Study of dead time estimation method based on pulse interval distribution

被引:0
|
作者
Liu, Junlong [1 ]
Hong, Xu [1 ]
Huang, Haixi [1 ]
Song, Xinru [1 ]
Li, Lin [1 ]
Zhou, Chengzhuo [1 ]
机构
[1] Chengdu Univ Technol, Coll Nucl Technol & Automat Engn, Chengdu 610059, Peoples R China
基金
中国国家自然科学基金;
关键词
Dead time; Pulse interval distribution; True input counting rate;
D O I
10.1016/j.apradiso.2024.111321
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In nuclear radiation detection, some measured radiation counts are lost due to dead time. Estimation of dead time is necessary to restore the counting information in quantitative analysis. This study aims to propose a method for estimating dead time at high counting rate. First of all, a measurement system of pulse interval distribution was used to estimate the true input counting rate of the detection system and implemented in FPGA. Then, a digital pole-zero cancellation technique was introduced in the measurement system to decrease the decay time constant of pulse. This was aimed at mitigating the impact of pile-up on the pulse interval distribution spectrum, especially at high counting rate. The feasibility of this measurement system at high counting rate was verified by the experimental platform of EDXRF. The dead time was calculated using the measured counting rate and the estimated true input counting rate. The accuracy of the dead time relies on the precision of the estimated true count rate. When the counting rate reaches 600 kCPS, the relative error between the theoretical counting rate and the estimated true input counting rate is less than 5%.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] A Gesture Recognition Method Based on Interval Distribution Probability
    Zhou, Song-Bin
    Lu, Shan-Dan
    Li, Chang
    2015 International Conference on Software Engineering and Information System (SEIS 2015), 2015, : 731 - 737
  • [42] Estimation of signal arrival time based on UWB pulse
    Meng, Hongwei
    Wang, Shuxun
    Sun, Xiaoying
    WCICA 2006: SIXTH WORLD CONGRESS ON INTELLIGENT CONTROL AND AUTOMATION, VOLS 1-12, CONFERENCE PROCEEDINGS, 2006, : 1759 - 1762
  • [43] Load Position Estimation Method for Wearable Devices Based on Difference in Pulse Wave Arrival Time
    Yoshida, Kazuki
    Murao, Kazuya
    SENSORS, 2022, 22 (03)
  • [44] Pulse repetition interval estimation in moving passive sensors based on observation calibration
    Ye, H.H.
    Liu, Z.
    Jiang, W.L.
    Progress In Electromagnetics Research C, 2012, 28 : 257 - 270
  • [45] A 9 ps time interval digitizer based on pulse repetition and averaging
    Szplet, Ryszard
    Jarzynski, Slawomir
    2019 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2019, : 12 - 16
  • [46] An attenuated time measurement based on pulse interval for oscillating cup viscometer
    Zhu, Ping
    Lai, Jiahui
    Wu, Kesen
    Zhang, Zhao
    Huang, Xingcan
    Zhang, Lili
    Liu, Jian
    2017 CHINESE AUTOMATION CONGRESS (CAC), 2017, : 3109 - 3111
  • [47] Unknown weak light pulse measurement using time interval distribution photon counting
    Zhu, Xinyi
    Wang, Yurong
    Li, Xue
    Li, Zhaohui
    Pan, Haifeng
    Chen, Xiuliang
    Zhang, Pengcheng
    Huang, Kun
    Wu, Guang
    OPTICS LETTERS, 2025, 50 (04) : 1353 - 1356
  • [48] A likelihood-based method for real-time estimation of the serial interval and reproductive number of an epidemic
    White, L. Forsberg
    Pagano, M.
    STATISTICS IN MEDICINE, 2008, 27 (16) : 2999 - 3016
  • [49] Interval estimation for the scale parameter of the two-parameter exponential distribution based on time-censored data
    Thiagarajah, K
    Paul, SR
    JOURNAL OF STATISTICAL PLANNING AND INFERENCE, 1997, 59 (02) : 279 - 289
  • [50] A General Dead Time Compensation Method Based on Current Ripple Prediction and Pulse Delay Measurement for Voltage Source Inverter
    Shen, Zewei
    Jiang, Dong
    Chen, Jianan
    Liu, Zicheng
    2020 IEEE ENERGY CONVERSION CONGRESS AND EXPOSITION (ECCE), 2020, : 1927 - 1931