A fully automatic parenchyma extraction method for MRI T2*relaxometry of iron loaded liver in transfusion-dependent patients

被引:0
|
作者
Lian, Zifeng [1 ,2 ,3 ,4 ,5 ,6 ]
Lu, Qiqi [1 ,2 ,3 ,4 ,5 ,6 ]
Lin, Bingquan [7 ]
Chen, Lingjian [8 ]
Gong, Jian [1 ]
Hu, Qiugen [1 ,8 ]
Wang, Huafeng [1 ,8 ]
Feng, Yanqiu [1 ,2 ,3 ,4 ,5 ,6 ,8 ]
机构
[1] Southern Med Univ, Sch Biomed Engn, Guangzhou, Peoples R China
[2] Southern Med Univ, Guangdong Prov Key Lab Med Image Proc, Guangzhou, Peoples R China
[3] Southern Med Univ, Guangdong Prov Engn Lab Med Imaging & Diagnost Tec, Guangzhou, Peoples R China
[4] Southern Med Univ, Guangdong Hong Kong Macao Greater Bay Area Ctr Bra, Guangdong Hong Kong Macao Greater Bay Area, Guangzhou, Peoples R China
[5] Southern Med Univ, Key Lab Mental Hlth, Minist Educ, Guangzhou, Peoples R China
[6] Southern Med Univ, Guangdong Hong Kong Joint Lab Psychiat Disorders, Guangzhou, Peoples R China
[7] Southern Med Univ, Nanfang Hosp, Dept Med Imaging Ctr, Guangzhou, Peoples R China
[8] Southern Med Univ, Shunde Hosp, Peoples Hosp Shunde 1, Dept Radiol, Foshan, Peoples R China
基金
中国国家自然科学基金;
关键词
Magnetic resonance imaging; T2*relaxometry; Liver iron loaded; Deep learning; Segmentation; SICKLE-CELL-DISEASE; WHOLE-LIVER; QUANTIFICATION; OVERLOAD; THALASSEMIA; RELAXOMETRY; THERAPY; STORES;
D O I
10.1016/j.mri.2024.02.017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop a fully automatic parenchyma extraction method for the T2* relaxometry of iron overload liver. Methods: A retrospective multicenter collection of liver MR examinations from 177 transfusion-dependent patients was conducted. The proposed method extended a semiautomatic parenchyma extraction algorithm to a fully automatic approach by introducing a modified TransUNet on the R2* (1/T2*) map for liver segmentation. Axial liver slices from 129 patients at 1.5 T were allocated to training (85%) and internal test (15%) sets. Two external test sets separately included 1.5 T data from 20 patients and 3.0 T data from 28 patients. The final T2* measurement was obtained by fitting the average signal of the extracted liver parenchyma. The agreement between T2* measurements using fully and semiautomatic parenchyma extraction methods was assessed using coefficient of variation (CoV) and Bland-Altman plots. Results: Dice of the deep network-based liver segmentation was 0.970 +/- 0.019 on the internal dataset, 0.960 +/- 0.035 on the external 1.5 T dataset, and 0.958 +/- 0.014 on the external 3.0 T dataset. The mean difference bias between T2* measurements of the fully and semiautomatic methods were separately 0.12 (95% CI: -0.37, 0.61) ms, 0.04 (95% CI: -1.0, 1.1) ms, and 0.01 (95% CI: -0.25, 0.23) ms on the three test datasets. The CoVs between the two methods were 4.2%, 4.8% and 2.0% on the internal test set and two external test sets. Conclusions: The developed fully automatic parenchyma extraction approach provides an efficient and operatorindependent T2* measurement for assessing hepatic iron content in clinical practice.
