The biomechanical effect of lumbopelvic distance reduction on reconstruction after total sacrectomy: a comparative finite element analysis of four techniques

被引:0
|
作者
Turbucz, Mate [1 ,2 ]
Pokorni, Agoston Jakab [1 ,2 ]
Hajnal, Benjamin [1 ,2 ]
Koch, Kristof [1 ,3 ]
Szoverfi, Zsolt [3 ]
Varga, Peter Pal [3 ]
Lazary, Aron [3 ,4 ]
Eltes, Peter Endre [2 ,3 ,4 ]
机构
[1] Semmelweis Univ, Sch PhD Stud, Ulloi Str 26, Budapest, Hungary
[2] Natl Ctr Spinal Disorders, In Sil Biomech Lab, Kiralyhago Str 1-3, Budapest, Hungary
[3] Natl Ctr Spinal Disorders, Kiralyhago Str 1-3, H-1126 Budapest, Hungary
[4] Semmelweis Univ, Dept Spine Surg, Dept Orthopaed, Ulloi Str 78-b, H-1082 Budapest, Hungary
来源
SPINE JOURNAL | 2024年 / 24卷 / 10期
基金
匈牙利科学研究基金会;
关键词
Biomechanics; Finite element analysis; Lumbopelvic distance reduction; Lumbopelvic fixation; Lumbopelvic reconstruction; Lumbopelvic stabilization; Total sacrectomy; LUMBAR MOTION SEGMENT; SACRAL TUMORS; MECHANICAL-PROPERTIES; ROD INSTRUMENTATION; CALIBRATION METHOD; 4-ROD TECHNIQUE; PELVIC RING; MODEL; SPINE; BONE;
D O I
10.1016/j.spinee.2024.04.024
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND CONTEXT: Following total sacrectomy, lumbopelvic reconstruction is essential to restore continuity between the lumbar spine and pelvis. However, to achieve long-term clinical stability, bony fusion between the lumbar spine and the pelvic ring is crucial. Reduction of the lumbopelvic distance can promote successful bony fusion. Although many lumbopelvic reconstruction techniques (LPRTs) have been previously analyzed, the biomechanical effect of lumbopelvic distance reduction (LPDR) has not been investigated yet. PURPOSE: To evaluate and compare the biomechanical characteristics of four different LPRTs while considering the effect of LPDR. STUDY DESIGN/SETTING: A comparative finite element (FE) study. METHODS: The FE models following total sacrectomy were developed to analyze four different LPRTs, with and without LPDR. The closed-loop reconstruction (CLR), the sacral-rod reconstruction (SRR), the four-rod reconstruction (FRR), and the improved compound reconstruction (ICR) techniques were analyzed in flexion, extension, lateral bending, and axial rotation. Lumbopelvic stability was assessed through the shift-down displacement and the relative sagittal rotation of L5, while implant safety was evaluated based on the stress state at the bone-implant interface and within the rods. RESULTS: Regardless of LPDR, both the shift-down displacement and relative sagittal rotation of L5 consistently ranked the LPRTs as ICR25% in CLR, by 61% in SRR, by 15% in FRR, and by 46% in ICR, as well as reduced the relative sagittal rotation values by 21% in CLR, by 73% in SRR, by 11% in FRR, and by 53% in ICR. Considering the stress at the bone- implant interface, without LPDR, the ICR yielded the smallest stress values for flexion, lateral bending, and axial rotation with 131.4 MPa, 68.2 MPa, and 70.3 MPa, respectively, and the second smallest in extension with 36.1 MPa. Due to LPDR, these stress values were reduced by 31% in flexion, by 17% in extension, by 29% in lateral bending, and by 29% in axial rotation. Within the rods, without LPDR, the ICR yielded the smallest stress values for flexion, extension, lateral bending, and axial rotation with 346.5 MPa, 108.0 MPa, 186.2 MPa, and 199.7 MPa, respectively. With LPDR, these stress values were reduced by 16% in flexion, by 9% in extension, by 11% in lateral CONCLUSIONS: LPDR significantly improved both lumbopelvic stability and implant safety in all reconstruction techniques after total sacrectomy. LPDR reduced the shift-down displacement of L5, the relative sagittal rotation of L5, and the stress values at the bone-implant interface. Furthermore, in the ICR and SRR techniques, LPDR decreased the peak stress values within the rods. All four investigated LPRTs demonstrated suitability for lumbopelvic reconstruction, with the ICR following total sacrectomy; therefore, it has the potential to impact the design of custom-made 3D-printed or traditional LPRTs. However, to confirm the findings of the current FE study, longterm clinical trials are recommended.
