nielsen-Kudsk, Jens-Erick
    Ghanem, Alexander
    Schmidt, Boris
    Hildick-Smith, David
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2020, 59 (01) : 71 - 78
  • [24] Temporal changes and clinical significance of peridevice leak following left atrial appendage occlusion with Amplatzer devices
    Korsholm, Kasper
    Jensen, Jesper M.
    Norgaard, Bjarne L.
    Nielsen-Kudsk, Jens E.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2022, 99 (07) : 2071 - 2079
  • [25] Residual leak following left atrial appendage occlusion with lariat closed with amplatzer PFO occluder device
    Bawa, Danish
    Ahmed, Adnan
    Darden, Douglas
    Lakkireddy, Dhanunjaya
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2023, 46 (05): : 440 - 442
  • [26] Thrombus formation after left atrial appendage occlusion with the Amplatzer Amulet Device
    Sedaghat, A.
    Andrie, R.
    Schueler, R.
    Schrickel, J. W.
    Nickenig, G.
    Hammerstingl, C.
    EUROPEAN HEART JOURNAL, 2016, 37 : 276 - 276
  • [27] Echocardiographic Guidance of AMPLATZER Amulet Left Atrial Appendage Occlusion Device Placement
    Subramaniam, Kathirvel
    Ibarra, Andrea
    Boisen, Michael L.
    SEMINARS IN CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2019, 23 (02) : 248 - 255
  • [28] Usefulness of a novel sizing chart for left atrial appendage occlusion with the Amplatzer Amulet
    Freixa, Xavier
    Cepas-Guillen, Pedro
    Korsholm, Kasper
    Kramer, Anders
    Regueiro, Ander
    Flores-Umanzor, Eduardo
    Sanchis, Laura
    Arenas-Loriente, Andrea
    Jensen, Jesper Moller
    Nielsen-Kudsk, Jens Erik
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2022, 100 (05) : 801 - 809
  • [29] Transcatheter left atrial appendage occlusion with Amplatzer devices to obviate anticoagulation in patients with atrial fibrillation
    Meier, B
    Palacios, I
    Windecker, S
    Rotter, M
    Cao, QL
    Keane, D
    Ruiz, CE
    Hijazi, ZM
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 60 (03) : 417 - 422
  • [30] Peridevice Leak Following Amplatzer Left Atrial Appendage Occlusion Cardiac Computed Tomography Classification and Clinical Outcomes
    Korsholm, Kasper
    Jensen, Jesper Moller
    Norgaard, Bjarne Linde
    Samaras, Athanasios
    Saw, Jacqueline
    Berti, Sergio
    Tzikas, Apostolos
    Nielsen-Kudsk, Jens Erik
    JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (01) : 83 - 93