In silico study of the electronic portal imaging devices sensitivity to parotid glands shrinkage during radiotherapy for nasopharyngeal cases

被引:0
|
作者
Alzahrani H.M. [1 ]
机构
[1] Radiology Technology, College of Applied Medical Sciences, Taibah University, Medina
关键词
Anatomical changes; Dosimetry; EGSnrc; EPID; Monte Carlo simulation; Parotid gland shrinkage;
D O I
10.1007/s43994-023-00103-z
中图分类号
学科分类号
摘要
This computational investigation examines the capability of electronic portal imaging devices (EPIDs) in detecting anatomical changes during radiation therapy. A photon beam for 6 MV was simulated using compliant phase-space files via BEAMnrc (an EGSnrc user code). Initially, the dose calculation was carried out using DOSXYZnrc, an EGSnrc user code, on a water-based calibration phantom made of water. Phantoms, constructed utilizing patient computed tomography data, were subsequently used to simulate clinically relevant parotid gland (PG) shrinkage through image deformation using ImSimQA software. Two EPID geometries were implemented; a simple water slab, and a comprehensive multi-layer representation of the EPID (referred to as the 17-slab model). Using a 1%/1 mm gamma index the 17-slab geometry was sensitive to volume reductions in the PG ≥ − 26% from its original volume, while the water slab detected volume reduction of ≥ − 28.5%. In a clinical setting, replanning of head and neck patients is initiated when a reduction of 24–30% in PG volume is detected through cone-beam computed tomography. The sensitivity index, indicative of the signal to error ratio between the reference and evaluated images, showed an increase as the gland volume decreased for both models. Notably, the 17-slab model consistently more sensitive than the water slab. This computational study highlights the prospective use of EPIDs in monitoring and detecting clinically significant volume changes in the PG during radiation treatment. Advances in knowledge: While EPIDs are frequently employed for patient setup and alignment, this computational work constructed a virtual EPID to assess its detection abilities for the anatomical changes in the PG due to radiation exposure in head and neck cancer cases. © The Author(s) 2024.
引用
收藏
页码:279 / 289
页数:10
相关论文
共 38 条
  • [31] Electronic portal imaging device detection of radioopaque markers for the evaluation of prostate position during megavoltage irradiation: A clinical study
    Vigneault, E
    Pouliot, J
    Laverdiere, J
    Roy, J
    Dorion, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (01): : 205 - 212
  • [32] Electronic portal cine imaging of implanted marker seeds during respiratory gated treatment to evaluate interfraction movement for stereotactic body radiotherapy of liver metastases
    Serago, C. F.
    Vallow, L. A.
    Paz-Fumagalli, R.
    Kim, S.
    Gale, A. A.
    Magalhaes, W. L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S543 - S544
  • [33] Planning target volume assessment in lung tumors during 3D conformal radiotherapy by means of an aSi electronic portal imaging device in cine mode
    R. Caivano
    S. Clemente
    A. Fiorentino
    C. Chiumento
    P. Pedicini
    G. Califano
    M. Cozzolino
    V. Fusco
    Clinical and Translational Oncology, 2013, 15 : 638 - 642
  • [34] Planning target volume assessment in lung tumors during 3D conformal radiotherapy by means of an aSi electronic portal imaging device in cine mode
    Caivano, R.
    Clemente, S.
    Fiorentino, A.
    Chiumento, C.
    Pedicini, P.
    Califano, G.
    Cozzolino, M.
    Fusco, V.
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2013, 15 (08): : 638 - 642
  • [35] A Prospective Weekly Volumetric Magnetic Resonance Imaging (MRI) Study of Parotid Glands and Neck Nodes in Oropharyngeal Cancer (OPC) Patients Treated On a Phase II Trial of Adaptive Radiotherapy (ART)
    Polvorosa, C.
    Hu, Y.
    Hunt, M.
    Mechalakos, J.
    Zhang, P.
    Fontenla, S.
    Shi, C.
    Li, X.
    Allgood, N.
    McBride, S.
    Riaz, N.
    Lee, N.
    Tsai, C. J.
    MEDICAL PHYSICS, 2018, 45 (06) : E495 - E495
  • [36] Comparison of low-dose, half-rotation, cone-beam CT with electronic portal imaging device for registration of fiducial markers during prostate radiotherapy
    Ung, Ngie Min
    Wee, Leonard
    Hackett, Sara Lyons
    Jones, Andrew
    Lim, Tee Sin
    Harper, Christopher Stirling
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (04): : 171 - 181
  • [37] Prospective study on set-up random errors in 3D-conformal prostate cancer radiotherapy by using an electronic portal imaging device (EPID)
    Zerini, D
    Antognoni, P
    Gandini, S
    Monti, L
    Visconti, M
    Orecchia, R
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S288 - S288
  • [38] A Study to Assess the Dosimetric Impact of the Anatomical Changes Occurring in the Parotid Glands and Tumour Volume during Intensity Modulated Radiotherapy using Simultaneous Integrated Boost (IMRT-SIB) in Head and Neck Squamous Cell Cancers
    Ghosh, Arunima
    Gupta, Seema
    Johny, Danial
    Vidyadhar Bhosale, Vivek
    Pal Singh Negi, Mahendra
    CANCER MEDICINE, 2021, 10 (15): : 5175 - 5190