SATURATION BY PULSE OXIMETRY - COMPARISON OF THE RESULTS OBTAINED BY INSTRUMENTS OF DIFFERENT BRANDS

被引:46
|
作者
THILO, EH
ANDERSEN, D
WASSERSTEIN, ML
SCHMIDT, J
LUCKEY, D
机构
[1] CHILDRENS HOSP, DEPT RESP THERAPY, DENVER, CO 80218 USA
[2] CHILDRENS HOSP, DEPT PATHOL, DENVER, CO 80218 USA
[3] UNIV COLORADO, SCH MED, DEPT PEDIAT, DENVER, CO 80202 USA
来源
JOURNAL OF PEDIATRICS | 1993年 / 122卷 / 04期
关键词
D O I
10.1016/S0022-3476(05)83549-8
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
We noticed that arterial oxygen saturation by pulse oximetry (Spo2) was generally lower when determined by the Ohmeda Biox 3700 pulse oximeter than when determined by the Nellcor N-100 pulse oximeter, and we investigated whether this finding was consistent and the reason for the discrepancy. We placed both oximeters simultaneously on 30 infants with indwelling arterial catheters and measured arterial partial pressure of oxygen (Pao2), percentage of fetal hemoglobin, and complete cooximetry, including arterial oxygen saturation (Sao2) with a Radiometer OSM-3 cooximeter, with and without correction for fetal hemoglobin levels, in four samples of blood from each infant during a 12-hour period for a total of 120 samples. The Nellcor SPO2 was consistently higher than the Ohmeda SPO2 by a mean (+/- SD) of 1.61% +/- 2.69% (p <0.001). The Nellcor SPO2 correlated best with functional SaO2 ((oxyhemoglobin(HbO2)/(HbO2 + reduced hemoglobin)) x 100); Ohmedo SPO2 correlated best with fractional SaO2 ((HbO2/(HbO2 + reduced hemoglobin + carboxyhemoglobin + methemoglobin)) X 100), reflecting a fundamental difference in the calibration algorithms used in the two instruments. A desired Pao2 of 50 to 100 mm Hg, is maintained when the range of SPO2 is 90% to 96% for Ohmeda Spo2 and 92% to 98% for Nellcor Spo2 in the neonate, giving a positive predictive value in this study of 94% to 95%. We conclude that Spo2 determined by pulse oximeters of different brands is not interchangeable, and this may be of clinical importance in predicting Pao2 on the basis of Spo2.
引用
收藏
页码:620 / 626
页数:7
相关论文
共 50 条
  • [1] SATURATION BY PULSE OXIMETRY - ARE DIFFERENT BRANDS RESULTS INTERCHANGEABLE
    THILO, EH
    SCHMIDT, J
    ANDERSEN, D
    LUCKEY, D
    CLINICAL RESEARCH, 1992, 40 (01): : A133 - A133
  • [2] PULSE OXIMETRY, A CONTINUOUS AND NONINVASIVE MONITOR - A COMPARISON OF 2 INSTRUMENTS
    ALTEMEYER, KH
    MAYER, J
    BERGSEITER, S
    FOSEL, T
    ANAESTHESIST, 1986, 35 (01): : 43 - 45
  • [3] THE EFFECT OF HENNA PASTE ON OXYGEN-SATURATION READING OBTAINED BY PULSE OXIMETRY
    ALMAJED, SA
    HARAKATI, MS
    TROPICAL AND GEOGRAPHICAL MEDICINE, 1994, 46 (01): : 38 - 39
  • [4] Comparison of oxygen saturation by pulse oximetry and co-oximetry during exercise testing in patients with COPD
    McGovern, JP
    Sasse, SA
    Stansbury, DW
    Causing, LA
    Light, RW
    CHEST, 1996, 109 (05) : 1151 - 1155
  • [5] Reflectance pulse oximetry from core body in neonates and infants: Comparison to arterial blood oxygen saturation and to transmission pulse oximetry
    Kugelman A.
    Wasserman Y.
    Mor F.
    Goldinov L.
    Geller Y.
    Bader D.
    Journal of Perinatology, 2004, 24 (6) : 366 - 371
  • [6] Accuracy of the Apple Watch in measuring oxygen saturation: comparison with pulse oximetry and ABG
    Banu Arslan
    Kemal Sener
    Ramazan Guven
    Mucahit Kapci
    Semih Korkut
    Mehmet N. Sutasir
    Mustafa A. Tekindal
    Irish Journal of Medical Science (1971 -), 2024, 193 : 477 - 483
  • [7] Accuracy of the Apple Watch in measuring oxygen saturation: comparison with pulse oximetry and ABG
    Arslan, Banu
    Sener, Kemal
    Guven, Ramazan
    Kapci, Mucahit
    Korkut, Semih
    Sutasir, Mehmet N.
    Tekindal, Mustafa A.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2024, 193 (01) : 477 - 483
  • [8] A comparison of results of different instruments for measuring
    Bryant, GL
    Porter, FR
    JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1936, 19 : 153 - 158
  • [9] Pulse Oximetry Overestimates Oxygen Saturation in COPD
    Amalakanti, Sridhar
    Pentakota, Mohan Rao
    RESPIRATORY CARE, 2016, 61 (04) : 423 - 427
  • [10] Design and validation of pulse oximetry for low saturation
    Mannheimer, PD
    ANESTHESIA AND ANALGESIA, 2002, 94 (01): : S21 - S25