Clinical implementation of a commercial synthetic computed tomography solution for radiotherapy treatment of glioblastoma

被引:2
|
作者
Emin, Sevgi [1 ]
Rossi, Elia [2 ]
Rooth, Elisabeth Myrvold [2 ]
Dorniok, Torsten [1 ]
Hedman, Mattias [2 ,3 ]
Gagliardi, Giovanna [1 ,3 ]
Villegas, Fernanda [1 ,3 ]
机构
[1] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, S-17176 Stockholm, Sweden
[2] Karolinska Univ Hosp, Dept Radiat Oncol, S-17176 Stockholm, Sweden
[3] Karolinska Inst, Dept Oncol Pathol, S-17177 Stockholm, Sweden
关键词
MR-only; Glioblastoma; Radiotherapy; Synthetic-CT; ACCURACY; MRI;
D O I
10.1016/j.phro.2024.100589
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and Purpose: Magnetic resonance (MR)-only radiotherapy (RT) workflow eliminates uncertainties due to computed tomography (CT)-MR image registration, by using synthetic CT (sCT) images generated from MR. This study describes the clinical implementation process, from retrospective commissioning to prospective validation stage of a commercial artificial intelligence (AI)-based sCT product. Evaluation of the dosimetric performance of the sCT is presented, with emphasis on the impact of voxel size differences between image modalities. Materials and methods: sCT performance was assessed in glioblastoma RT planning. Dose differences for 30 patients in both commissioning and validation cohorts were calculated at various dose-volume-histogram (DVH) points for target and organs-at-risk (OAR). A gamma analysis was conducted on regridded image plans. Quality assurance (QA) guidelines were established based on commissioning phase results. Results: Mean dose difference to target structures was found to be within +/- 0.7 % regardless of image resolution and cohort. OARs' mean dose differences were within +/- 1.3 % for plans calculated on regridded images for both cohorts, while differences were higher for plans with original voxel size, reaching up to -4.2 % for chiasma D2% in the commissioning cohort. Gamma passing rates for the brain structure using the criteria 1 %/1mm, 2 %/2mm and 3 %/3mm were 93.6 %/99.8 %/100 % and 96.6 %/99.9 %/100 % for commissioning and validation cohorts, respectively. Conclusions: Dosimetric outcomes in both commissioning and validation stages confirmed sCT's equivalence to CT. The large patient cohort in this study aided in establishing a robust QA program for the MR-only workflow, now applied in glioblastoma RT at our center.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Evaluation of a commercial synthetic computed tomography generation solution for magnetic resonance imaging-only radiotherapy
    Gonzalez-Moya, A.
    Dufreneix, S.
    Ouyessad, N.
    Guillerminet, C.
    Autret, D.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (06): : 191 - 197
  • [2] Feasibility of magnetic resonance imaging-only rectum radiotherapy with a commercial synthetic computed tomography generation solution
    Maspero, Matteo
    Tyyger, Marcus D.
    Tijssen, Rob H. N.
    Seevinck, Peter R.
    Intven, Martijn P. W.
    van den Berg, Cornelis A. T.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 7 : 58 - 64
  • [3] Challenges and opportunities in the development and clinical implementation of artificial intelligence based synthetic computed tomography for magnetic resonance only radiotherapy
    Villegas, Fernanda
    Dal Bello, Riccardo
    Alvarez-Andres, Emilie
    Dhont, Jennifer
    Janssen, Tomas
    Milan, Lisa
    Robert, Charlotte
    Salagean, Ghizela-Ana-Maria
    Tejedor, Natalia
    Trnkova, Petra
    Fusella, Marco
    Placidi, Lorenzo
    Cusumano, Davide
    RADIOTHERAPY AND ONCOLOGY, 2024, 198
  • [4] Clinical target volume definition for glioblastoma radiotherapy planning: magnetic resonance imaging and computed tomography
    A. Fiorentino
    R. Caivano
    P. Pedicini
    V. Fusco
    Clinical and Translational Oncology, 2013, 15 : 754 - 758
  • [5] Clinical target volume definition for glioblastoma radiotherapy planning: magnetic resonance imaging and computed tomography
    Fiorentino, A.
    Caivano, R.
    Pedicini, P.
    Fusco, V.
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2013, 15 (09): : 754 - 758
  • [6] Metal Artifact Reduction in Radiotherapy - Implementation of iMAR corrected Computed Tomography and clinical Evaluation
    Sabatino, M.
    Kretschmer, M.
    Kasch, A.
    Zimmermann, M.
    Clos, S.
    Heyden, S.
    Dahle, J.
    Wuerschmidt, F.
    Allianz, Radiologische
    STRAHLENTHERAPIE UND ONKOLOGIE, 2017, 193 : S102 - S102
  • [7] COMPUTED-TOMOGRAPHY AND RADIOTHERAPY IN THE TREATMENT OF CANCER
    JOSE, B
    CHU, AM
    SHARMA, SC
    TOBIN, DA
    SCOTT, RM
    JOURNAL OF SURGICAL ONCOLOGY, 1983, 23 (02) : 83 - 92
  • [8] Current and future developments of synthetic computed tomography generation for radiotherapy
    van Elmpt, Wouter
    Taasti, Vicki Trier
    Redalen, Kathrine Roe
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 28
  • [9] A treatment-site-specific evaluation of commercial synthetic computed tomography solutions for proton therapy
    Yeap, Ping Lin
    Wong, Yun Ming
    Lee, Kang Hao
    Koh, Calvin Wei Yang
    Lew, Kah Seng
    Chua, Clifford Ghee Ann
    Wibawa, Andrew
    Master, Zubin
    Lee, James Cheow Lei
    Park, Sung Yong
    Tan, Hong Qi
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 31
  • [10] Clinical implementation: 4D computed tomography for thoracic tumor treatment planning
    Rietzel, E
    Liu, AK
    Chen, GT
    Doppke, KP
    Choi, NC
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S609 - S610