Effectiveness of the Attachment Design and Thickness of Clear Aligners during Orthodontic Anterior Retraction: Finite Element Analysis

被引:7
|
作者
Alhasyimi, Ananto Ali [1 ]
Ayub, Aulia
Farmasyanti, Cendrawasih Andusyana [1 ]
机构
[1] Univ Gadjah Mada, Fac Dent, Dept Orthodont, Yogyakarta 55281, Indonesia
关键词
attachment; clear aligner; force; orthodontic; retraction; thickness;
D O I
10.1055/s-0043-1761452
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective Clear aligner treatment (CAT) provides orthodontic patients with a comfortable treatment alternative; however, this device has limited capacity to facilitate tooth movements. Although composite attachment has been proposed to facilitate tooth displacement, some of its aspects, such as aligner thickness, can influence CAT's precision. This work aimed to compare the stress distribution patterns produced by clear aligners with different thicknesses and composite attachment shapes during anterior retraction. Materials and Methods Maxillary models consisting of clear aligners, maxillary teeth, and various attachments to the upper central incisor's labial surface were generated. Three models were built to mimic the retraction of the upper central incisors. Each had a distinct attachment design (rectangular attachment, ellipsoid attachment, and pyramidal attachment) and various aligner thicknesses (0.75, 0.85, 0.95, 1.05, and 1.15 mm). Upper central incisor retraction was accomplished using clear aligners. Finite element analysis was used to examine the built models. Stress distribution pattern was examined. Results The greater the thickness of the aligner, the higher the stress experienced by the teeth. The 0.75 mm-thick aligner induces the lightest stress with a minimum of 0.0037623 MPa and a maximum of 0.32859 MPa. Meanwhile, the 1.5 mm-thick aligner has the highest stress with a minimum of 0.004679 MPa and a maximum of 0.43858 MPa. The force distribution on rectangular attachments appears evenly distributed. The maximum pressure force on rectangular attachments has a minimum of 0.38828 MPa, which is smaller than the maximum on ellipsoid and pyramidal attachments at 0.40933 and 0.45099 MPa, respectively. Conclusion The best aligner thickness is 0.75 to 0.85 mm for anterior retraction. An aligner with 0.95 mm thickness can still be used when a remarkable amount of tooth movement force is needed; however, this exception is only applicable to a limited number of clear aligner trays. The ellipsoid attachment is the best type of attachment because the resulting force is substantial and evenly distributed.
引用
收藏
页码:174 / 181
页数:8
相关论文
共 50 条
  • [1] Effectiveness of the attachment position in molar intrusion with clear aligners: a finite element study
    Fan, Dian
    Liu, Hao
    Yuan, Chang-Yong
    Wang, Shi-Yu
    Wang, Peng-Lai
    BMC ORAL HEALTH, 2022, 22 (01)
  • [2] Effectiveness of the attachment position in molar intrusion with clear aligners: a finite element study
    Dian Fan
    Hao Liu
    Chang-Yong Yuan
    Shi-Yu Wang
    Peng-Lai Wang
    BMC Oral Health, 22
  • [3] Effectiveness of palatal anchorage device (PAD) for maxillary anterior retraction with clear aligners using finite element study
    Lin, Shuang
    Lu, Tong
    Zhang, Hengbing
    Wang, Wei
    Han, Lei
    Mei, Li
    Li, Huang
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [4] Effects of different intrusion patterns during anterior teeth retraction using clear aligners in extraction cases: an iterative finite element analysis
    Zhang, Yiyan
    Wang, Kaixin
    Li, Mengyu
    Liu, Cuiyu
    Tang, Li
    Wan, Chunyan
    Fan, Cunhui
    Liu, Yang
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12
  • [5] Effectiveness of an anterior mini-screw in achieving incisor intrusion and palatal root torque for anterior retraction with clear aligners: A finite element study
    Liu, Lu
    Zhan, Qi
    Zhou, Jing
    Kuang, Qianyun
    Yan, Xinyu
    Zhang, Xiaoqi
    Shan, Yue
    Li, Xiaolong
    Lai, Wenli
    Long, Hu
    ANGLE ORTHODONTIST, 2021, 91 (06) : 794 - 803
  • [6] Clear aligners for maxillary anterior en masse retraction: a 3D finite element study
    Jiang, Ting
    Wu, Rui Ying
    Wang, Jian Kai
    Wang, Hong Hong
    Tang, Guo Hua
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [7] The effect of enhanced structure in the posterior segment of clear aligners during anterior retraction: a three-dimensional finite element and experimental model analysis
    Jin, Xiaohan
    Tian, Xue
    Lee Zhi Hui, Victoria
    Zheng, Yikan
    Song, Jinlin
    Han, Xianglong
    PROGRESS IN ORTHODONTICS, 2024, 25 (01)
  • [8] Clear aligners for maxillary anterior en masse retraction: a 3D finite element study
    Ting Jiang
    Rui Ying Wu
    Jian Kai Wang
    Hong Hong Wang
    Guo Hua Tang
    Scientific Reports, 10
  • [9] The effect of enhanced structure in the posterior segment of clear aligners during anterior retraction: a three-dimensional finite element and experimental model analysis
    Xiaohan Jin
    Xue Tian
    Victoria Lee Zhi Hui
    Yikan Zheng
    Jinlin Song
    Xianglong Han
    Progress in Orthodontics, 25
  • [10] Effects of canine movement on maxillary anterior en-masse retraction with clear aligners: a finite element study
    Tang, Xin Yue
    Jiang, Ting
    Su, Han
    Chin, Daphne Yuchen
    Chen, Jia Yi
    Qin, Yu Qi
    Tang, Guo Hua
    BMC ORAL HEALTH, 2025, 25 (01):