DELIVERY OF A NEBULIZED AEROSOL TO A LUNG MODEL DURING MECHANICAL VENTILATION - EFFECT OF VENTILATOR SETTINGS AND NEBULIZER TYPE, POSITION, AND VOLUME OF FILL

被引:110
|
作者
ODOHERTY, MJ [1 ]
THOMAS, SHL [1 ]
PAGE, CJ [1 ]
TREACHER, DF [1 ]
NUNAN, TO [1 ]
机构
[1] ST THOMAS HOSP,INTENS CARE UNIT,LONDON SE1 7EH,ENGLAND
来源
AMERICAN REVIEW OF RESPIRATORY DISEASE | 1992年 / 146卷 / 02期
关键词
D O I
10.1164/ajrccm/146.2.383
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Several factors may affect the delivery of a nebulized aerosol to the lung through an endotracheal tube during mechanical ventilation. To study these factors in vitro, a model representing ventilation of an adult patient was constructed by linking a Servo 900C ventilator to a standard humidified circuit and an endotracheal (ET) tube positioned within a pipe representing the trachea. This was connected via a filter to a lung simulator. Nebulizers filled with Tc-99m human serum albumin were positioned in the circuit, and the delivery of nebulized aerosol through the ET tube into the filter was measured using a gamma camera. With the use of an inspiratory phase-activated System 22 Acorn jet nebulizer, typical adult ventilator settings, and a 3-ml nebulizer solution volume, 5.4% of the nebulizer dose reached beyond the end of the ET tube. This was increased by increasing the inspiratory time, reducing the respiratory rate or respiratory minute volume, and by repositioning the nebulizer on the inspiratory limb of the Y-piece and was reduced by slowing the driving gas flow to the nebulizer. Under the same conditions, delivery was 3.1 and 4.4% using the Samsonic and Fisoneb ultrasonic nebulizers, respectively. Increasing the fill volume and the addition of an aerosol storage chamber increased delivery with all three nebulizers. These experiments suggest some simple ways of improving aerosol delivery during mechanical ventilation, including increasing the volume of nebulizer fill, repositioning the nebulizer in the ventilator circuit, adding an aerosol storage chamber, and adjusting ventilator settings to maximize delivery.
引用
收藏
页码:383 / 388
页数:6
相关论文
共 50 条
  • [31] EFFECT OF DOSE VOLUME ON AEROSOL DELIVERY WITH MESH AND JET NEBULIZERS DURING NONINVASIVE VENTILATION: AN IN VITRO MODEL
    Fink, J. B.
    Ari, A.
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2013, 26 (02) : A33 - A34
  • [32] AEROSOL DELIVERY OF INHALED VANCOMYCIN DURING MECHANICAL VENTILATION USING I-VENT SYSTEM, A UNIQUE SMALL VOLUME, BREATH ENHANCED JET NEBULIZER
    Toddywala, Rohinton
    Shukla, Vijay
    Hande, Akanksha
    Smaldone, Gerald
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2019, 32 (03) : A16 - A17
  • [33] Aerosol Delivery of Inhaled Drugs During Mechanical Ventilation Using I-AIRE System, a Unique Small Volume, Breath Enhanced Jet Nebulizer
    Toddywala, R.
    Shukla, V.
    Hande, A.
    Smaldone, G. C.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201
  • [34] Effect of driving pressure and nebulizer model on aerosol output during intermittent delivery with a dosimeter
    Alkins, SA
    Hurwitz, KM
    Sierra, A
    JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2002, 15 (01): : 1 - 6
  • [35] In vitro study and semiempirical model for aerosol delivery control during mechanical ventilation
    Vecellio, L
    Guérin, C
    Grimbert, D
    De Monte, M
    Diot, P
    INTENSIVE CARE MEDICINE, 2005, 31 (06) : 871 - 876
  • [36] In vitro study and semiempirical model for aerosol delivery control during mechanical ventilation
    Laurent Vecellio
    Claude Guérin
    Daniel Grimbert
    Michele De Monte
    Patrice Diot
    Intensive Care Medicine, 2005, 31 : 871 - 876
  • [37] EFFECT OF DROPLET SIZE ON AEROSOL DELIVERY DURING SIMULATED NEONATAL MECHANICAL VENTILATION
    Bennett, Gavin
    Joyce, Mary
    Fernandez, Elena Fernandez
    MacLoughlin, Ronan
    ARCHIVES OF DISEASE IN CHILDHOOD, 2019, 104 : A36 - A36
  • [38] EFFECT OF TURNING OFF HUMIDIFICATION ON AEROSOL DELIVERY AND HUMIDITY DURING MECHANICAL VENTILATION
    Li, Jie
    Jacquier, Sophie
    Lin, Hui-Ling
    Sheridan, Caylie
    Karabelas, Paul
    Liu, Jui-Fang
    Ehrmann, Stephan
    Fink, James
    CRITICAL CARE MEDICINE, 2024, 52
  • [39] Effect of Nebulizer Location and Spontaneous Breathing on Aerosol Delivery During Airway Pressure Release Ventilation in Bench Testing
    Ge, Hui-Qing
    Wang, Ji-Mei
    Lin, Hui-Ling
    Fink, James B.
    Luo, Ronghua
    Xu, Peifeng
    Ying, Kejing
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2019, 32 (01) : 34 - 39
  • [40] The Effect of Targeting Scheme on Tidal Volume Delivery During Volume Control Mechanical Ventilation
    Volsko, Teresa A.
    Hoffman, Justin
    Conger, Alecia
    Chatburn, Robert L.
    RESPIRATORY CARE, 2012, 57 (08) : 1297 - 1304