Frequency-Difference Electrical Impedance Imaging of Breast Specimens

被引:0
|
作者
Ouypornkochagorn, Taweechai [1 ]
Ngamdi, Napatsawan [1 ]
Buathong, Piyathida [1 ]
Laor-Iam, Phanpasorn [1 ]
Phisaiphan, Aiyada [1 ]
Ouypornkochagorn, Sairoong [2 ]
Sriwilai, Jaruwan [3 ]
Trongwongsa, Therdkiat [3 ]
机构
[1] Srinakharinwirot Univ, Fac Engn, Bangkok 26120, Thailand
[2] Naresuan Univ, Fac Sci, Phitsanulok 65000, Thailand
[3] Srinakharinwirot Univ, Fac Med, Dept Pathol, Bangkok 26120, Thailand
关键词
Conductivity; Electrodes; Voltage measurement; Probes; Electrical impedance tomography; Pathology; Imaging; Breast; conductivity; electrode probe; frequency-difference electrical impedance tomography (fdEIT); pathological specimens; DIELECTRIC-PROPERTIES; CANCER; TISSUES; SYSTEM; MODELS; WOMEN;
D O I
10.1109/TIM.2024.3418087
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Imaging abnormality in pathological breast specimens can greatly help pathologists select sites for conducting pathological examination through a microscope. In this study, the frequency-difference electrical impedance tomography (fdEIT) technique was proposed for localizing the abnormality in eight breast specimens. Three 16-electrode probes were developed for this purpose which worked with a high-signal-to-noise ratio (SNR) multifrequency electrical impedance tomography (EIT) system. The frequencies of 10 and 50 kHz were selected. The conductivity of the carcinoma masses was found 2.7-3.8 times higher than that of adipose tissues and 2%-21% higher than that of fibroglandular tissues. The conductivity difference of the masses over the selected frequencies was 2.5-3.5 times higher than that of the surrounding tissues. The probes could sense to a depth of similar to 12 mm from the specimen surface. Five of seven specimens (71.4%) were achieved to visualize the masses. Extremely and heterogeneously dense specimens provided better mass distinction in the images than the scattered dense specimens. A mass with a diameter over 20 mm and near the measuring surface was preferable for detection. However, the mass distinction in the images also depends on the conductivity of the mass itself. Mass could be hard and soft, and the soft masses could have different conductivity. The high conductive mass tended to have a better mass distinction in the images.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Frequency-Difference Electrical Impedance Imaging of Cervical Specimens
    Ouypornkochagorn, Taweechai
    Ngamdi, Napatsawan
    Ouypornkochagorn, Sairoong
    Sriwilai, Jaruwan
    Trongwongsa, Therdkiat
    IEEE ACCESS, 2024, 12 : 92087 - 92097
  • [2] Combining frequency-difference and ultrasound modulated electrical impedance tomography
    Harrach, Bastian
    Lee, Eunjung
    Ullrich, Marcel
    INVERSE PROBLEMS, 2015, 31 (09)
  • [3] Multiple Weighted Frequency-difference Method for Electrical Impedance Tomography
    Hu, Qingwei
    Xu, Yanbin
    Liu, Ziqi
    Li, Changbin
    Dong, Feng
    2021 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC 2021), 2021,
  • [4] Method of Tikhonov Regularization for Weighted Frequency-Difference Electrical Impedance Tomography
    Han, Bing
    Xu, Yanbin
    Dong, Feng
    2017 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2017, : 242 - 247
  • [5] Weighted Frequency-difference Electrical Impedance Tomography of Lung Based on RBFNN*
    Bai, Shi-Zhan
    Li, Wen-Sheng
    Lin, Hai-Jun
    Li, Jian-Min
    Zhang, Fu
    Yang, Yu-Xiang
    PROGRESS IN BIOCHEMISTRY AND BIOPHYSICS, 2023, 50 (07) : 1755 - 1766
  • [6] Asymptotic analysis of the membrane structure to sensitivity of frequency-difference electrical impedance tomography
    Kim, Sungwhan
    Lee, Eun Jung
    Woo, Eung Je
    Seo, Jin Keun
    INVERSE PROBLEMS, 2012, 28 (07)
  • [7] Frequency-difference electrical impedance tomography (fdEIT): algorithm development and feasibility study
    Seo, Jin Keun
    Lee, Jeehyun
    Kim, Sung Wan
    Zribi, Habib
    Woo, Eung Je
    PHYSIOLOGICAL MEASUREMENT, 2008, 29 (08) : 929 - 944
  • [8] Factorization Method and Its Physical Justification in Frequency-Difference Electrical Impedance Tomography
    Harrach, Bastian
    Seo, Jin Keun
    Woo, Eung Je
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (11) : 1918 - 1926
  • [9] Calibrated Frequency-Difference Electrical Impedance Tomography for 3D Tissue Culture Monitoring
    Wu, Hancong
    Yang, Yunjie
    Bagnaninchi, Pierre-Olivier
    Jia, Jiabin
    IEEE SENSORS JOURNAL, 2019, 19 (18) : 7813 - 7821
  • [10] Cervical Intraepithelial Neoplasia Localization With Frequency-Difference Electrical Impedance Tomography-Simulation and Phantom Study
    Ouypornkochagorn, Taweechai
    Ngamdi, Napatsawan
    Ouypornkochagorn, Sairoong
    Trongwongsa, Therdkiat
    IEEE ACCESS, 2023, 11 : 67745 - 67754