Continuous non-invasive blood pressure during continuous repositioning by pulse transit time

被引:1
|
作者
Jeong, Hieyong [1 ]
Yoshimoto, Kayo [2 ]
Wang, Tianyi [3 ]
Ohno, Takafumi [1 ]
Yamada, Kenji [1 ]
Kido, Michiko [3 ]
Ohno, Yuko [3 ]
机构
[1] Osaka Univ, Grad Sch Med, Dept R&D Innovat Healthcare, 1-7 Yamada Oka, Suita, Osaka 5650871, Japan
[2] Osaka City Univ, Grad Sch Engn, Dept Elect & Informat Engn, Osaka, Japan
[3] Osaka Univ, Grad Sch Med, Div Hlth Sci, Suita, Osaka, Japan
来源
COGENT ENGINEERING | 2016年 / 3卷 / 01期
关键词
continuous non-invasive blood pressure; pulse transit time; pulse wave velocity; severely handicapped children; side effects;
D O I
10.1080/23311916.2016.1221202
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The purpose of the present study was to propose a method to measure continuous non-invasive arterial pressure (CNAP) of severely handicapped patient during continuous repositioning. We have been interested in healthcare for severely handicapped children. Although their long stay in hospital could be led to a circulatory disorder or a serious secondary disease, there was little way to give an adequate healthcare service for them due to several kinds of side effects. Thus, we developed the non-invasive sensor system to measure pulse transit time (PTT) through the different output signal between the two different measuring points. The proposed extended model of algorithm enabled us to measure continuous blood pressure (BP) with the measured PTT. The algorithm considered the influence of gravity on BP; there was no need to re-tune the system even if the posture of patient was changed. We performed the continuous repositioning experiment from the supine posture to 45 or 75 degrees postures after the effectiveness of the proposed algorithm was confirmed. Through experimental results, it was found that the proposed method has the strongest correlation with commercial BP device. The experimental results showed that it was possible for the proposed algorithm to measure CNAP during continuous repositioning without the re-tune of system. It was recommended that some points of developed prototype hardware were necessary to be improved and the integration of other bio signals was more helpful for providing important insight into the overall physical condition.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Methods for Continuous Non-invasive Measurement of Blood Pressure: Pulse Transit Time a Review
    Kaur, Harleen
    Bhutto, Javed Khan
    Choudhary, Bhawani Shankar
    ASIAN JOURNAL OF PHARMACEUTICS, 2018, 12 (03) : S863 - S872
  • [2] Continuous Non-Invasive Blood Pressure Measurement Using Pulse Transit Time in Children and Adolescents
    Zagrada, Elena
    Krickau, Tobias
    Rascher, Wolfgang
    KLINISCHE PADIATRIE, 2019, 231 (02): : 67 - 73
  • [3] CONTINUOUS AND NON-INVASIVE BLOOD PRESSURE MEASUREMENT BASED ON PULSE TRANSIT TIME COMPARED TO AMBULATORY BLOOD PRESSURE MEASUREMENT
    Haberl, R.
    Wesely, M.
    Thimel, K.
    Gehring, J.
    Kuechler, G.
    JOURNAL OF HYPERTENSION, 2016, 34 : E287 - E287
  • [4] A review of methods for non-invasive and continuous blood pressure monitoring: Pulse transit time method is promising?
    Peter, L.
    Noury, N.
    Cerny, M.
    IRBM, 2014, 35 (05) : 271 - 282
  • [5] A novel method for non-invasive blood pressure estimation based on continuous pulse transit time: An observational study
    Shin, Hangsik
    PSYCHOPHYSIOLOGY, 2023, 60 (02)
  • [6] Pulse transit time as an indicator of non-invasive blood pressure in the obstetric patient
    Moultrie, C. E. J.
    Heidemann, B. H.
    Drummond, G. B.
    Parmar, D.
    BRITISH JOURNAL OF ANAESTHESIA, 2007, 99 (05) : 755P - 756P
  • [7] Optimizing Non-Invasive Blood Pressure Estimation Using Pulse Transit Time
    Jeong, In Cheol
    Finkelstein, Joseph
    MEDINFO 2013: PROCEEDINGS OF THE 14TH WORLD CONGRESS ON MEDICAL AND HEALTH INFORMATICS, PTS 1 AND 2, 2013, 192 : 1198 - 1198
  • [8] An Investigation of Pulse Transit Time as a Non-Invasive Blood Pressure Measurement Method
    McCarthy, B. M.
    O'Flynn, B.
    Mathewson, A.
    SENSORS & THEIR APPLICATIONS XVI, 2011, 307
  • [9] NON-INVASIVE, CONTINUOUS, PULSE PRESSURE MONITORING METHOD
    Kuwahara, Makiko
    Yavari, Ehsan
    Boric-Lubecke, Olga
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 6574 - 6577
  • [10] PTT Based Continuous Time Non-invasive Blood Pressure System
    Atef, Mohamed
    Li Xiyan
    Wang, Guoxing
    Lian, Yong
    2016 IEEE 59TH INTERNATIONAL MIDWEST SYMPOSIUM ON CIRCUITS AND SYSTEMS (MWSCAS), 2016, : 329 - 332