Investigation on the automatic cell counting in feline blood

被引:0
|
作者
Mischke, R [1 ]
Deniz, A [1 ]
Weiss, J [1 ]
机构
[1] SERONO DIAGNOST GMBH,FREIBURG,GERMANY
来源
DEUTSCHE TIERARZTLICHE WOCHENSCHRIFT | 1995年 / 102卷 / 11期
关键词
D O I
暂无
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The usability of the haematology analyzers ''F-800'' and ''System 9000+'' was investigated in cats using 40 blood samples. Reference measurements were done with a haemocytometer or the microhaematocrit method. Reliable leukocyte counts were measured with the analyzer F-800 in cases only, when the histogram rendered possible a clear demarcation of the leukocyte curve (r = 0.97, n = 25). With the analyzer System 9000+ a changed basic adjustment of the inferior discriminator (49 fl) vouched for a close correlation with the values measured visually (r = 0.94). The volume distribution curves of the erythrocytes and thrombocytes which were partly overlapping allowed an approximate counting of the thrombocytes with both analyzers only of one half of the samples, The inaccuracy resulting from the inexact separation had only a relatively unimportant effect on the automatically measured erythrocyte count due to the numerical relation of erythro- and thrombocytes. Hereat a close correlation to the haemocytometer method (r = 0.95) was found as well as slightly higher values with the automatic counting. The haematocrit measured with the analyzer F-800 and to a smaller extent by the use of the System 9000+ lay distinctly higher in relation to the Values measured by microhaematocrit, when the analyzers were calibrated with human standards. Besides a species specific calibration of the haematocrit and a particular basic adjustment for leukocyte counting with the analyzer System 9000+, the use of the investigated haematology analyzers on feline blood requires an individual estimation of the histograms and removing of the discriminators as well as frequently a repetition of the leukocyte- and thrombocyte counting with a haemocytometer.
引用
收藏
页码:435 / 440
页数:6
相关论文
共 50 条
  • [41] ERRORS IN CONDUCTOMETRIC BLOOD-CELL COUNTING
    BRYSKIN, AI
    PARASHIN, VB
    [J]. MEASUREMENT TECHNIQUES, 1975, 18 (05) : 771 - 774
  • [42] Automatic Detection and Counting of Blood Cells in Smear Images Using RetinaNet
    Dralus, Grzegorz
    Mazur, Damian
    Czmil, Anna
    [J]. ENTROPY, 2021, 23 (11)
  • [43] The improved Neubauer ruling for blood cell counting
    Trenner, S
    [J]. JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 1924, 10 (1/3): : 0056 - 0058
  • [44] White blood cell counting techniques in birds
    Walberg, J
    [J]. SEMINARS IN AVIAN AND EXOTIC PET MEDICINE, 2001, 10 (02): : 72 - 76
  • [45] A COMPARISON OF RED BLOOD CELL COUNTING TECHNICS
    CONNOLLY, VJ
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1947, 17 (03) : 254 - 255
  • [46] AN ELECTRONIC BLOOD-CELL-COUNTING MACHINE
    CROSLANDTAYLOR, P
    STEWART, JW
    HAGGIS, G
    [J]. BLOOD, 1958, 13 (04) : 398 - 409
  • [47] Investigation of immune cell markers in feline oral squamous cell carcinoma
    Sparger, Ellen E.
    Murphy, Brian G.
    Kamal, Farina Mustaffa
    Arzi, Boaz
    Naydan, Diane
    Skouritakis, Chrisoula T.
    Cox, Darren P.
    Skorupski, Katherine
    [J]. VETERINARY IMMUNOLOGY AND IMMUNOPATHOLOGY, 2018, 202 : 52 - 62
  • [48] OSLO: AUTOMATIC CELL COUNTING AND SEGMENTATION FOR OLIGODENDROCYTE PROGENITOR CELLS
    Ma, Haoyi
    Beiter, Rebecca
    Gaultier, Alban
    Acton, Scott T.
    Lin, Zongli
    [J]. 2018 25TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2018, : 2431 - 2435
  • [49] Automatic cell identification and counting of leaf epidermis for plant phenotyping
    Mele, Giacomo
    Gargiulo, Laura
    [J]. METHODSX, 2020, 7
  • [50] Automatic cell counting in vivo in the larval nervous system of Drosophila
    Forero, M. G.
    Kato, K.
    Hidalgo, A.
    [J]. JOURNAL OF MICROSCOPY, 2012, 246 (02) : 202 - 212