Deformations and cyclic fatigue resistance of nickel-titanium instruments inside a sequence

被引:5
|
作者
Gambarini, Gianluca [1 ]
Plotino, Gianluca [1 ]
Piasecki, Lucila [2 ]
Al-Sudani, Dina [3 ]
Testarelli, Luca [1 ]
Sannino, Gianpaolo [4 ]
机构
[1] Sapienza Univ Rome, Endodont Unit, Dept Oral & Maxillofacial Sci, Rome, Italy
[2] PUC Univ, Curitiba, Parana, Brazil
[3] King Saud Univ, Coll Dent, Dept Restorat Dent Sci, Riyadh, Saudi Arabia
[4] Univ Rome II, Tor Vergata, Italy
关键词
nickel-titanium; brushing action; deformations; cyclic fatigue;
D O I
10.11138/ads/2015.6.1.006
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim. To compare the effect of brushing motion on torsional and cyclic fatigue resistance of TF Adaptive instruments after clinical use. Methods. 20 packs of TFA small sequence (SybronEndo, Orange, CA, USA) were used for this study and divided into two groups. Each instrument prepared one resin tooth, consisting in 4 canals with a complex anatomy. In group A, no brushing motion was performed. In group B, after the green instrument reached the working length, brushing motion with circumferential filing was performed for 15 seconds in each canal (overall 1 minute). All the instruments were then subjected to cyclic fatigue test and mean values and standard deviation for time to fracture were evaluated. Data were subjected to one-way analysis of variance and Bonferroni t-test procedure with a significance set at P < 0.05. Results. No instruments were broken during preparation of root canals. Two TF Adaptive green and 5 yellow showed unwinding after intracanal clinical use. No statistically significant differences were found between green instruments of both groups (P > 0.05), while a statistically significant difference was found between the yellow instruments (P < 0.05), with group B showing an higher resistance to cyclic fatigue. Conclusions. A prolonged passive brushing motion did not adversely affected mechanical resistance of the instrument used for this purpose. Resistance to both deformations and cyclic fatigue of the second instrument within the TFA small sequence was enhanced by the coronal flaring provided by the brushing action of the first instrument used.
引用
收藏
页码:6 / 9
页数:4
相关论文
共 50 条
  • [41] Evaluation of cyclic and torsional fatigue resistance of several heat-treated reciprocating nickel-titanium instruments
    Weissheimer, Theodoro
    Duarte, Marco Antonio Hungaro
    So, Marcus Vinicius Reis
    da Rosa, Ricardo Abreu
    Espinosa, Michel Klymus
    Vivan, Rodrigo Ricci
    Alcalde, Murilo Priori
    AUSTRALIAN ENDODONTIC JOURNAL, 2023, 49 (03) : 524 - 529
  • [42] Cyclic fatigue resistance of XP-endo Shaper compared with different nickel-titanium alloy instruments
    Elnaghy, Amr
    Elsaka, Shaymaa
    CLINICAL ORAL INVESTIGATIONS, 2018, 22 (03) : 1433 - 1437
  • [43] Influence of Continuous or Reciprocating Motion on Cyclic Fatigue Resistance of 4 Different Nickel-Titanium Rotary Instruments
    Pedulla, Eugenio
    Grande, Nicola M.
    Plotino, Gianluca
    Gambarini, Gianluca
    Rapisarda, Ernesto
    JOURNAL OF ENDODONTICS, 2013, 39 (02) : 258 - 261
  • [44] Cyclic Fatigue Resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium Instruments
    Gundogar, Mustafa
    Ozyurek, Taha
    JOURNAL OF ENDODONTICS, 2017, 43 (07) : 1192 - 1196
  • [45] Effect of Repeated Autoclaving on Cyclic Fatigue of Three Nickel-Titanium Rotary Instruments
    Elbatal, Laila M.
    El Batouty, Kariem Mostafa
    Obeid, Maram Mohamed
    ENDO-ENDODONTIC PRACTICE TODAY, 2016, 10 (01): : 15 - 22
  • [46] In vitro cyclic fatigue analysis of different endodontic Nickel-Titanium rotary instruments
    Cutro, Giuseppe
    Amateis, Valentina
    Provenzano, Salvatore C.
    Cusimano, Paolo
    Cumbo, Enzo M.
    Gallina, Giuseppe
    GIORNALE ITALIANO DI ENDODONZIA, 2012, 26 (03): : 145 - 149
  • [47] An In Vitro Cyclic Fatigue Analysis of Different Endodontic Nickel-Titanium Rotary Instruments
    Bhagabati, Nilav
    Yadav, Seema
    Talwar, Sangeeta
    JOURNAL OF ENDODONTICS, 2012, 38 (04) : 515 - 518
  • [48] Survival Analysis of Factors Influencing Cyclic Fatigue of Nickel-Titanium Endodontic Instruments
    Fiserova, Eva
    Chvostekova, Martina
    Bjlaskova, Silvie
    Bumbalek, Michal
    Joska, Zdenek
    ADVANCES IN MATERIALS SCIENCE AND ENGINEERING, 2015, 2015
  • [49] Torsion property and cyclic fatigue fracture behavior of nickel-titanium endodontic instruments
    Zheng, YF
    Li, C
    Huang, C
    Zhao, LC
    ASBM6: ADVANCED BIOMATERIALS VI, 2005, 288-289 : 603 - 606
  • [50] Influence of Different Angles of Reciprocation on the Cyclic Fatigue of Nickel-Titanium Endodontic Instruments
    Gambarini, Gianluca
    Rubini, Alessio Giansiracusa
    Al Sudani, Dina
    Gergi, Richard
    Culla, Antonio
    De Angelis, Francesca
    Di Carlo, Stefano
    Pompa, Giorgio
    Osta, Nada
    Testarelli, Luca
    JOURNAL OF ENDODONTICS, 2012, 38 (10) : 1408 - 1411