A comparison of intraocular pressure measurement using SUOER SW-500 rebound tonometer and conventional reusable Goldmann prisms

被引:0
|
作者
Chaung, Jia Quan [1 ,2 ]
Sangapillai, Thanendthire [1 ]
Quilat, Karen Kate [1 ]
Perera, Shamira [1 ,2 ,3 ]
机构
[1] Singapore Natl Eye Ctr, Singapore City, Singapore
[2] Singapore Eye Res Inst, Singapore City, Singapore
[3] Duke NUS Med Sch, Singapore City, Singapore
关键词
glaucoma; rebound tonometer; IOP; intraocular pressure; Goldmann applanation tonometry; APPLANATION TONOMETER; ICARE; GLAUCOMA; AGREEMENT;
D O I
10.3389/fmed.2024.1269332
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: To determine the agreement between intraocular pressure (IOP) measurements using conventional Goldmann applanation tonometry (GA(1,2)T) and SUOER SW-500 Rebound Tonometer. Methods: This was a retrospective observational study where 205 eyes of 106 glaucoma patients had their IOPs measured by 2 fellowship trained ophthalmologists. Data were analyzed using the Bland-Altman method of differences. Correlation was measured using the Pearson coefficient. Results: Most of our patients were Chinese (88.7%) and female (51.9%). The average age was 66.9 years. The range of IOPs as measured by GAT was 2 to 58 mm Hg. Using the Bland-Altman method to compare GAT and SUOER SW-500 Rebound Tonometer. The tonometer overestimated the IOP by 0.5 mm Hg in the right eye and underestimated it by 0.1 mm Hg in the left eye. Overall, the tonometer overestimated the IOP by 0.2 mmHg. The Tonometer IOP correlated well with GAT, with a Pearson coefficient of correlation(r) of 0.89 (p < 0.001) for the right eye and 0.86 (p < 0.001) for the left eye, respectively. In patients with GAT IOP >= 21 mm Hg (n = 25), the Tonometer underestimated the IOP by 2.96 mm Hg. Discussion: The IOP measurements from the SUOER SW-500 Rebound Tonometer correlates well with the conventional GAT in measuring the IOP within normal ranges of IOP. SUOER SW-500 Rebound Tonometer may be of use, especially if the risk of transmission of infection is high considering that the probes are disposable. It is easy to use and its small size and portability makes it useful in situations where the patient is unable to be examined at the slit lamp.
引用
收藏
页数:7
相关论文
共 35 条
  • [1] A Comparison of Intraocular Pressure Measurement using SUOER SW-500 Rebound Tonometer and Conventional Reusable Goldmann Prisms
    Chaung, Jia Quan
    Sangapillai, Thanendthire
    Quilat, Karen Kate
    Perera, Shamira
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [2] A Comparison of Applanation Tonometry Using Conventional Reusable Goldmann Prisms and Disposable Prisms
    Tsai, Andrew S. H.
    How, Alicia C.
    Su, Daniel H. W.
    Lee, Kelvin Y.
    Wong, Tina T.
    Perera, Shamira A.
    JOURNAL OF GLAUCOMA, 2014, 23 (08) : 521 - 525
  • [3] Intraocular pressure measurement in mice: a comparison between Goldmann and rebound tonometry
    Kim, C. Y.
    Kuehn, M. H.
    Anderson, M. G.
    Kwon, Y. H.
    EYE, 2007, 21 (09) : 1202 - 1209
  • [4] Intraocular pressure measurement in mice: a comparison between Goldmann and rebound tonometry
    C Y Kim
    M H Kuehn
    M G Anderson
    Y H Kwon
    Eye, 2007, 21 : 1202 - 1209
  • [5] Comparison of Intraocular Pressure Measurements and Assessment of Intraobserver and Interobserver Reproducibility With the Portable ICare Rebound Tonometer and Goldmann Applanation Tonometer in Glaucoma Patients
    Salim, Sarwat
    Du, Haiming
    Wan, Jim
    JOURNAL OF GLAUCOMA, 2013, 22 (04) : 325 - 329
  • [6] Comparison of the Intraocular Pressure Measured Using the New Rebound Tonometer Icare ic100 and Icare TA01i or Goldmann Applanation Tonometer
    Nakakura, Shunsuke
    Mori, Etsuko
    Fujio, Yuki
    Fujisawa, Yasuko
    Matsuya, Kanae
    Kobayashi, Yui
    Tabuchi, Hitoshi
    Asaoka, Ryo
    Kiuchi, Yoshiaki
    JOURNAL OF GLAUCOMA, 2019, 28 (02) : 172 - 177
  • [7] Reproducibility of intraocular pressure self-measurement by ICare Home rebound tonometer and comparison with Goldman applanation tonometer
    Gaillard, R.
    Fenolland, J. R.
    Bonnel, S.
    Rosenberg, R.
    Marill, A. F.
    Bostanci, K.
    Giraud, J. M.
    Renard, J. P.
    ACTA OPHTHALMOLOGICA, 2015, 93
  • [8] Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects
    Guler, Mete
    Bilak, Semsettin
    Bilgin, Burak
    Simsek, Ali
    Capkin, Musa
    Reyhan, Ali Hakim
    JOURNAL OF GLAUCOMA, 2015, 24 (08) : 613 - 618
  • [9] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Yuehua Li
    Jingming Shi
    Xuanchu Duan
    Fang Fan
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1765 - 1770
  • [10] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Li, Yuehua
    Shi, Jingming
    Duan, Xuanchu
    Fan, Fang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (12) : 1765 - 1770