Longitudinal Stent Deformation Caused by Retraction of the Looped Main Branch Guidewire

被引:0
|
作者
Lee, Hung-Hao [1 ]
Hsu, Po-Chao [1 ,3 ]
Lee, Wen-Hsien [1 ,2 ,3 ]
Chu, Chun-Yuan [1 ]
Su, Ho-Ming [1 ,2 ,3 ]
Lin, Tsung-Hsien [1 ,3 ]
Voon, Wen-Chol [1 ,3 ]
Lai, Wen-Ter [1 ,3 ]
Sheu, Sheng-Hsiung [1 ,3 ]
Chiu, Cheng-An [1 ]
机构
[1] Kaohsiung Med Univ Hosp, Dept Internal Med, Div Cardiol, 100,Tzyou 1st Rd, Kaohsiung 80708, Taiwan
[2] Kaohsiung Municipal Hsiaokang Hosp, Dept Internal Med, Kaohsiung, Taiwan
[3] Kaohsiung Med Univ, Coll Med, Fac Med, Dept Internal Med, Kaohsiung, Taiwan
关键词
Complications; Percutaneous coronary intervention; Stent fracture;
D O I
10.6515/ACS20151115A
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 66-year-old male was treated percutaneously for a bifurcation lesion of the left anterior descending coronary artery by provisional stenting using the jailed wire technique. After successfully stenting the main branch, retraction of the looped main branch guidewire was impossible. After using an intravascular ultrasound we discovered the guidewire was entangled with a stent strut. Thereafter, the proximal stent elongated after retraction. With the support of an over-the-wire microcatheter, we finally pulled out the entrapped guidewire. This rare complication should remind physicians that it is important to prevent the distal guidewire from being looped while retracting it through a stent, regardless of whether it is in the side branch or main vessel. If the guidewire becomes entangled with a stent, a microcatheter or low-profile balloon can be advanced to rescue it before the stent is damaged. Furthermore, the microcather should be maintained after successful retraction of the entangled guidewire to facilitate further wiring and subsequent rescue angioplasty as necessary.
引用
收藏
页码:616 / 618
页数:3
相关论文
共 50 条
  • [41] How serious an adverse event is longitudinal stent deformation? Incidence and severity
    Williams, Paul D.
    Mamas, Mamas A.
    Fraser, Douglas G.
    EXPERT REVIEW OF MEDICAL DEVICES, 2012, 9 (05) : 449 - 451
  • [42] Simulation of longitudinal stent deformation in a patient-specific coronary artery
    Ragkousis, Georgios E.
    Curzen, Nick
    Bressloff, Neil W.
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (04) : 467 - 476
  • [43] Impact of stent platform on longitudinal stent deformation: an in vivo frequency domain optical coherence tomography study
    Yamada R.
    Okura H.
    Kume T.
    Fukuhara K.
    Koyama T.
    Higa T.
    Neishi Y.
    Yoshida K.
    Uemura S.
    Cardiovascular Intervention and Therapeutics, 2017, 32 (3) : 199 - 205
  • [44] Deformation characteristics of main cable in suspension bridge caused by live load
    Chai, Shengbo
    Xiao, Rucheng
    Sun, Bin
    Tongji Daxue Xuebao/Journal of Tongji University, 2012, 40 (10): : 1452 - 1457
  • [45] Retrograde Balloon Dilation outside the Main Branch Stent to Restore the Occlusion of Side Branch in Chronic Total Occlusion Bifurcation Lesions
    Gao Hao-kao
    Li Cheng-xiang
    CASE REPORTS IN CARDIOLOGY, 2020, 2020
  • [46] Clinical and Angiographic Outcomes of Patients Undergoing Entrapped Guidewire Retrieval in Stent-Jailed Side Branch Using a Balloon Catheter
    Sakamoto, Shingo
    Taniguchi, Norimasa
    Mizuguchi, Yukio
    Yamada, Takeshi
    Nakajima, Shunsuke
    Hata, Tetsuya
    Takahashi, Akihiko
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2014, 84 (05) : 750 - 756
  • [47] The Degree of Stent Apposition Measured by Stent Enhancement at the Level of the Side Branch as a Novel Predictor of Procedural Success in Left Main PCI
    Mot, Stefan Dan Cezar
    Serban, Adela Mihaela
    Dadarlat-Pop, Alexandra
    Tomoaia, Raluca
    Pop, Dana
    JOURNAL OF PERSONALIZED MEDICINE, 2023, 13 (05):
  • [48] The dedicated Tryton Side Branch Stent System in PCI of left main lesions; feasibility and results
    Sonck, Jeroen
    Scott, Benjamin
    ACTA CARDIOLOGICA, 2012, 67 (01) : 132 - 132
  • [49] Improved Outcomes of Combined Main Branch Stenting and Side Branch Drug-Coated Balloon versus Two-Stent Strategy in Patients with Left Main Bifurcation Lesions
    Liu, Hengdao
    Tao, Hailong
    Han, Xufei
    Lu, Yang
    Xue, Xiaofei
    Feng, Ruihan
    Lv, Fenghua
    Liu, Yanwei
    Jin, Hongrui
    Li, Lianjie
    Gu, Heping
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2022, 2022
  • [50] Bench test and in vivo evaluation of longitudinal stent deformation during proximal optimisation
    Toth, Gabor G.
    Achim, Alexandru
    Kafka, Marcel
    Wu, Xinlei
    Lunardi, Mattia
    Biswas, Sinjini
    Shahzad, Atif
    Thury, Attila
    Ruzsa, Zoltan
    Johnson, Thomas W.
    Wijns, William
    EUROINTERVENTION, 2022, 18 (01) : 83 - +