Effect of photons and electrons on the over-response of optical fiber X-ray sensors

被引:0
|
作者
He, Bo [1 ]
Xie, Tianci [1 ]
Zhang, Bin [1 ]
Qian, Jinqian [1 ]
Li, Haopeng [1 ]
Shi, Qieming [1 ]
Hao, Wenjing [1 ]
Qin, Zhuang [2 ]
Lewis, Elfed [3 ]
Sun, Weimin [1 ]
机构
[1] Harbin Engn Univ, Key Lab In Fiber Integrated Opt, Minist Educ China, Harbin 150001, Peoples R China
[2] Heilongjiang Univ, Elect Engn Coll, Minist Educ China, Harbin 50001, Peoples R China
[3] Univ Limerick, Opt Fiber Sensors Res Ctr, Limerick, Ireland
基金
爱尔兰科学基金会; 中国国家自然科学基金;
关键词
Radiotherapy fiber X-ray sensors; Over-response; Monte Carlo simulation; RADIATION DOSIMETER; DETECTORS;
D O I
10.1016/j.sna.2024.115649
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Optical fiber X-ray sensors are prone to over -response compared to ionization chambers (ICs) when used to measure the percentage depth dose (PDD) curves and off axis ratio (OAR) curves. Investigating this phenomenon is crucial for the successful adoption of these sensors in clinical radiotherapy. Compared with ion chambers and other water equivalent dosimeters, the fiber X-ray sensors using high Z eff scintillators (Gd 2 O 2 S:Tb) have generally higher mass attenuation coefficients. In this case, the response of scintillators to photons needs to be considered. In this article, a Monte -Carlo (MC) software package BEAMnrc was used to simulate the distribution of photon and electron fluxes in a clinical based water phantom, besides the energy deposition of air and Gd 2 O 2 S:Tb in varying energy was also simulated. The distribution of the photon and electron fluxes in the horizontal (radial) direction at a fixed 10 cm water depth was studied, as well as the fluxes in the vertical (depth) direction. An embedded -structure fiber X-ray sensor was used for the experimental measurements. The simulation results indicate that, in the horizontal direction, there are many photons remain outside the radiation field. In the vertical direction, as the size of the radiation field increases, the maximum point of the photon flux shifts towards a deeper water depth. These behavior of photon flux is consistent with the fiber X-ray sensors measurement. However, the maximum point of the electron flux remains unchanged in the vertical direction. In the horizontal direction, the electron flux decreases rapidly, and this scenario is similar to the case when using the ion chamber. Results of analyzing the spectra inside and outside the radiation field show that there is a large number of photons with energy below 0.2 MeV outside the radiation field, and the proportion of these low -energy photons is greater than 60 %. The findings indicate that the fiber X-ray sensors ' over -response can be primarily attributed to the lower energy scattered photons.
引用
收藏
页数:9
相关论文
共 50 条
  • [11] Energetic electrons and x-ray photons from multiterawatt Ti:sapphire lasers
    Nickles, PV
    Kalachnikov, MP
    Warwick, PJ
    Janulewicz, KA
    Sandner, W
    Jahnke, U
    Hilscher, D
    Schnürer, M
    Nolte, R
    Rousse, A
    [J]. QUANTUM ELECTRONICS, 1999, 29 (05) : 444 - 448
  • [12] THE SCATTERING OF X-RAY PHOTONS
    COMPTON, AH
    [J]. AMERICAN JOURNAL OF PHYSICS, 1946, 14 (02) : 80 - 84
  • [13] X-ray electrons
    Whiddington, R
    [J]. PHILOSOPHICAL MAGAZINE, 1922, 43 (258): : 1116 - +
  • [14] MICROSTRUCTURED OPTICAL FIBER FOR X-RAY DETECTION
    DeHaven, S. L.
    [J]. REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 29A AND 29B, 2010, 1211 : 541 - 548
  • [15] The classical distribution of the electrons in the photoelectric effect of the x-ray
    Kunz, J
    [J]. PHYSIKALISCHE ZEITSCHRIFT, 1933, 34 : 219 - 220
  • [16] New model for explaining the over-response phenomenon in percentage of depth dose curve measured using inorganic scintillating materials for optical fiber radiation sensors
    Qin, Zhuang
    Xie, Tianci
    Dai, Xinyu
    Zhang, Bin
    Ma, Yu
    Khan, Ihsan Ullah
    Zhang, Xu
    Li, Haopeng
    Yan, Yongji
    Zhao, Wenhui
    Li, Song
    Chen, Ziyin
    Zhang, Daxin
    Xu, Jun
    Hu, Xiaokang
    Xing, Lina
    Feng, Kun
    Lewis, Elfed
    Sun, Weimin
    [J]. OPTICS EXPRESS, 2019, 27 (17): : 23693 - 23706
  • [17] Triple coincidences of x-ray photons
    Ikonen, E.
    Holopainen, S.
    [J]. PHYSICAL REVIEW A, 2007, 76 (03):
  • [18] X-ray electrons.
    Rankine, AO
    [J]. NATURE, 1922, 110 : 681 - 681
  • [19] Optical imaging in tissue with X-ray excited luminescent sensors
    Chen, Hongyu
    Longfield, David E.
    Varahagiri, Venkata S.
    Nguyen, KhanhVan T.
    Patrick, Amanda L.
    Qian, Haijun
    VanDerveer, Donald G.
    Anker, Jeffrey N.
    [J]. ANALYST, 2011, 136 (17) : 3438 - 3445
  • [20] X-RAY PHOTOELECTRON-SPECTROSCOPY WITH X-RAY PHOTONS OF HIGHER ENERGY
    WAGNER, CD
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1978, 15 (02): : 518 - 523