Viscoelastic blood coagulation testing system enabled by a non-contact triboelectric angle sensor

被引:3
|
作者
Wang, Baocheng [1 ,2 ]
Wei, Xuelian [1 ,2 ]
Zhou, Hanlin [1 ]
Cao, Xiaole [1 ,2 ]
Zhang, Enyang [1 ]
Wang, Zhong Lin [1 ,3 ]
Wu, Zhiyi [1 ,2 ]
机构
[1] Chinese Acad Sci, Beijing Inst Nanoenergy & Nanosyst, Beijing 101400, Peoples R China
[2] Univ Chinese Acad Sci, Sch Nanosci & Technol, Beijing, Peoples R China
[3] Georgia Inst Technol, Atlanta, GA USA
来源
EXPLORATION | 2024年 / 4卷 / 01期
基金
中国国家自然科学基金;
关键词
angle sensor; blood coagulation testing; non-contact; thromboelastography; triboelectric nanogenerator; POINT;
D O I
10.1002/EXP.20230073
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Thromboelastography (TEG) remains a convenient and effective viscoelastic blood coagulation testing device for guiding blood component transfusion and assessing the risk of thrombosis. Here, a TEG enabled by a non-contact triboelectric angle sensor (NTAS) with a small size (similar to 7 cm3) is developed for assessing the blood coagulation system. With the assistance of a superelastic torsion wire structure, the NTAS-TEG realizes the detection of blood viscoelasticity. Benefiting from a grating and convex design, the NTAS holds a collection of compelling features, including accurate detection of rotation angles from -2.5 degrees to 2.5 degrees, high linearity (R2 = 0.999), and a resolution of 0.01 degrees. Besides, the NTAS exhibits merits of low cost and simplified fabrication. Based on the NTAS-TEG, a viscoelastic blood coagulation detection and analysis system is successfully constructed, which can provide a graph and parameters associated with clot initiation, formation, and stability for clinicians by using 0.36 mL of whole blood. The system not only validates the feasibility of the triboelectric coagulation testing sensor, but also further expands the application of triboelectric sensors in healthcare. Thromboelastography (TEG) remains a convenient and effective viscoelastic coagulation testing device. This paper presents a novel TEG enabled by a miniaturized triboelectric coagulation sensor, which can realize a comprehensive assessment of the coagulation system and is expected to promote the miniaturization of traditional TEGs. image
引用
收藏
页数:11
相关论文
共 50 条
  • [41] An Innovative Non-Contact ECG and Biopotential Sensor
    Sun, Y.
    Yu, X.
    Berilla, J.
    STRUCTURAL HEALTH MONITORING 2011: CONDITION-BASED MAINTENANCE AND INTELLIGENT STRUCTURES, VOL 2, 2013, : 2203 - 2211
  • [42] A non-contact broad bandwidth current sensor
    Zhang, Rui
    Zhang, Qiaogen
    Geng, Jiangtao
    Liu, Fuli
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2022, 33 (10)
  • [43] NON-CONTACT PNEUMATIC SENSOR BOLT.
    Bechtler, H.J.
    Cachon, R.
    O'Dell, R.
    IBM technical disclosure bulletin, 1985, 27 (08): : 4922 - 4923
  • [44] Micrometric non-contact position magnetoimpedance sensor
    Beato-Lopez, J. J.
    Royo-Silvestre, I
    Gomez-Polo, C.
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2018, 465 : 489 - 494
  • [45] A New Kind of Non-Contact Stress Sensor
    Shi, Yanping
    Qian, Cheng
    ICIEA: 2009 4TH IEEE CONFERENCE ON INDUSTRIAL ELECTRONICS AND APPLICATIONS, VOLS 1-6, 2009, : 3273 - +
  • [46] Simulation model of a non-contact triboelectric nanogenerator based on electrostatic induction
    You, Jing
    Shao, Jiajia
    He, Yahua
    Guo, Xin
    See, K. W.
    Wang, Zhong Lin
    Wang, Xiaolin
    ECOMAT, 2023, 5 (10)
  • [47] System for Non-Contact and Multispectral Examination of Blood Supply to Cutaneous Tissue
    Labuda, Michal
    Smondrk, Maros
    Babusiak, Branko
    Borik, Stefan
    ELECTRONICS, 2022, 11 (18)
  • [48] New method for non-contact material testing
    Adhaesion Kleben und Dichten, 2019, 63 (09): : 38 - 41
  • [49] High throughput non-contact SiP testing
    Moore, B.
    Sellathamby, C.
    Cauvet, P.
    Fleury, H.
    Paulson, M.
    Reja, M.
    Fu, L.
    Bai, B.
    Reid, E.
    Filanovsky, I.
    Slupsky, S.
    2007 IEEE INTERNATIONAL TEST CONFERENCE, VOLS 1 AND 2, 2007, : 328 - +
  • [50] Non-contact Measurement of Rotation Angle with Solo Camera
    Gan Xiaochuan
    Sun Anbin
    Ye Xin
    Ma Liqun
    NINTH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2015, 9446