引用
收藏
页码:1981 / 1992
页数:12
相关论文
共 50 条
  • [41] Biomechanical effect of implant design on four implants supporting mandibular full-arch fixed dentures: In vitro test and finite element analysis
    Wu, Aaron Yu-Jen
    Hsu, Jui-Ting
    Fuh, Lih-Jyh
    Huang, Heng-Li
    JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2020, 119 (10) : 1514 - 1523
  • [42] Biomechanical comparison between metal block and cement-screw techniques for the treatment of tibial bone defects in total knee arthroplasty based on finite element analysis
    Liu, Yang
    Zhang, Aobo
    Wang, Chenyu
    Yin, Weihuang
    Wu, Naichao
    Chen, Hao
    Chen, Bingpeng
    Han, Qing
    Wang, Jincheng
    COMPUTERS IN BIOLOGY AND MEDICINE, 2020, 125
  • [43] Reconstruction of anterior talofibular ligament and posterior tibiotalar ligament enhance ankle stability after total talus replacement by finite element analysis
    Li, Hao
    Xie, Haitao
    Kang, Shujing
    Xu, Kuixue
    Huang, Xiaoyi
    Xie, Haiqiong
    Cai, Xu
    Chen, Wan
    Wei, Kai
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2025,
  • [44] Comparative study of biomechanical effects between two types of 2 × 4 techniques employing a rocking-chair archwire: a three-dimensional finite element analysis
    Shaotai Wang
    Min Hu
    Sihan Wang
    Huichuan Qi
    Dongsheng Song
    Huan Jiang
    Clinical Oral Investigations, 2023, 27 : 4617 - 4631
  • [45] 3-DIMENSIONAL SHAPE RECONSTRUCTION AND FINITE-ELEMENT ANALYSIS OF FEMUR BEFORE AND AFTER THE CEMENTLESS TYPE OF TOTAL HIP-REPLACEMENT
    KANG, YK
    PARK, HC
    YOUM, Y
    LEE, IK
    AHN, MH
    IHN, JC
    JOURNAL OF BIOMEDICAL ENGINEERING, 1993, 15 (06): : 497 - 504
  • [46] Wear Risk Prevention and Reduction in Total Hip Arthroplasty. A Personalized Study Comparing Cement and Cementless Fixation Techniques Employing Finite Element Analysis
    Gonzalez-Bravo, Carlos
    Ortega, Miguel A.
    Bujan, Julia
    de la Torre, Basilio
    Barrios, Loreto
    JOURNAL OF PERSONALIZED MEDICINE, 2021, 11 (08):
  • [47] Biomechanical effect of anterior talofibular ligament injury in Weber B lateral malleolus fractures after lateral plate fixation: A finite element analysis
    Zhou, Lin
    Lin, Jiaming
    Wang, Bata
    Gan, Weidong
    Huang, Anfei
    Lin, Yongxin
    FOOT AND ANKLE SURGERY, 2020, 26 (08) : 871 - 875
  • [48] Biomechanical effect of bone resorption of the spinous process after single-segment interspinous dynamic stabilization device implantation A finite element analysis
    Zhu, Zhen-Qi
    Duan, Shuo
    Wang, Kai-Feng
    Liu, Hai-Ying
    Xu, Shuai
    Liu, Chen-Jun
    MEDICINE, 2018, 97 (27)
  • [49] A finite element analysis of the pelvic reconstruction using fibular transplantation fixed with four different rod-screw systems after type I resection
    Jia Yong-Wei
    Cheng Li-Ming
    Yu Guang-Rong
    Du Cheng-fei
    Yang Zhi-yong
    Yu Yan
    Ding Zu-quan
    CHINESE MEDICAL JOURNAL, 2008, 121 (04) : 321 - 326
  • [50] Comparative study of biomechanical effects between two types of 2 x 4 techniques employing a rocking-chair archwire: a three-dimensional finite element analysis
    Wang, Shaotai
    Hu, Min
    Wang, Sihan
    Qi, Huichuan
    Song, Dongsheng
    Jiang, Huan
    CLINICAL ORAL INVESTIGATIONS, 2023, 27 (08) : 4617 - 4631