5-YEAR FOLLOW-UP EVALUATION OF THE NONCEMENTED PRESS-FIT TITANIUM HIP-JOINT ENDOPROSTHESIS

被引:0
|
作者
SERAL, F
VILLAR, JM
ESTELLER, A
VIVAR, FG
ABAD, I
GRANDE, MM
JORDA, E
ESPINAR, E
机构
[1] HOSP CRUCES, VIZCAYA, SPAIN
[2] HOSP MARQUES VALDECILLA, SANTANDER, SPAIN
[3] HOSP REG CARLOS HAYA, MALAGA, SPAIN
[4] HOSP VIRGEN CAMINO, PAMPLONA, SPAIN
[5] HOSP GEN VIRGEN LLUCH, PALMA DE MALLORCA, SPAIN
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
ties with a titanium alloy stem and fixation by the Zweymuller press-fit and an Endler polyethylene threaded cup was reviewed in detail. The minimum follow-up period was 48 months and the maximum 72 months, with an average of 60 months. A scale from zero to five points was applied to evaluate pain, mobility, and motion for a total possible accumulation of 15 points. The global results of the different ethiologic groups (arthrosis, femoral head necrosis, rheumatoid arthritis, and subcapital hip fractures) have been very good and good (12-15 points) in 67.5% of the cases and fairly good and bad in 32.3%. These results have been better in femoral head necrosis than in arthrosis or rheumatoid arthritis, but not as good in subcapital hip fractures. The age groups below 60 had better results than the above 60 groups. The Singh index higher than 3 was correlated with better-than-average results. The polyethylene cup migrated horizontally (more than 4 mm) in 7.6% of the cases and vertically (more than 5 mm) in 10%. The nonevolutionary cortical remodelation of the femur does not influence the results. Prosthetic stem sinking less than 4 mm has been found in 62% of the cases, from 4 to 9 mm in 21%, and greater than 9 mm in 6%. No alterations with clinical consequences attributable to stress-shielding have been detected.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 50 条
  • [41] Ceramic-on-ceramic articulation in press-fit total hip arthroplasty as a potential reason for early failure, what about the survivors: a ten year follow-up
    J. van Loon
    A. M. J. S. Vervest
    H. M. van der Vis
    I. N. Sierevelt
    D. C. Baas
    K. T. M. Opdam
    G. M. M. J. Kerkhoffs
    D. Haverkamp
    International Orthopaedics, 2021, 45 : 1447 - 1454
  • [42] Ceramic-on-ceramic articulation in press-fit total hip arthroplasty as a potential reason for early failure, what about the survivors: a ten year follow-up
    van Loon, J.
    Vervest, A. M. J. S.
    van der Vis, H. M.
    Sierevelt, I. N.
    Baas, D. C.
    Opdam, K. T. M.
    Kerkhoffs, G. M. M. J.
    Haverkamp, D.
    INTERNATIONAL ORTHOPAEDICS, 2021, 45 (06) : 1447 - 1454
  • [43] Uncemented press-fit total hip arthroplasty using the identifit custom-molding technique - A prospective minimum 2-year follow-up study
    Robinson, RP
    Clark, JE
    JOURNAL OF ARTHROPLASTY, 1996, 11 (03): : 247 - 254
  • [44] Early follow-up of a custom non-fluted diaphyseal press-fit tumour prosthesis
    O'Donnell, Patrick W.
    Griffin, Anthony M.
    Eward, William C.
    Sternheim, Amir
    Wunder, Jay S.
    Ferguson, Peter C.
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (01) : 123 - 127
  • [45] LONG-TERM FOLLOW-UP OF INTERTROCHANTERIC OSTEOTOMY IN OSTEOARTHRITIS IN HIP-JOINT
    GOLDIE, IF
    ANDERSSON, GB
    OLSSON, SS
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1973, (93) : 265 - 270
  • [46] Radiographic evaluation of short-stem press-fit total shoulder arthroplasty: short-term follow-up
    Casagrande, Danielle J.
    Parks, Di L.
    Torngren, Travis
    Schrumpf, Mark A.
    Harmsen, Samuel M.
    Norris, Tom R.
    Kelly, James D., II
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2016, 25 (07) : 1163 - 1169
  • [47] CONGENITAL DISLOCATION OF THE HIP-JOINT - FOLLOW-UP OF LATE-DIAGNOSIS CASES
    PALMEN, K
    ACTA ORTHOPAEDICA SCANDINAVICA, 1983, 54 (03): : 497 - 497
  • [48] WALKING PATTERNS IN 20 YEARS FOLLOW-UP HIP-JOINT ARTHRODESIS PATIENTS
    CATANI, F
    GIANNINI, S
    BOMBARDI, F
    MARCACCI, M
    CECCARELLI, F
    PROCEEDINGS OF THE ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, PTS 1-4, 1988, : 620 - 621
  • [49] A prospective 24 months follow-up of a three component press-fit prosthesis for hallux rigidus
    Wassink, S.
    Burger, B. J.
    Saragas, N. P.
    Marquez, J. Asuncion
    Trtik, L.
    Harlaar, J.
    FOOT AND ANKLE SURGERY, 2017, 23 (03) : 157 - 162
  • [50] Early follow-up of a custom non-fluted diaphyseal press-fit tumour prosthesis
    Patrick W. O’Donnell
    Anthony M. Griffin
    William C. Eward
    Amir Sternheim
    Jay S. Wunder
    Peter C. Ferguson
    International Orthopaedics, 2014, 38 : 123 - 127