Simulated Fewer-Angle SPECT/CT Imaging Protocol for Parathyroid Adenoma

被引:0
|
作者
Ficke, Jennifer B. [1 ]
Watson, Nora L. [2 ]
Stocker, Derek J. [1 ]
Schlegel, Kevin E. [1 ]
Sahajwalla, Maya C. [1 ]
Adams, Chad C. [3 ]
机构
[1] Walter Reed Natl Mil Med Ctr, Dept Radiol, Bethesda, MD 20814 USA
[2] Walter Reed Natl Mil Med Ctr, Dept Res Programs, Bethesda, MD USA
[3] Eisenhower Army Med Ctr, Dept Radiol, Augusta, GA USA
关键词
SPECT/CT; FASpecT; parathyroid adenoma; RECONSTRUCTION ALGORITHMS; LOCALIZATION; SCINTIGRAPHY; QUALITY;
D O I
10.2967/jnmt.122.264649
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new SPECT/CT protocol for parathyroid imaging detailing fewer image -angle acquisitions (fewer -angle SPECT/CT [FASpecT/CT]) was evaluated for identification of parathyroid adenoma. The motivation for validating this protocol was to be able to use it in the future to decrease patient imaging time in our clinic. Methods: This was a retrospective review of existing data performed as a simulated case control study evaluating 50 parathyroid SPECT/CT scans acquired using the standard 60 -stop protocol and the tested 15 -stop FASpecT protocol acquired using angular sampling software. Agreement on the final interpretations between imaging methods was evaluated using the McNemar test and the Cohen K . Interrater reliability among the 3 readers was described for each method using the Fleiss K interpreted as in the strength -ofagreement guidelines by Landis and Koch. Results: Of the 50 evaluated images, 45 (90%) had concordant final image interpretations between imaging methods. The sensitivity of FASpecT/CT relative to SPECT/CT was 17 of 19 (89.5%; 95% CI, 66.9%-98.7%), and the specificity was 28 of 31 (90.3%; 95% CI, 74.2%-98.0%). Additionally, there was statistically significant substantial agreement between protocols and among readers for each protocol. Conclusion: Adequate diagnostic -quality SPECT/CT images can be acquired using significantly fewer imaging stops given advances in camera quality and processing algorithms such as iterative reconstruction. ize surgical localization ble nonpathologic SPECT/CT added diagnosis planar nation ciency, patient 10 delayed clinics delayed lar will compared approximately
引用
收藏
页码:132 / 136
页数:5
相关论文
共 50 条
  • [41] Comparison of dual-tracer subtraction imaging and sestamibi SPECT in preoperative localization of parathyroid adenoma
    Lezaic, L.
    Fettich, J.
    Rep, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S102 - S102
  • [42] Parathyroid adenoma mimicking a cervical isthmic thyroid nodule: Value of SPECT/CT in the localization of ectopic parathyroid. A case report
    El Alaoui, M. Chafai
    Aschawa, H.
    Bellaoui, W.
    Aaouini, R.
    Bssis, A.
    Ben Rais, N.
    MEDECINE NUCLEAIRE-IMAGERIE FONCTIONNELLE ET METABOLIQUE, 2015, 39 : E91 - E94
  • [43] Optimization of myocardial perfusion imaging protocol for digital SPECT/CT imaging system
    Hirvilammi, R.
    Seppanen, M.
    Knuuti, J.
    Noponen, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S219 - S220
  • [44] Validation of resolution recovery CT-attenuation corrected SPECT/CT MIBI parathyroid imaging
    Nichols, Kenneth
    Tronco, Gene
    Jing, Kai
    Heller, Keith
    Sznyter, Laura
    Palestro, Christopher
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [45] Diagnostic Value Of Early SPECT-CT Imaging In The Diagnosis Of Parathyroid Pathologies
    Semirgin, S. Ucak
    Sahin, M.
    Atmaca, B.
    Yapici, O.
    Basoglu, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S288 - S289
  • [46] SPECT-CT using dual-isotope subtraction for parathyroid imaging
    Baudewijns, B.
    Gelderblom, A.
    Habraken, J.
    Karremans, M.
    Kok-Kleinveld, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S458 - S458
  • [47] Comparison of SPECT/CT, Planar Imaging and Conventional Ultrasonography in Localization of Parathyroid Adenomas
    Ince, S.
    Karacalioglu, A. O.
    Emer, O.
    Alagoz, E.
    Arslan, N.
    Gunalp, B.
    Ozguven, M. A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S419 - S419
  • [48] The role of Tc-99m-MIBI SPECT/CT in localization diagnosis of parathyroid adenoma and hyperplasia
    Xu, Lian
    Sun, Xiaoguang
    Song, Shaoli
    Han, Yuan
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [49] The key role of 99mTc-MIBI SPECT/CT in the diagnosis of parathyroid adenoma: a case report
    Li, Qing
    Pan, Jiemin
    Luo, Quanyong
    Wang, Yan
    Bao, Yuqian
    Jia, Weiping
    ARCHIVES OF ENDOCRINOLOGY METABOLISM, 2015, 59 (03): : 265 - 269
  • [50] Atypical non-[99mTc]Tc-MIBI avid parathyroid adenoma with a sternal brown tumor, mimicking ectopic parathyroid adenoma in planar and SPECT images: Importance of SPECT/CT acquisition
    Roustaei, Hessamoddin
    Yaghoubi, Mohammad Ali
    Golestani, Zahra Bakhshi
    Sadrizadeh, Ali
    Sadeghi, Ramin
    Barashki, Somayeh
    Aghaee, Atena
    IRANIAN JOURNAL OF NUCLEAR MEDICINE, 2022, 30 (01): : 76 - 79