Dual threshold input receiver FPGA-only signal digitization method for time-of-flight positron emission tomography

被引:3
|
作者
Ko, Guen Bae [1 ,2 ]
Lee, Jae Sung [1 ,2 ,3 ,4 ]
机构
[1] Brightonix Imaging Inc, Seoul 04782, South Korea
[2] Seoul Natl Univ, Inst Radiat Med, Med Res Ctr, Seoul 03080, South Korea
[3] Seoul Natl Univ, Coll Med, Dept Nucl Med, Seoul 03080, South Korea
[4] Seoul Natl Univ, Coll Engn, Interdisciplinary Program Bioengn, Seoul 03080, South Korea
关键词
Positron emission tomography; Field-programmable gate array; Silicon photomultiplier; Time-to-digital converter; Time-over-threshold; PET; PET/MRI;
D O I
10.1007/s13534-024-00380-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As silicon photomultiplier (SiPM)-based time-of-flight (TOF) positron emission tomography (PET) becomes popular, the need for sophisticated PET data acquisition (DAQ) systems is increasing. One promising solution to this challenge is the adoption of a field-programmable gate array (FPGA)-only signal digitization method. In this paper, we propose a new approach to efficiently implement an FPGA-only digitizer. We configured the input/output (IO) port of the FPGA to function as a dual-threshold voltage comparator through the use of simple passive circuitry and heterogeneous IO standards. This configuration overcomes the limitations of existing methods by allowing different threshold voltages for adjacent IO pins, effectively reducing routing complexity and lowering manufacturing costs. An FPGA-only digitizer was implemented by integrating the dual-threshold voltage comparator and FPGA-based time-to-digital converter. By combining the dual-threshold time-over-threshold (TOT) method and curve fitting, precise energy information could be obtained. The performance of the FPGA-only digitizer was assessed using a detector setup comprising a 3 x 3 x 20 mm3 LYSO scintillation crystal and a single pixel SiPM. Using the configured evaluation setup, an energy resolution of 12.5% and a time resolution of 146 +/- 9 ps were achieved for a 20 mm scintillation crystal. The dual-threshold TOT implemented using the proposed method showed consistent linearity across an energy range of 100 keV to 600 keV. The proposed method is well-suited for the development of cost-effective DAQ systems in highly integrated TOF PET systems.
引用
收藏
页码:847 / 858
页数:12
相关论文
共 50 条
  • [21] Event-Time Determination by Waveform Analysis for Time-of-Flight Positron Emission Tomography
    Kao, Chien-Min
    Kim, Heejong
    Chen, Chin-Tu
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3874 - 3879
  • [22] Single Channel Optimization for an Endoscopic Time-of-Flight Positron Emission Tomography Detector
    Garutti, Erika
    Gadow, Karsten
    Goettlich, Martin
    Silenzi, Alessandro
    Xu, Chen
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 54 - 58
  • [23] Poisson-noise weighted filter for time-of-flight positron emission tomography
    Gengsheng L. Zeng
    Li Lv
    Qiu Huang
    Visual Computing for Industry, Biomedicine, and Art, 3
  • [24] Image quality improvements with time-of-flight positron emission tomography for molecular imaging
    Manjeshwar, RM
    Shao, YP
    Jansen, FP
    2005 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH, AND SIGNAL PROCESSING, VOLS 1-5: SPEECH PROCESSING, 2005, : 853 - 856
  • [25] Poisson-noise weighted filter for time-of-flight positron emission tomography
    Zeng, Gengsheng L.
    Lv, Li
    Huang, Qiu
    VISUAL COMPUTING FOR INDUSTRY BIOMEDICINE AND ART, 2020, 3 (01)
  • [26] HRFlexToT: A High Dynamic Range ASIC for Time-of-Flight Positron Emission Tomography
    Sanchez, David
    Gomez, Sergio
    Mauricio, Joan
    Freixas, Lluis
    Sanuy, Andreu
    Guixe, Gerard
    Lopez, Albert
    Manera, Rafel
    Marin, Jesus
    Perez, Jose M.
    Picatoste, Eduardo
    Pujol, Carolina
    Sanmukh, Anand
    Rato, Pedro
    Vela, Oscar
    Gascon, David
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2022, 6 (01) : 51 - 67
  • [27] Pilot study of timing performance for time-of-flight positron emission tomography detector
    Liu, Jun-Hui
    Li, Dao-Wu
    Zhang, Zhi-Ming
    Liao, Yan-Fei
    Huang, Xian-Chao
    Wang, Pei-Lin
    Chen, Yan
    Yun, Ming-Kai
    Feng, Bao-Tong
    Shuai, Lei
    Wang, Bao-Yi
    Zhang, Tian-Bao
    Wei, Long
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2011, 45 (12): : 1522 - 1526
  • [28] TIME-OF-FLIGHT POSITRON-EMISSION TOMOGRAPHY STATUS RELATIVE TO CONVENTIONAL PET
    TERPOGOSSIAN, MM
    FICKE, DC
    JOURNAL OF NUCLEAR MEDICINE, 1984, 25 (01) : 136 - 137
  • [29] Sensitivity estimation in time-of-flight list-mode positron emission tomography
    Herraiz, J. L.
    Sitek, A.
    MEDICAL PHYSICS, 2015, 42 (11) : 6690 - 6702
  • [30] A New Direction to Explore to Advance Coincidence Time Resolution for Time-of-Flight Positron Emission Tomography
    Tao, L.
    Levin, C. S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S305 - S306