SILICON DETECTOR SYSTEM FOR HIGH-RATE EXAFS APPLICATIONS

被引:8
|
作者
PULLIA, A [1 ]
KRANER, HW [1 ]
SIDDONS, DP [1 ]
FURENLID, LR [1 ]
BERTUCCIO, G [1 ]
机构
[1] POLITECN MILAN,DIPARTIMENTO ELETTRON & INFORMAZ,I-20133 MILAN,ITALY
关键词
D O I
10.1109/23.467909
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A multichannel silicon pad detector for EXAFS (Extended X-ray Absorption Fine Structure) applications has been designed and built. The X-ray spectroscopic measurements demonstrate that an adequate energy resolution of 230 eV FWHM (corresponding to 27 rms electrons in silicon) can be achieved reliably at -35 degrees C. A resolution of 190 eV FWHM (corresponding to 22 rms electrons) has been obtained from individual pads at -35 degrees C. At room temperature (25 degrees C) an average energy resolution of 380 eV FWHM is achieved and a resolution of 350 eV FWHM (41 rms electrons) is the best performance. A simple cooling system constituted of Pettier cells is sufficient to reduce the reverse currents of the pads and their related shot noise contribution, in order to achieve resolutions better than 300 eV FWHM which is adequate for the EXAFS applications.
引用
收藏
页码:585 / 589
页数:5
相关论文
共 50 条
  • [1] Development of a high-rate, high-resolution detector for EXAFS experiments
    De Geronimo, G
    O'Connor, P
    Beuttenmuller, RH
    Li, Z
    Kuczewski, AJ
    Siddons, DP
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 19 - 23
  • [2] Development of a high-rate high-resolution detector for EXAFS experiments
    De Geronimo, G
    O'Connor, P
    Beuttenmuller, RH
    Li, Z
    Kuczewski, AJ
    Siddons, DP
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (04) : 885 - 891
  • [3] HIGH-RATE NAI DETECTOR SYSTEM
    JOHNSON, LO
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (01) : 465 - 475
  • [4] A transient recorder system for applications in high-rate detector systems WCC 1998
    Stiehler, R
    Adamek, M
    Besch, HJ
    Junk, M
    Menzel, G
    Neuser, E
    Meissner, W
    Pavel, N
    Sarvestani, A
    Sauer, N
    Volkov, S
    Walenta, AH
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 419 (2-3): : 711 - 714
  • [5] PERFORMANCE OF A SILICON MICROSTRIP DETECTOR IN A VERY HIGH-RATE ENVIRONMENT
    MCMANUS, A
    BODEN, A
    CHEN, TY
    COOPER, M
    COX, B
    FORTNEY, L
    GOLOVATYUK, V
    HANLET, P
    KOWALD, W
    MAZUR, P
    NELSON, K
    RAMACHANDRAN, S
    RECAGNI, M
    SEGAL, J
    SELOVE, W
    SMITH, R
    SUN, J
    TRISCHUK, J
    VANBERG, R
    YAO, N
    ZOU, B
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (05) : 1249 - 1253
  • [6] A HIGH-RATE, LOW-NOISE, X-RAY SILICON STRIP DETECTOR SYSTEM
    LUDEWIGT, B
    JAKLEVIC, J
    KIPNIS, I
    ROSSINGTON, C
    SPIELER, H
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 1037 - 1041
  • [7] A HIGH-RATE FASTBUS SILICON STRIP READOUT SYSTEM
    SWOBODA, C
    BARSOTTI, E
    BOWDEN, M
    CHRISTIAN, D
    DEMAAT, R
    FACHIN, M
    GONZALEZ, H
    HANCE, R
    HALDEMAN, M
    HOFF, J
    LARWILL, M
    ROTOLO, C
    TRENDLER, R
    TREPTOW, K
    URISH, J
    WALSH, D
    YAREMA, R
    ZIMMERMAN, T
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) : 342 - 346
  • [8] Comparison of readout systems for high-rate silicon photomultiplier applications
    Wong, M. L.
    Kolodziej, M.
    Briggl, K.
    Hetzel, R.
    Korcyl, G.
    Lalik, R.
    Malige, A.
    Magiera, A.
    Ostrzolek, G.
    Rusiecka, K.
    Stahl, A.
    Urbanevych, V.
    Wiebusch, M.
    Wronska, A.
    JOURNAL OF INSTRUMENTATION, 2024, 19 (01)
  • [9] CHARGED-PARTICLE DETECTOR SYSTEM FOR HIGH-RATE EXPERIMENTS
    PLATNER, ED
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1978, 25 (01) : 35 - 37
  • [10] Silicon array detector system for high-rate, low-noise X-ray spectroscopy
    Ludewigt, BA
    Krieger, B
    Lindstrom, D
    Maier, MR
    Rutgersson, M
    Tull, CR
    Yaver, H
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (04) : 886 - 889