Study of dead time estimation method based on pulse interval distribution

被引:0
|
作者
Liu, Junlong [1 ]
Hong, Xu [1 ]
Huang, Haixi [1 ]
Song, Xinru [1 ]
Li, Lin [1 ]
Zhou, Chengzhuo [1 ]
机构
[1] Chengdu Univ Technol, Coll Nucl Technol & Automat Engn, Chengdu 610059, Peoples R China
基金
中国国家自然科学基金;
关键词
Dead time; Pulse interval distribution; True input counting rate;
D O I
10.1016/j.apradiso.2024.111321
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In nuclear radiation detection, some measured radiation counts are lost due to dead time. Estimation of dead time is necessary to restore the counting information in quantitative analysis. This study aims to propose a method for estimating dead time at high counting rate. First of all, a measurement system of pulse interval distribution was used to estimate the true input counting rate of the detection system and implemented in FPGA. Then, a digital pole-zero cancellation technique was introduced in the measurement system to decrease the decay time constant of pulse. This was aimed at mitigating the impact of pile-up on the pulse interval distribution spectrum, especially at high counting rate. The feasibility of this measurement system at high counting rate was verified by the experimental platform of EDXRF. The dead time was calculated using the measured counting rate and the estimated true input counting rate. The accuracy of the dead time relies on the precision of the estimated true count rate. When the counting rate reaches 600 kCPS, the relative error between the theoretical counting rate and the estimated true input counting rate is less than 5%.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] APPROXIMATION OF THE NONPARALYZABLE DEAD TIME BY THE SHORTEST TIME INTERVAL
    SABOL, J
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (09): : 2086 - 2087
  • [22] Impact of heart disease and calibration interval on accuracy of pulse transit time-based blood pressure estimation
    Ding, Xiaorong
    Zhang, Yuanting
    Tsang, Hon Ki
    PHYSIOLOGICAL MEASUREMENT, 2016, 37 (02) : 227 - 237
  • [23] A novel method for non-invasive blood pressure estimation based on continuous pulse transit time: An observational study
    Shin, Hangsik
    PSYCHOPHYSIOLOGY, 2023, 60 (02)
  • [24] Estimation of the incubation time distribution in the singly and doubly interval censored model
    Groeneboom, Piet
    STATISTICA NEERLANDICA, 2024, 78 (04) : 617 - 635
  • [25] PULSE ARRIVAL TIME-ESTIMATION BASED ON PULSE SAMPLE RATIOS
    HO, KC
    CHAN, YT
    INKOL, RJ
    IEE PROCEEDINGS-RADAR SONAR AND NAVIGATION, 1995, 142 (04) : 153 - 157
  • [26] Recognition Method of Abnormal Data Based on Interval Estimation
    Gu, Jingqiu
    Wang, Zhihai
    Wu, Huarui
    MATERIALS, INFORMATION, MECHANICAL, ELECTRONIC AND COMPUTER ENGINEERING (MIMECE 2016), 2016, : 212 - 218
  • [27] Study on the parameters estimation based on interval observer
    Wang, Zhe
    Li, Zetao
    Dahhou, Boutaieb
    Proceedings of the 2012 24th Chinese Control and Decision Conference, CCDC 2012, 2012, : 1630 - 1632
  • [28] Study on the Parameters Estimation Based on Interval Observe
    Wang, Zhe
    Li, Zetao
    Dahhou, Boutaieb
    PROCEEDINGS OF THE 2012 24TH CHINESE CONTROL AND DECISION CONFERENCE (CCDC), 2012, : 1630 - 1632
  • [29] Revised constraint-propagation method for distribution interval state estimation
    Ngo, VietCuong
    Wu, Wenchuan
    Lou, Yining
    IET GENERATION TRANSMISSION & DISTRIBUTION, 2020, 14 (07) : 1329 - 1336
  • [30] A Decentralized Stay-Time Based Occupant Distribution Estimation Method for Buildings
    Jia, Qing-Shan
    Wang, Hengtao
    Lei, Yulin
    Zhao, Qianchuan
    Guan, Xiaohong
    IEEE TRANSACTIONS ON AUTOMATION SCIENCE AND ENGINEERING, 2015, 12 (04) : 1482 - 1491