A comparison of intraocular pressure measurement using SUOER SW-500 rebound tonometer and conventional reusable Goldmann prisms

被引:0
|
作者
Chaung, Jia Quan [1 ,2 ]
Sangapillai, Thanendthire [1 ]
Quilat, Karen Kate [1 ]
Perera, Shamira [1 ,2 ,3 ]
机构
[1] Singapore Natl Eye Ctr, Singapore City, Singapore
[2] Singapore Eye Res Inst, Singapore City, Singapore
[3] Duke NUS Med Sch, Singapore City, Singapore
关键词
glaucoma; rebound tonometer; IOP; intraocular pressure; Goldmann applanation tonometry; APPLANATION TONOMETER; ICARE; GLAUCOMA; AGREEMENT;
D O I
10.3389/fmed.2024.1269332
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: To determine the agreement between intraocular pressure (IOP) measurements using conventional Goldmann applanation tonometry (GA(1,2)T) and SUOER SW-500 Rebound Tonometer. Methods: This was a retrospective observational study where 205 eyes of 106 glaucoma patients had their IOPs measured by 2 fellowship trained ophthalmologists. Data were analyzed using the Bland-Altman method of differences. Correlation was measured using the Pearson coefficient. Results: Most of our patients were Chinese (88.7%) and female (51.9%). The average age was 66.9 years. The range of IOPs as measured by GAT was 2 to 58 mm Hg. Using the Bland-Altman method to compare GAT and SUOER SW-500 Rebound Tonometer. The tonometer overestimated the IOP by 0.5 mm Hg in the right eye and underestimated it by 0.1 mm Hg in the left eye. Overall, the tonometer overestimated the IOP by 0.2 mmHg. The Tonometer IOP correlated well with GAT, with a Pearson coefficient of correlation(r) of 0.89 (p < 0.001) for the right eye and 0.86 (p < 0.001) for the left eye, respectively. In patients with GAT IOP >= 21 mm Hg (n = 25), the Tonometer underestimated the IOP by 2.96 mm Hg. Discussion: The IOP measurements from the SUOER SW-500 Rebound Tonometer correlates well with the conventional GAT in measuring the IOP within normal ranges of IOP. SUOER SW-500 Rebound Tonometer may be of use, especially if the risk of transmission of infection is high considering that the probes are disposable. It is easy to use and its small size and portability makes it useful in situations where the patient is unable to be examined at the slit lamp.
引用
收藏
页数:7
相关论文
共 35 条
  • [21] The effect of corneal pathology on intraocular pressure measurement using the OBF pneumotonometer, Tonopen and Goldmann applanation tonometer.
    Browning, AC
    Bhan, A
    Shah, S
    Dua, HS
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2000, 41 (04) : S282 - S282
  • [22] Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry
    Annette Lösch
    Alexander Scheuerle
    Volker Rupp
    Gerd Auffarth
    Matthias Becker
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2005, 243 : 313 - 316
  • [23] Intraocular pressure measurement using ICare rebound tonometer in different positions of eye and different locations on cornea
    Wongwanwatana, Sirada
    Treesit, Isaraporn
    Funarunart, Panrapee
    Iemsomboon, Wallop
    Choontanom, Raveewan
    MEDICINE, 2023, 102 (36) : E34874
  • [24] A comparison of postural and diurnal variations in intraocular pressure using the iCare rebound tonometer and Perkins applanation tonometer in admitted adults in Kenya
    Chikasirimobi, Timothy G.
    Ndinyo, Mercy
    Ondieki, Maxine N.
    Miti, Isaac
    Ejiochi-Iyoke, Ndidi J.
    Osuagwu, Uchechukwu L.
    Zeried, Ferial M.
    Agho, Kingsley
    Ogbuehi, Kelechi C.
    Mashige, Khathutshelo P.
    AFRICAN VISION AND EYE HEALTH JOURNAL, 2021, 80 (01):
  • [25] Reliability and reproducibility of introcular pressure (IOP) measurement with the Icare® Home rebound tonometer (model TA022) and comparison with Goldmann applanation tonometer in glaucoma patients
    Valero, B.
    Fenolland, J. -R.
    Rosenberg, R.
    Sendon, D.
    Mesnard, C.
    Sigaux, M.
    Giraud, J. -M.
    Renard, J. -P.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2017, 40 (10): : 865 - 875
  • [26] Comparison of intraocular pressure in healthy brachycephalic and nonbrachycephalic cats using the Icare® TONOVET Plus rebound tonometer
    Kerdchuchuen, Kamolchanok
    Samathayanon, Kingkarn
    Phientong, Pitchapa
    Chattraphirat, Suprapha
    Jaturakan, Orapun
    Tuntivanich, Nalinee
    VETERINARY OPHTHALMOLOGY, 2021, 24 (05) : 484 - 490
  • [27] Evaluation of biomechanically corrected intraocular pressure using Corvis ST and comparison of the Corvis ST, noncontact tonometer, and Goldmann applanation tonometer in patients with glaucoma
    Nakao, Yoshitaka
    Mohan, Rajiv R.
    Kiuchi, Yoshiaki
    Okumichi, Hideaki
    PLOS ONE, 2020, 15 (09):
  • [28] COMPARATIVE MEASUREMENT OF INTRAOCULAR-PRESSURE USING GOLDMANN APPLANATION TONOMETER AND A NEW OCULAR BLOOD-FLOW TONOGRAPH
    RONZANI, M
    LANG, GE
    SPRAUL, CW
    PILLUNAT, LE
    HOGEL, J
    LANG, GK
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1995, 36 (04) : S81 - S81
  • [29] Measurement of intraocular pressure (IOP) in chickens using a rebound tonometer: Quantitative evaluation of variance due to position inaccuracies
    Prashar, A.
    Guggenheim, J. A.
    Erichsen, J. T.
    Hocking, Pm.
    Morgan, J. E.
    EXPERIMENTAL EYE RESEARCH, 2007, 85 (04) : 563 - 571
  • [30] Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients
    Vincent, Stephen J.
    Vincent, Roslyn A.
    Shields, David
    Lee, Graham A.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (04): : E163 - E170