Synthesising two-dimensional mammographic images using compressed sensing-reconstructed digital breast tomosynthesis images

被引:0
|
作者
Han, Seohee [1 ]
Yang, Hyesun [1 ]
Xu, Wenting [1 ]
Seo, Changwoo [1 ]
Cho, Hyosung [1 ]
Cha, Bo Kyung [2 ]
机构
[1] Yonsei Univ, Dept Radiat Convergence Engn, Wonju, South Korea
[2] Korea Electrotechnol Res Inst, Precis Med Device Res Ctr, Ansan, South Korea
来源
JOURNAL OF INSTRUMENTATION | 2024年 / 19卷 / 08期
关键词
Data processing methods; Medical-image reconstruction methods and algorithms; computer-aided diagnosis; X-ray mammography and scinto- and MRI-mammography;
D O I
10.1088/1748-0221/19/08/P08012
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This study presents a method synthesizing a two-dimensional (2D) mammographic image of reasonable quality directly from a 3D digital breast tomosynthesis (DBT) image, reconstructed using an advanced compressed sensing (CS)-based algorithm. This approach aims to reduce the radiation dose required for complementary imaging technologies of digital mammography (DM) and DBT, eliminating the need for additional DM examinations. The method involves three main steps: projection data acquisition from a DBT system, CS-based DBT reconstruction, and synthesis of a 2D mammographic image from the reconstructed DBT image. To verify the efficacy of the proposed method, we conducted both a simulation and an experiment on a numerical breast and commercially available BR3D phantoms, respectively, prior to practical implementation in real-world DBT systems. Our simulation and experimental results indicated that the CS-based algorithm yielded markedly improved DBT reconstruction quality, preserving superior image homogeneity, better edge contour and sharpening, and fewer image artifacts. The measured contrast-to-noise ratio and structural similarity values of the CS-reconstructed DBT images were 10.31 and 0.78, respectively, which were approximately 2.2 and 3.1 times larger, respectively, than those of the filtered-backprojectionreconstructed DBT images in the simulation. The quality of the synthetic mammographic images using the CS-reconstructed DBT images was similar to that of conventional mammographic images obtained using a full dose, indicating the efficacy of the proposed method. Consequently, we successfully reconstructed DBT images of substantially high quality using the CS-based algorithm and synthesised 2D mammographic images of reasonable quality, potentially reducing the radiation dose to patients in the complementary imaging technologies of DM and DBT.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Two-View Digital Breast Tomosynthesis Screening with Synthetically Reconstructed Projection Images: Comparison with Digital Breast Tomosynthesis with Full-Field Digital Mammographic Images
    Skaane, Per
    Bandos, Andriy I.
    Eben, Ellen B.
    Jebsen, Ingvild N.
    Krager, Mona
    Haakenaasen, Unni
    Ekseth, Ulrika
    Izadi, Mina
    Hofvind, Solveig
    Gullien, Randi
    [J]. RADIOLOGY, 2014, 271 (03) : 655 - 663
  • [2] Breast Cancer Conspicuity on Simultaneously Acquired Digital Mammographic Images versus Digital Breast Tomosynthesis Images
    Korhonen, Katrina E.
    Conant, Emily E.
    Cohen, Eric A.
    Synnestvedt, Marie
    McDonald, Elizabeth S.
    Weinstein, Susan P.
    [J]. RADIOLOGY, 2019, 292 (01) : 69 - 76
  • [3] Utility of Digital Breast Tomosynthesis with Two-Dimensional Synthesized Mammography Images: A Pictorial Essay
    Hande, Pradipta C.
    Desai, Sabita S.
    Arneja, Sarabjeet K.
    Sathian, Sreedevi
    [J]. INDIAN JOURNAL OF RADIOLOGY AND IMAGING, 2021, 31 (03): : 678 - 688
  • [4] Effect of Denoising on the Quality of Reconstructed Images in Digital Breast Tomosynthesis
    Vieira, Marcelo A. C.
    Bakic, Predrag R.
    Maidment, Andrew D. A.
    [J]. MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [5] Semi-automated Segmentation and Classification of Digital Breast Tomosynthesis Reconstructed Images
    Vedantham, Srinivasan
    Shi, Linxi
    Karellas, Andrew
    Michaelsen, Kelly E.
    Krishnaswamy, Venkataramanan
    Pogue, Brian W.
    Paulsen, Keith D.
    [J]. 2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 6188 - 6191
  • [6] Computerized detection of mammographic masses in digital breast tomosynthesis images using radial gradient index filtering
    Reiser, I
    Nishikawa, RM
    Giger, ML
    Rafferty, EA
    Kopans, DB
    Moore, R
    Wu, T
    [J]. CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 1352 - 1352
  • [7] Homotopy in two-dimensional digital images
    Malgouyres, R
    [J]. THEORETICAL COMPUTER SCIENCE, 2000, 230 (1-2) : 221 - 233
  • [8] Digital breast tomosynthesis: a comparison of the accuracy of digital breast tomosynthesis, two-dimensional digital mammography and two-dimensional screening mammography (film-screen)
    MJ Michell
    RK Wasan
    P Whelehan
    A Iqbal
    CP Lawinski
    AN Donaldson
    DR Evans
    C Peacock
    ARM Wilson
    [J]. Breast Cancer Research, 11 (Suppl 2)
  • [9] Image Quality Comparison between Digital Breast Tomosynthesis Images and 2D Mammographic Images Using the CDMAM Test Object
    Tsalafoutas, Ioannis A.
    Epistatou, Angeliki C.
    Delibasis, Konstantinos K.
    [J]. JOURNAL OF IMAGING, 2022, 8 (08)
  • [10] Comparison of Digital Breast Tomosynthesis and Two-dimensional Mammography
    Korporaal, Johannes G.
    Mertelmeier, Thomas
    [J]. RADIOLOGY, 2016, 280 (03) : 980 - 981