Frequency-Difference Electrical Impedance Imaging of Breast Specimens

被引:0
|
作者
Ouypornkochagorn, Taweechai [1 ]
Ngamdi, Napatsawan [1 ]
Buathong, Piyathida [1 ]
Laor-Iam, Phanpasorn [1 ]
Phisaiphan, Aiyada [1 ]
Ouypornkochagorn, Sairoong [2 ]
Sriwilai, Jaruwan [3 ]
Trongwongsa, Therdkiat [3 ]
机构
[1] Srinakharinwirot Univ, Fac Engn, Bangkok 26120, Thailand
[2] Naresuan Univ, Fac Sci, Phitsanulok 65000, Thailand
[3] Srinakharinwirot Univ, Fac Med, Dept Pathol, Bangkok 26120, Thailand
关键词
Conductivity; Electrodes; Voltage measurement; Probes; Electrical impedance tomography; Pathology; Imaging; Breast; conductivity; electrode probe; frequency-difference electrical impedance tomography (fdEIT); pathological specimens; DIELECTRIC-PROPERTIES; CANCER; TISSUES; SYSTEM; MODELS; WOMEN;
D O I
10.1109/TIM.2024.3418087
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Imaging abnormality in pathological breast specimens can greatly help pathologists select sites for conducting pathological examination through a microscope. In this study, the frequency-difference electrical impedance tomography (fdEIT) technique was proposed for localizing the abnormality in eight breast specimens. Three 16-electrode probes were developed for this purpose which worked with a high-signal-to-noise ratio (SNR) multifrequency electrical impedance tomography (EIT) system. The frequencies of 10 and 50 kHz were selected. The conductivity of the carcinoma masses was found 2.7-3.8 times higher than that of adipose tissues and 2%-21% higher than that of fibroglandular tissues. The conductivity difference of the masses over the selected frequencies was 2.5-3.5 times higher than that of the surrounding tissues. The probes could sense to a depth of similar to 12 mm from the specimen surface. Five of seven specimens (71.4%) were achieved to visualize the masses. Extremely and heterogeneously dense specimens provided better mass distinction in the images than the scattered dense specimens. A mass with a diameter over 20 mm and near the measuring surface was preferable for detection. However, the mass distinction in the images also depends on the conductivity of the mass itself. Mass could be hard and soft, and the soft masses could have different conductivity. The high conductive mass tended to have a better mass distinction in the images.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Simultaneous Imaging of Bio- and Non-Conductive Targets by Combining Frequency and Time Difference Imaging Methods in Electrical Impedance Tomography
    Bai, Xue
    Liu, Dun
    Wei, Jinzhao
    Bai, Xu
    Sun, Shijie
    Tian, Wenbin
    BIOSENSORS-BASEL, 2021, 11 (06):
  • [42] Dual-frequency HeNe lasers of 3-40MHz frequency-difference
    Li, Y.
    Zhang, S.L.
    Han, Y.M.
    Gaojishu Tongxin/High Technology Letters, 2001, 11 (03):
  • [43] Character analysis and recognition of the imaging in electrical impedance scanning for breast
    Ji Zhenyu
    Dong Xiuzhen
    Liu Ruigang
    Fu Feng
    Shi Xuetao
    You Fusheng
    2007 IEEE/ICME INTERNATIONAL CONFERENCE ON COMPLEX MEDICAL ENGINEERING, VOLS 1-4, 2007, : 828 - 832
  • [44] Multifrequency electrical impedance imaging:: preliminary in vivo experience in breast
    Osterman, KS
    Kerner, TE
    Williams, DB
    Hartov, A
    Poplack, SP
    Paulsen, KD
    PHYSIOLOGICAL MEASUREMENT, 2000, 21 (01) : 99 - 109
  • [45] High-resolution frequency-difference beamforming for a short linear array
    Sheng, Xueli
    Dong, Chaoping
    Guo, Longxiang
    Wang, Xin
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2023, 154 (02): : 1191 - 1201
  • [46] Depth Measurement of Underwater Target Based on Laser Frequency-difference Scanning
    Zheng, Bing
    Li, Wenbo
    Zheng, Haiyong
    Ji, Guangqiang
    Zhao, Lifeng
    2012 OCEANS, 2012,
  • [47] SIZE AND RELIABILITY OF FREQUENCY-DIFFERENCE THRESHOLDS DETERMINED WITH OPERANT TRACKING PROCEDURE
    FRAZIER, LF
    ELLIOTT, DN
    JOURNAL OF THE EXPERIMENTAL ANALYSIS OF BEHAVIOR, 1963, 6 (02) : 189 - &
  • [48] Long-range frequency-difference source localization in the Philippine Sea
    Geroski, David J.
    Dowling, David R.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2019, 146 (06): : 4727 - 4739
  • [49] DETERMINATION OF ABSOLUTE-INTENSITY THRESHOLDS AND FREQUENCY-DIFFERENCE THRESHOLDS IN CATS
    ELLIOTT, DN
    STEIN, L
    HARRISON, MJ
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1960, 32 (03): : 380 - 384
  • [50] A New Strategy to Overcome Cycle Skipping: Frequency-Difference Waveform Inversion
    Ali, Rehman
    Mitcham, Trevor
    Duric, Nebojsa
    MEDICAL IMAGING 2024: ULTRASONIC IMAGING AND TOMOGRAPHY, 2024, 12932