Application of Deep Learning for Real-Time Ablation Zone Measurement in Ultrasound Imaging

被引:0
|
作者
Zimmermann, Corinna [1 ]
Michelmann, Adrian [1 ]
Daniel, Yannick [1 ]
Enderle, Markus D. [1 ]
Salkic, Nermin [1 ,2 ]
Linzenbold, Walter [1 ]
机构
[1] Erbe Elektromed GmbH, D-72072 Tubingen, Germany
[2] Univ Tuzla, Fac Med, Tuzla 75000, Bosnia & Herceg
关键词
radiofrequency ablation; ultrasonography; artificial intelligence; image processing; computer-assisted; ablation techniques; RADIOFREQUENCY ABLATION; MICROWAVE ABLATION; LIVER; LENGTH; ACCURACY; TISSUE;
D O I
10.3390/cancers16091700
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The accurate delineation of ablation zones (AZs) is crucial for assessing radiofrequency ablation (RFA) therapy's efficacy. Manual measurement, the current standard, is subject to variability and potential inaccuracies. Aim: This study aims to assess the effectiveness of Artificial Intelligence (AI) in automating AZ measurements in ultrasound images and compare its accuracy with manual measurements in ultrasound images. Methods: An in vitro study was conducted using chicken breast and liver samples subjected to bipolar RFA. Ultrasound images were captured every 15 s, with the AI model Mask2Former trained for AZ segmentation. The measurements were compared across all methods, focusing on short-axis (SA) metrics. Results: We performed 308 RFA procedures, generating 7275 ultrasound images across liver and chicken breast tissues. Manual and AI measurement comparisons for ablation zone diameters revealed no significant differences, with correlation coefficients exceeding 0.96 in both tissues (p < 0.001). Bland-Altman plots and a Deming regression analysis demonstrated a very close alignment between AI predictions and manual measurements, with the average difference between the two methods being -0.259 and -0.243 mm, for bovine liver and chicken breast tissue, respectively. Conclusion: The study validates the Mask2Former model as a promising tool for automating AZ measurement in RFA research, offering a significant step towards reducing manual measurement variability.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Contrast-enhanced ultrasound improves real-time imaging of ablation region during radiofrequency ablation: Preliminary results
    Wiggermann, P.
    Zuber-Jerger, I.
    Zausig, Y.
    Loss, M.
    Scherer, M. N.
    Schreyer, A. G.
    Stroszczynski, C.
    Jung, E. M.
    CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2011, 49 (1-4) : 43 - 54
  • [42] Deep Learning for Real -time, Automatic, and Scanner-adapted Prostate (Zone) Segmentation of Transrectal Ultrasound, for Example, Magnetic Resonance Imaging ?transrectal Ultrasound
    van Sloun, Ruud J. G.
    Wildeboer, Rogier R.
    Mannaerts, Christophe K.
    Postema, Arnoud W.
    Gayet, Maudy
    Beerlage, Harrie P.
    Salomon, Georg
    Wijkstra, Hessel
    Mischi, Massimo
    EUROPEAN UROLOGY FOCUS, 2021, 7 (01): : 78 - 85
  • [43] Deep learning for real-time multi-class segmentation of artefacts in lung ultrasound
    Howell, Lewis
    Ingram, Nicola
    Lapham, Roger
    Morrell, Adam
    McLaughlan, James R.
    ULTRASONICS, 2024, 140
  • [44] Deep Learning-Based Algorithms for Real-Time Lung Ultrasound Assisted Diagnosis
    Muñoz, Mario
    Rubio, Adrián
    Cosarinsky, Guillermo
    Cruza, Jorge F.
    Camacho, Jorge
    Applied Sciences (Switzerland), 2024, 14 (24):
  • [45] REAL-TIME ULTRASOUND FOR THE DETECTION OF DEEP VENOUS THROMBOSIS
    CHANCE, JF
    ABBITT, PL
    TEGTMEYER, CJ
    POWERS, RD
    ANNALS OF EMERGENCY MEDICINE, 1991, 20 (05) : 494 - 496
  • [46] THE MEASUREMENT OF GASTRIC MOTOR FUNCTION BY REAL-TIME ULTRASOUND
    BATEMAN, DN
    WHITTINGHAM, TA
    BRITISH JOURNAL OF PHARMACOLOGY, 1980, 70 (01) : P188 - P188
  • [47] GASTRIC-EMPTYING MEASUREMENT BY REAL-TIME ULTRASOUND
    BATEMAN, DN
    WHITTINGHAM, TA
    ZEITSCHRIFT FUR GASTROENTEROLOGIE, 1981, 19 (08): : 417 - 417
  • [48] Generic Application of Deep Learning Framework for Real-Time Engineering Data Analysis
    Majdani, Farzan
    Petrovski, Andrei
    Petrovski, Sergei
    2018 INTERNATIONAL JOINT CONFERENCE ON NEURAL NETWORKS (IJCNN), 2018,
  • [49] Real-time noise cancellation with deep learning
    Porr, Bernd
    Daryanavard, Sama
    Bohollo, Lucia Munoz
    Cowan, Henry
    Dahiya, Ravinder
    PLOS ONE, 2022, 17 (11):
  • [50] Application of Deep Reinforcement Learning in Real-time Plan Scheduling of Power Grid
    Liu J.
    Song X.
    Yang N.
    Wan X.
    Cai Y.
    Huang Y.
    Dianli Xitong Zidonghua/Automation of Electric Power Systems, 2023, 47 (14): : 157 - 166