Polymer composite fiber-optic Fabry-Perot cavity sensor based on Ti3C2TX MXene enhanced photothermal effect for low-concentration bilirubin detection

被引:0
|
作者
Zhang, Yangyuan [1 ]
Miao, Yinping [1 ]
Chen, Xuanyi [1 ]
Bai, Yangbo [1 ]
Yao, Jianquan [2 ]
机构
[1] Tianjin Univ Technol, Sch Integrated Circuit Sci & Engn, Tianjin Key Lab Film Elect & Commun Device, Tianjin 300384, Peoples R China
[2] Tianjin Univ, Inst Laser & Optoelect, Coll Precis Instrument & Optoelect Engn, Tianjin 300072, Peoples R China
基金
中国国家自然科学基金;
关键词
Absorption spectroscopy - Composite materials - Diagnosis - Fabry-Perot interferometers - Neurodegenerative diseases - Single mode fibers;
D O I
10.1063/5.0215454
中图分类号
O59 [应用物理学];
学科分类号
摘要
The World Health Statistics 2023, as presented by the World Health Organization (WHO), underscores a significant global health concern: 73% of global mortality is attributed to chronic diseases. Within the domain of medical diagnostics, serum bilirubin levels serve as a critical biomarker for identifying several prevalent chronic conditions in adult and geriatric populations. In this study, a polymer composite fiber-optic Fabry-P & eacute;rot interferometer is prepared, wherein the reflective surfaces are, respectively, constituted by polyurethane (PU), with characteristics marked by high thermal responsiveness and rapid thermal transfer rates, and a single-mode fiber. Subsequently, the outer surface of the PU reflective end undergoes deposition with Ti3C2TX MXene material, which has the capability to enhance the adsorption of bilirubin molecules. This work proposes a detection method using the photothermal effect of bilirubin. When the sensor is operational, bilirubin generates a thermal effect under 405 nm laser irradiation. The Fabry-P & eacute;rot air cavity rapidly expands with the increased heat absorption, leading to shifts in the interference spectrum, which enables the effective detection of low concentrations of bilirubin within the serum. The experimental results indicate that the sensor exhibits a sensitivity of 7.51 [nm/(mu mol/L)] and a detection limit of a mere 0.002 19 mu mol/L within the linear dynamic range of 1-10 mu mol/L. Furthermore, a tenfold improvement is observed when compared to the previously reported lowest detection limit. The sensor delineated herein is characterized by heightened sensitivity, minimal detection limit, rapid response time, and substantial robustness, rendering it suitable for potential applications in the early diagnosis and prognosis of chronic diseases such as Alzheimer's disease.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Cu2Se/MXene (Ti3C2Tx) composite achieved ultra-low thermal conductivity and enhanced thermoelectric
    Gu, Xueke
    Fan, Pengya
    Wang, Chao
    Cui, Shengqiang
    Wang, Xinxin
    You, Guangmeng
    Peng, Chengxiao
    Yang, Gui
    PHYSICA B-CONDENSED MATTER, 2024, 695
  • [22] Novel Pd-doped-CuO/Ti3C2Tx integrated nanocomposite with improved performance for low-concentration hexanal detection
    Chai, Xiaojie
    Zhang, Kaibin
    Wang, Miaoru
    Sang, Shengbo
    Hu, Xiaobing
    CERAMICS INTERNATIONAL, 2025, 51 (02) : 1918 - 1928
  • [23] Stretchable Ti3C2Tx MXene/Carbon Nanotube Composite Based Strain Sensor with Ultrahigh Sensitivity and Tunable Sensing Range
    Cai, Yichen
    Shen, Jie
    Ge, Gang
    Zhang, Yizhou
    Jin, Wanqin
    Huang, Wei
    Shao, Jinjun
    Yang, Jian
    Dong, Xiaochen
    ACS NANO, 2018, 12 (01) : 56 - 62
  • [24] High-Performance Ammonia Detection of Polymeric BaTiO3/Ti3C2Tx MXene Composite-Based Sensor for Gas Emission and Leakage
    Sun, Guoqing
    Wang, Chenglin
    Jia, Jie
    Zhang, Hao
    Hu, Yaqing
    Liu, Yukun
    Zhang, Dongzhi
    JOURNAL OF ELECTRONIC MATERIALS, 2024, 53 (07) : 3426 - 3437
  • [25] A fuel cell type gas sensor based on Pt/Ti3C2Tx MXene for highly selective detection of NH3
    Li, Bingyang
    Liu, Jiaxin
    Lin, Zhen
    Mai, Chunwei
    Cao, Anjie
    Yao, Dong
    Chen, Zhanshen
    Qu, Fengdong
    MATERIALS LETTERS, 2025, 383
  • [26] Simultaneous detection of 4-chlorophenol and 4-nitrophenol using a Ti3C2Tx MXene based electrochemical sensor
    Lei, Ling
    Li, Caoling
    Huang, Wensheng
    Wu, Kangbing
    ANALYST, 2021, 146 (24) : 7593 - 7600
  • [27] Wearable CNT/Ti3C2Tx MXene/PDMS composite strain sensor with enhanced stability for real-time human healthcare monitoring
    Xiaowen Xu
    Yucheng Chen
    Pei He
    Song Wang
    Kai Ling
    Longhui Liu
    Pengfei Lei
    Xianjun Huang
    Hu Zhao
    Jianyun Cao
    Junliang Yang
    Nano Research, 2021, 14 : 2875 - 2883
  • [28] Wearable CNT/Ti3C2Tx MXene/PDMS composite strain sensor with enhanced stability for real-time human healthcare monitoring
    Xu, Xiaowen
    Chen, Yucheng
    He, Pei
    Wang, Song
    Ling, Kai
    Liu, Longhui
    Lei, Pengfei
    Huang, Xianjun
    Zhao, Hu
    Cao, Jianyun
    Yang, Junliang
    NANO RESEARCH, 2021, 14 (08) : 2875 - 2883
  • [29] A High-Performance Wearable Ag/Ti3C2Tx MXene-Based Fiber Sensor for Temperature Sensing, Pressure Sensing, and Human Motion Detection
    Liu, Hailian
    Yang, Ning
    Zhang, Qi
    Wang, Fang
    Yan, Xin
    Zhang, Xuenan
    Cheng, Tonglei
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2022, 71
  • [30] Long period fiber grating sensor coated with MXene (Ti3C2TX) functionalized by poly-levodopa for rapid detection of Cu2+
    Yan M.
    Li Y.
    Li Y.
    Liu B.
    Wang R.
    Jiang M.
    Optik, 2023, 288