PULSE-HEIGHT DEFECT IN AN IONIZATION-CHAMBER INVESTIGATED BY COLD FISSION MEASUREMENTS

被引:36
|
作者
SIMON, G [1 ]
TROCHON, J [1 ]
BRISARD, F [1 ]
SIGNARBIEUX, C [1 ]
机构
[1] CENS,CEA,F-91191 GIF SUR YVETTE,FRANCE
关键词
D O I
10.1016/0168-9002(90)90224-T
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The cold fragmentation of 235U thermal neutron fission has been studied in a twin ionization chamber working with pure methane gas. Due to the very high fragment mass resolution, a careful investigation of the pulse height defect in this detector has been achieved. No dependence of the pulse height defect on the fragment mass has been observed. Therefore, it is easy to calibrate the energy response of such a detector. An accurate calibration method is proposed. © 1990.
引用
收藏
页码:220 / 229
页数:10
相关论文
共 50 条
  • [41] EFFECTS OF IONIZATION-CHAMBER CONSTRUCTION ON MEASUREMENTS OF HETEROGENEITY CORRECTION FACTORS
    MAUCERI, TC
    KASE, KR
    MEDICAL PHYSICS, 1986, 13 (04) : 605 - 605
  • [42] PRECISION IONIZATION-CHAMBER MEASUREMENTS OF NATURAL BACKGROUND-RADIATION
    BRONSON, FL
    GEORGE, DL
    HARDING, WH
    PHYSICS IN MEDICINE AND BIOLOGY, 1974, 19 (02): : 239 - 239
  • [43] AN IONIZATION-CHAMBER FOR MEASUREMENTS OF HIGH-LEVEL TRITIUM GAS
    CARSTENS, DHW
    DAVID, WR
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1980, 31 (08): : 482 - 482
  • [44] DRIFT CHAMBER AND PULSE-HEIGHT READOUT SYSTEMS USING ANALOG MULTIPLEXING
    CISNEROS, EL
    KANG, HK
    HALL, JN
    LARSEN, RS
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (01) : 413 - 418
  • [45] A PULSE-COUNTING IONIZATION-CHAMBER FOR MEASURING THE RADON CONCENTRATION IN AIR
    BALTZER, P
    GORSTEN, KG
    BACKLIN, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 317 (1-2): : 357 - 364
  • [46] THE MASS DEPENDENCE OF THE SIGNAL PEAK HEIGHT OF A BRAGG-CURVE IONIZATION-CHAMBER
    SHENHAV, NJ
    STELZER, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 228 (2-3): : 359 - 364
  • [47] Pulse-height measurements with a cooled avalanche-photodiode detector
    Kishimoto, S
    JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 : 883 - 885
  • [48] Pulse-height defect in single-crystal CVD diamond detectors
    Beliuskina, O.
    Strekalovsky, A. O.
    Aleksandrov, A. A.
    Aleksandrova, I. A.
    Devaraja, H. M.
    Heinz, C.
    Heinz, S.
    Hofmann, S.
    Ilich, S.
    Imai, N.
    Kamanin, D. V.
    Kis, M.
    Knyazheva, G. N.
    Kozhuharov, C.
    Kuznetsova, E. A.
    Maurer, J.
    Mishinsky, G. V.
    Pomorski, M.
    Pyatkov, Yu. V.
    Strekalovsky, O. V.
    Traeger, M.
    Zhuchko, V. E.
    EUROPEAN PHYSICAL JOURNAL A, 2017, 53 (02):
  • [49] Pulse-height defect in single-crystal CVD diamond detectors
    O. Beliuskina
    A. O. Strekalovsky
    A. A. Aleksandrov
    I. A. Aleksandrova
    H. M. Devaraja
    C. Heinz
    S. Heinz
    S. Hofmann
    S. Ilich
    N. Imai
    D. V. Kamanin
    M. Kis
    G. N. Knyazheva
    C. Kozhuharov
    E. A. Kuznetsova
    J. Maurer
    G. V. Mishinsky
    M. Pomorski
    Yu. V. Pyatkov
    O. V. Strekalovsky
    M. Träger
    V. E. Zhuchko
    The European Physical Journal A, 2017, 53
  • [50] IONIZATION-CHAMBER BACKSCATTER MEASUREMENTS OF DIAGNOSTIC-X-RAY BEAMS
    STANTON, L
    BRATTELLI, SD
    MEDICAL PHYSICS, 1980, 7 (04) : 416 - 417