APPLICATION OF A HIGH-RESOLUTION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER TO THE MEASUREMENT OF LONG-LIVED RADIONUCLIDES

被引:81
|
作者
KIM, CK
SEKI, R
MORITA, S
YAMASAKI, S
TSUMURA, A
TAKAKU, Y
IGARASHI, Y
YAMAMOTO, M
机构
[1] NATL INST RADIOL SCI,DIV RADIOECOL,3609 ISOZAKI,IBARAKI 31112,JAPAN
[2] UNIV TSUKUBA,DEPT CHEM,SAKURA,IBARAKI 305,JAPAN
[3] NATL INST AGROENVIRONM SCI,DIV WATER QUAL SCI,IBARAKI 305,JAPAN
[4] MARUBUN,CHUO KU,TOKYO 170,JAPAN
[5] KANAZAWA UNIV,LOW LEVEL RADIOACT LAB,TATSUNOKUCHI,ISHIKAWA 92312,JAPAN
关键词
HIGH RESOLUTION INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; LONG-LIVED RADIONUCLIDE; DETECTION LIMIT;
D O I
10.1039/ja9910600205
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Some long-lived radionuclides, such as Tc-99, Ra-226, Th-232, Np-237, U-238, Pu-239 and Pu-240 were measured using high resolution inductively coupled plasma mass spectrometry (HR-ICP-MS). By using HR-ICP-MS with an ultrasonic nebulizer, the detection limits of these nuclides were 0.002-0.02 pg ml-1 and the sensitivities were 10 times better than those obtained using HR-ICP-MS without the ultrasonic nebulizer. More accurate isotopic data were also obtained using HR-ICP-MS than with quadrupole ICP-MS at lower concentrations of the analyte because of the improvement in counting statistics that can be obtained with HR-ICP-MS due to the greater efficiency of ion transmission. A comparison of the measurement of the Pu-240 to Pu-239 ratio is shown.
引用
收藏
页码:205 / 209
页数:5
相关论文
共 50 条
  • [21] Selenium measurement in sulphides by hydride generation high-resolution inductively coupled plasma mass spectrometry
    Fitzpatrick, Alexander J.
    Kyser, T. Kurtis
    Chipley, Don
    GEOCHEMISTRY-EXPLORATION ENVIRONMENT ANALYSIS, 2009, 9 : 93 - 100
  • [22] HIGH-RESOLUTION TIME-OF-FLIGHT MASS-SPECTROMETER
    BERGMANN, T
    MARTIN, TP
    SCHABER, H
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (04): : 792 - 793
  • [23] Iron Isotopic Measurement using Large-Geometry High-Resolution Multi-Collector Inductively Coupled Plasma Mass Spectrometer
    Lei, Yvtian
    Li, Ming
    Wang, Zaicong
    Zhu, Yangtao
    Hu, Zhaochu
    Li, Yongsheng
    Chai, Xinna
    ATOMIC SPECTROSCOPY, 2022, 43 (03) : 214 - 222
  • [24] Miniature, high-resolution, quadrupole mass-spectrometer array
    Orient, OJ
    Chutjian, A
    Garkanian, V
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (03): : 1393 - 1397
  • [25] HIGH-RESOLUTION ON A QUADRUPOLE ION TRAP MASS-SPECTROMETER
    SCHWARTZ, JC
    SYKA, JEP
    JARDINE, I
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1991, 2 (03) : 198 - 204
  • [26] Ultratrace and isotopic analysis of long-lived radionuclides and depleted uranium by direct liquid introduction-inductively coupled plasma mass spectrometry.
    Montaser, A
    McLean, JA
    Becker, JS
    Dietze, HJ
    Westphal, CS
    Hodge, SJ
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2001, 221 : U90 - U90
  • [27] Inductively coupled plasma mass spectrometry (ICP-MS) and laser ablation ICP-MS for isotope analysis of long-lived radionuclides
    Becker, JS
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2005, 242 (2-3) : 183 - 195
  • [28] Oxide ion formation of long-lived radionuclides in double-focusing sector field inductively coupled plasma mass spectrometry and their analytical applications
    Becker, JS
    Dietze, HJ
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 202 (1-3) : 69 - 79
  • [29] PROTOTYPE INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER FOR QUANTITATIVE ELEMENTAL AND ISOTOPIC ANALYSIS
    GORKOVOI, VV
    ELOKHIN, VA
    CHERNETSKII, SM
    CHOPOROV, DY
    JOURNAL OF ANALYTICAL CHEMISTRY OF THE USSR, 1991, 46 (02): : 313 - 317
  • [30] ANALYTICAL PERFORMANCE EVALUATION OF A 40.68 MHZ INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER
    PARK, CJ
    LEE, KW
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (06) : 431 - 437