Deformations and cyclic fatigue resistance of nickel-titanium instruments inside a sequence

被引:5
|
作者
Gambarini, Gianluca [1 ]
Plotino, Gianluca [1 ]
Piasecki, Lucila [2 ]
Al-Sudani, Dina [3 ]
Testarelli, Luca [1 ]
Sannino, Gianpaolo [4 ]
机构
[1] Sapienza Univ Rome, Endodont Unit, Dept Oral & Maxillofacial Sci, Rome, Italy
[2] PUC Univ, Curitiba, Parana, Brazil
[3] King Saud Univ, Coll Dent, Dept Restorat Dent Sci, Riyadh, Saudi Arabia
[4] Univ Rome II, Tor Vergata, Italy
关键词
nickel-titanium; brushing action; deformations; cyclic fatigue;
D O I
10.11138/ads/2015.6.1.006
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim. To compare the effect of brushing motion on torsional and cyclic fatigue resistance of TF Adaptive instruments after clinical use. Methods. 20 packs of TFA small sequence (SybronEndo, Orange, CA, USA) were used for this study and divided into two groups. Each instrument prepared one resin tooth, consisting in 4 canals with a complex anatomy. In group A, no brushing motion was performed. In group B, after the green instrument reached the working length, brushing motion with circumferential filing was performed for 15 seconds in each canal (overall 1 minute). All the instruments were then subjected to cyclic fatigue test and mean values and standard deviation for time to fracture were evaluated. Data were subjected to one-way analysis of variance and Bonferroni t-test procedure with a significance set at P < 0.05. Results. No instruments were broken during preparation of root canals. Two TF Adaptive green and 5 yellow showed unwinding after intracanal clinical use. No statistically significant differences were found between green instruments of both groups (P > 0.05), while a statistically significant difference was found between the yellow instruments (P < 0.05), with group B showing an higher resistance to cyclic fatigue. Conclusions. A prolonged passive brushing motion did not adversely affected mechanical resistance of the instrument used for this purpose. Resistance to both deformations and cyclic fatigue of the second instrument within the TFA small sequence was enhanced by the coronal flaring provided by the brushing action of the first instrument used.
引用
收藏
页码:6 / 9
页数:4
相关论文
共 50 条
  • [1] Effect of Torsional Loading of Nickel-Titanium Instruments on Cyclic Fatigue Resistance
    Cheung, Gary Shun-Pan
    Oh, Seung-Hei
    Ha, Jung-Hong
    Kim, Sung Kyo
    Park, Se-Hee
    Kim, Hyeon-Cheol
    JOURNAL OF ENDODONTICS, 2013, 39 (12) : 1593 - 1597
  • [2] Effect of movement kinematics on the cyclic fatigue resistance of nickel-titanium instruments
    Karatas, E.
    Arslan, H.
    Buker, M.
    Seckin, F.
    Capar, I. D.
    INTERNATIONAL ENDODONTIC JOURNAL, 2016, 49 (04) : 361 - 364
  • [3] A comparison of cyclic fatigue resistance in reciprocating and rotary nickel-titanium instruments
    Wan, Jeffrey
    Rasimick, Brian J.
    Musikant, Barry L.
    Deutsch, Allan S.
    AUSTRALIAN ENDODONTIC JOURNAL, 2011, 37 (03) : 122 - 127
  • [4] Comparison of Cyclic Fatigue Resistance of Nickel-Titanium Coronal Flaring Instruments
    Capar, Ismail Davut
    Ertas, Huseyin
    Arslan, Hakan
    JOURNAL OF ENDODONTICS, 2014, 40 (08) : 1182 - 1185
  • [5] Comparison of cyclic fatigue resistance of novel nickel-titanium rotary instruments
    Capar, Ismail Davut
    Ertas, Huseyin
    Arslan, Hakan
    AUSTRALIAN ENDODONTIC JOURNAL, 2015, 41 (01) : 24 - 28
  • [6] Cyclic fatigue testing of nickel-titanium endodontic instruments
    Pruett, JP
    Clement, DJ
    Carnes, DL
    JOURNAL OF ENDODONTICS, 1997, 23 (02) : 77 - 85
  • [7] Effect of interrupted motion on the cyclic fatigue resistance of reciprocating nickel-titanium instruments
    Keskin, C.
    Inan, U.
    Demiral, M.
    INTERNATIONAL ENDODONTIC JOURNAL, 2018, 51 (05) : 549 - 555
  • [8] Cyclic fatigue resistance of two nickel-titanium rotary instruments in interrupted rotation
    Pedulla, E.
    Lizio, A.
    Scibilia, M.
    Grande, N. M.
    Plotino, G.
    Boninelli, S.
    Rapisarda, E.
    Lo Giudice, G.
    INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (02) : 194 - 201
  • [9] Cyclic fatigue resistance of Nickel-Titanium reciprocating instruments tested with an innovative kinematics
    Iacono, Francesco
    Pirani, Chiara
    Generali, Luigi
    Gatto, Maria Rosaria
    Gandolfi, Maria Giovanna
    Prati, Carlo
    GIORNALE ITALIANO DI ENDODONZIA, 2018, 32 (01): : 42 - 46
  • [10] Cyclic fatigue resistance of novel Genius and Edgefile nickel-titanium reciprocating instruments
    Emerenciano Bueno, Carlos Roberto
    Sandoval Cury, Marina Tolomei
    Veiga Vasques, Ana Maria
    Sivieri-Araujo, Gustavo
    Jacinto, Rogerio Castilho
    Gomes-Filho, Joao Eduardo
    Angelo Cintra, Luciano Tavares
    Dezan-Junior, Eloi
    BRAZILIAN ORAL RESEARCH, 2019, 33