引用
收藏
页码:18 / 26
页数:9
相关论文
共 50 条
  • [31] Serum Ferritin Levels Correlation With Heart and Liver MRI and LIC in Patients With Transfusion-Dependent Thalassemia
    Majd, Zahra
    Haghpanah, Sezaneh
    Ajami, Gholam Hossein
    Matin, Sara
    Namazi, Hamid
    Bardestani, Marzieh
    Karimi, Mehran
    IRANIAN RED CRESCENT MEDICAL JOURNAL, 2015, 17 (04)
  • [32] Correlates of splenic T2*measurements in transfusion-dependent thalassaemia and comparison between non-splenectomised and splenectomised patients
    Meloni, Antonella
    Pistoia, Laura
    Positano, Vincenzo
    Spasiano, Anna
    Putti, Maria Caterina
    Fotzi, Ilaria
    Corigliano, Elisabetta
    Cossu, Antonella
    Massa, Antonella
    Massei, Francesco
    Alberini, Gianna
    Cademartiri, Filippo
    BRITISH JOURNAL OF HAEMATOLOGY, 2023, 203 (03) : E93 - E97
  • [33] Imaging of Body Iron Stores in Transfusion-Dependent Patients By Liver Dual-Energy CT
    Kobayashi, Hironori
    Yoshimura, Norihiko
    Ushiki, Takashi
    Shibasaki, Yasuhiko
    Moriyama, Masato
    Takizawa, Jun
    Narita, Miwako
    Sone, Hirohito
    Masuko, Masayoshi
    BLOOD, 2014, 124 (21)
  • [34] USEFULNESS OF DUAL-ENERGY CT FOR DETECTION OF LIVER IRON DEPOSITION IN TRANSFUSION-DEPENDENT PATIENTS
    Kobayashi, H.
    Yoshimura, N.
    Kasami, T.
    Katagiri, T.
    Nemoto, H.
    Kawamoto, K.
    Ushiki, T.
    Shibasaki, Y.
    Moriyama, M.
    Takizawa, J.
    Narita, M.
    Sone, H.
    Masuko, M.
    HAEMATOLOGICA, 2015, 100 : 789 - 789
  • [35] Serial Measurements of Cardiac and Hepatic Iron with MRI T2*In Transfusion Dependent Patients with Beta-Thalassemia Major Receiving Deferasirox
    Ambati, Srikanth R.
    Kleinert, Dorothy A.
    Giardina, Patricia J.
    BLOOD, 2010, 116 (21) : 851 - 851
  • [36] Utility of Transient Elastography in Estimating Hepatic Iron Concentration in Comparison to MRI in Patients Who Are Transfusion-Dependent
    Ou, George
    Ko, Hin Hin
    Sandhu, Navdeep
    Lee, Terry
    Ezzat, Hatoon
    GASTROENTEROLOGY, 2015, 148 (04) : S230 - S230
  • [37] MRI for the diagnosis of cardiac and liver iron overload in patients with transfusion-dependent thalassemia: An algorithm to guide clinical use when availability is limited
    Viprakasit, Vip
    Ajlan, Amr
    Aydinok, Yesim
    Al Ebadi, Basim A. A.
    Dewedar, Hany
    Ibrahim, Ahmed S.
    Ragab, Lamis
    Trad, Omar
    Wataify, Ahmed S.
    Wong, Lily L. L.
    Taher, Ali T.
    AMERICAN JOURNAL OF HEMATOLOGY, 2018, 93 (06) : E135 - E137
  • [38] Multiple Sclerosis and the Accumulation of Iron in the Basal Ganglia: Quantitative Assessment of Brain Iron Using MRI T2 Relaxometry
    Burgetova, A.
    Seidl, Z.
    Krasensky, J.
    Horakova, D.
    Vaneckova, M.
    EUROPEAN NEUROLOGY, 2010, 63 (03) : 136 - 143
  • [39] T2*relaxometry tool for calibration and quantification of iron concentration based on multi echo MRI data
    Ioannidis, Georgios S.
    Nikiforaki, Katerina
    Kalaitzakis, Georgios
    Boursianis, Themistoklis
    Antonopoulos, Grigorios
    Maris, Thomas G.
    Marias, Kostas
    2022 IEEE INTERNATIONAL CONFERENCE ON IMAGING SYSTEMS AND TECHNIQUES (IST 2022), 2022,
  • [40] Author Correction: Optimisation of T2 and T2* sequences in MRI for better quantification of iron on transfused dependent sickle cell patients
    Azza Ahmed
    Amani Baldo
    A. Sulieman
    Hind Mirghani
    Fouad A. Abolaban
    I. I. Suliman
    Isam Salih
    Scientific Reports, 11