Viscoelastic blood coagulation testing system enabled by a non-contact triboelectric angle sensor

被引:3
|
作者
Wang, Baocheng [1 ,2 ]
Wei, Xuelian [1 ,2 ]
Zhou, Hanlin [1 ]
Cao, Xiaole [1 ,2 ]
Zhang, Enyang [1 ]
Wang, Zhong Lin [1 ,3 ]
Wu, Zhiyi [1 ,2 ]
机构
[1] Chinese Acad Sci, Beijing Inst Nanoenergy & Nanosyst, Beijing 101400, Peoples R China
[2] Univ Chinese Acad Sci, Sch Nanosci & Technol, Beijing, Peoples R China
[3] Georgia Inst Technol, Atlanta, GA USA
来源
EXPLORATION | 2024年 / 4卷 / 01期
基金
中国国家自然科学基金;
关键词
angle sensor; blood coagulation testing; non-contact; thromboelastography; triboelectric nanogenerator; POINT;
D O I
10.1002/EXP.20230073
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Thromboelastography (TEG) remains a convenient and effective viscoelastic blood coagulation testing device for guiding blood component transfusion and assessing the risk of thrombosis. Here, a TEG enabled by a non-contact triboelectric angle sensor (NTAS) with a small size (similar to 7 cm3) is developed for assessing the blood coagulation system. With the assistance of a superelastic torsion wire structure, the NTAS-TEG realizes the detection of blood viscoelasticity. Benefiting from a grating and convex design, the NTAS holds a collection of compelling features, including accurate detection of rotation angles from -2.5 degrees to 2.5 degrees, high linearity (R2 = 0.999), and a resolution of 0.01 degrees. Besides, the NTAS exhibits merits of low cost and simplified fabrication. Based on the NTAS-TEG, a viscoelastic blood coagulation detection and analysis system is successfully constructed, which can provide a graph and parameters associated with clot initiation, formation, and stability for clinicians by using 0.36 mL of whole blood. The system not only validates the feasibility of the triboelectric coagulation testing sensor, but also further expands the application of triboelectric sensors in healthcare. Thromboelastography (TEG) remains a convenient and effective viscoelastic coagulation testing device. This paper presents a novel TEG enabled by a miniaturized triboelectric coagulation sensor, which can realize a comprehensive assessment of the coagulation system and is expected to promote the miniaturization of traditional TEGs. image
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Non-contact thickness sensor for containers
    Glass, 2000, 77 (05):
  • [22] Non-Contact Nondestructive Rail Testing
    Kudinov, D.
    Shaydurov, G.
    SIBCON-2009: INTERNATIONAL SIBERIAN CONFERENCE ON CONTROL AND COMMUNICATIONS, 2009, : 290 - 295
  • [23] Functionally integrated conductive organohydrogel sensor for wearable motion detection, triboelectric nanogenerator and non-contact sensing
    Zhou, Zixuan
    Yuan, Weizhong
    COMPOSITES PART A-APPLIED SCIENCE AND MANUFACTURING, 2023, 172
  • [24] Hand rehabilitation training system integrating non-contact and contact triboelectric nanogenerators for enhanced gesture and handwriting recognition
    Yang, Lei
    Liang, Jiachang
    Liu, Guilei
    Jia, Youkai
    Yang, Shuai
    Li, Baotong
    Guo, Yanjie
    NANO ENERGY, 2025, 134
  • [25] An opto-electronic integrated non-contact angle measuring system for knife blade angle
    Shi, Wenqing
    Feng, Yujun
    Wang, Wenhua
    Li, Yongqiang
    Huang, Cunyou
    Xiong, Zhengye
    Tan, Zhongming
    Liang, Su
    EQUIPMENT MANUFACTURING TECHNOLOGY AND AUTOMATION, PTS 1-3, 2011, 317-319 : 1016 - +
  • [26] A non-contact flexible pyroelectric sensor for wireless physiological monitoring system
    Jian He
    Sen Li
    Xiaojuan Hou
    Yongjun Zhou
    Hao Li
    Min Cui
    Tao Guo
    Xiangdong Wang
    Jiliang Mu
    Wenping Geng
    Xiujian Chou
    Science China Information Sciences, 2022, 65
  • [27] A non-contact flexible pyroelectric sensor for wireless physiological monitoring system
    Jian HE
    Sen LI
    Xiaojuan HOU
    Yongjun ZHOU
    Hao LI
    Min CUI
    Tao GUO
    Xiangdong WANG
    Jiliang MU
    Wenping GENG
    Xiujian CHOU
    ScienceChina(InformationSciences), 2022, 65 (02) : 162 - 172
  • [28] A non-contact flexible pyroelectric sensor for wireless physiological monitoring system
    He, Jian
    Li, Sen
    Hou, Xiaojuan
    Zhou, Yongjun
    Li, Hao
    Cui, Min
    Guo, Tao
    Wang, Xiangdong
    Mu, Jiliang
    Geng, Wenping
    Chou, Xiujian
    SCIENCE CHINA-INFORMATION SCIENCES, 2022, 65 (02)
  • [29] Non-contact respiration monitoring system using fiber grating sensor
    Sato, I
    Nakajima, M
    MACHINE VISION AND ITS OPTOMECHATRONIC APPLICATIONS, 2004, 5603 : 262 - 271
  • [30] Non-contact transport system
    Morishita, M
    JOURNAL OF JAPANESE SOCIETY OF TRIBOLOGISTS, 1997, 42 (11) : 859 - 864