ACCURACY OF DISPOSABLE BLOOD-PRESSURE TRANSDUCERS USED IN THE CRITICAL CARE SETTING

被引:7
|
作者
BAILEY, RH
BAUER, JH
YANOS, J
机构
[1] UNIV MISSOURI, DIV PULM CRIT CARE & ENVIRONM MED, COLUMBIA, MO USA
[2] UNIV MISSOURI, DIV NEPHROL, HYPERTENS RES SECT, COLUMBIA, MO USA
[3] UNIV MISSOURI, DEPT INTERNAL MED, COLUMBIA, MO USA
[4] HARRY S TRUMAN MEM VET HOSP, COLUMBIA, MO 65201 USA
[5] CASE WESTERN RESERVE UNIV, DEPT MED, CLEVELAND, OH 44106 USA
关键词
TRANSDUCERS; PRESSURE; BLOOD PRESSURE DETERMINATION; CRITICAL CARE; BLOOD PRESSURE; CENTRAL VENOUS PRESSURE; MANOMETRY; MONITORING; PHYSIOLOGICAL;
D O I
10.1097/00003246-199501000-00029
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To assess the initial accuracy, drift in calibration over several hours, and decline in accuracy over daily use of blood pressure transducers used in the critical care setting. Design: Prospective, criterion standard. Setting: Three critical care units in a university hospital. Subjects: Twenty-seven consecutively available high-pressure transducers were used to measure intra-arterial blood pressures, and 11 consecutively available low-pressure transducers were used to measure central venous and pulmonary arterial blood pressures. Interventions: High-pressure transducers were compared with a mercury column manometer, and low-pressure transducers were compared with a water column manometer at three pressure levels. Data were collected initially after installation of the pressure transducer, at 3 hrs, and at 6 hrs. Data collections were repeated on subsequent days. Measurements and Main Results: a) The majority of pressure transducers are accurate; b) there is no drift in the accuracy of pressure transducers over several hours; c) the accuracy of pressure transducers does not decline over subsequent days. Some transducers overestimated standard pressures by 10 mm Hg and/or underestimated standard pressure by 17 mm Hg. Conclusions: The majority of pressure transducers used in the critical care setting are accurate, although there are some transducers that may compromise patient care. Blood pressure transducers should be checked against a standard manometer upon installation, and daily during use.
引用
收藏
页码:187 / 192
页数:6
相关论文
共 50 条
  • [1] THE MEASURING ACCURACY OF DEVICES FOR INVASIVE BLOOD-PRESSURE MEASUREMENT WITH DISPOSABLE TRANSDUCERS - RESULTS OF A USER POLL
    NINK, R
    SCHUSTER, HP
    BIOMEDIZINISCHE TECHNIK, 1990, 35 (10): : 235 - 237
  • [2] COMPARISON OF THE PERFORMANCE OF DISPOSABLE PRESSURE TRANSDUCERS FOR INVASIVE BLOOD-PRESSURE MEASUREMENT
    MATZEK, F
    BOENICK, U
    FRUCHT, U
    SCHROEDER, PM
    BIOMEDIZINISCHE TECHNIK, 1989, 34 (03): : 34 - 41
  • [3] A DISPOSABLE BLOOD-PRESSURE CUFF - AN ASSESSMENT OF ITS ACCURACY
    DODD, FM
    ARMSTRONG, P
    RAY, DC
    SCOTT, DHT
    ANAESTHESIA, 1990, 45 (08) : 666 - 668
  • [6] Accuracy and reliability of disposable pressure transducers coupled with modern pressure monitors
    Gardner, RM
    CRITICAL CARE MEDICINE, 1996, 24 (05) : 879 - 882
  • [7] The accuracy of the flowrate in flush-devices of disposable pressure transducers
    Michel, C
    Roth, F
    Feigenwinter, P
    Zbinden, AM
    INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1995, 12 (04): : 219 - 224
  • [8] DIRECT AND INDIRECT BLOOD-PRESSURE IN CRITICAL CARE PATIENTS
    BYRACOOK, CJ
    DRACUP, KA
    LAZIK, AJ
    NURSING RESEARCH, 1990, 39 (05) : 285 - 288
  • [9] MEASUREMENT OF THE VOLUME DISPLACEMENT OF BLOOD-PRESSURE TRANSDUCERS
    GEDDES, LA
    ATHENS, W
    ARONSON, S
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1984, 22 (06) : 613 - 614
  • [10] DIGITAL BLOOD-PRESSURE MONITORING IN CRITICAL CARE AND DIALYSIS PATIENTS
    MCDOUGAL, B
    SCHWARM, D
    JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 1977, 90 (01): : 57 - 60