Preclinical evaluation of acoustic radiation force impulse measurements in regions of heterogeneous elasticity

被引:10
|
作者
Hollerieth, Katharina [1 ]
Gassmann, Bernhard [2 ]
Wagenpfeil, Stefan [3 ]
Moog, Philipp [1 ]
Minh-Truc Vo-Cong [1 ]
Heemann, Uwe [1 ]
Stock, Konrad Friedrich [1 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Dept Nephrol, Str 22, D-81675 Munich, Germany
[2] Meso Int GmbH, Berlin, Germany
[3] Univ Saarland, Inst Med Biometry Epidemiol & Med Informat, Campus Homburg Saar, Homburg, Germany
关键词
Elasticity imaging techniques; Ultrasonography; Phantoms; imaging; Transducers; Heterogeneity;
D O I
10.14366/usg.16024
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to compare the reliability of ultrasound-based shear wave elastography in regions of homogeneous versus heterogeneous elasticity by using two different probes. Methods: Using acoustic radiation force impulse (ARFI) elastography, we measured the shear wave velocity (SWV) in different lesions of an elastography phantom with the convex 4C1 probe and the linear 9L4 probe. The region of interest (ROI) was positioned in such a way that it was partly filled by one of the lesions (0%, 25%, 50%, 75%, and 100%) and partly by the background of the phantom (100%, 75%, 50%, 25%, and 0%, respectively). Results: The success rate was 98.5%. The measured value and the reference value of SWV correlated significantly (r=0.89, P<0.001). Further, a comparison of the two probes revealed that there was no statistical difference in either the mean or the variance values. However, the deviation of SWV from the reference was higher in the case of the 9L4 probe than in the case of the 4C1 probe, both overall and in measurements in which the ROI contained structures of different elasticity (P=0.021 and P=0.002). Taking into account all data, for both probes, we found that there was a greater spread and deviation of the SWV from the reference value when the ROI was positioned in structures having different elastic properties (standard deviation, 0.02 +/- 0.01 m/sec vs. 0.04 +/- 0.04 m/sec; P=0.010; deviation from the reference value, 0.21 +/- 0.12 m/sec vs. 0.38 +/- 0.27 m/sec; P=0.050). Conclusion: Quantitative ARFI elastography was achievable in structures of different elasticity; however, the validity and the reliability of the SWV measurements decreased in comparison to those of the measurements performed in structures of homogeneous elasticity. Therefore, a convex probe is preferred for examining heterogeneous structures.
引用
收藏
页码:345 / 352
页数:8
相关论文
共 50 条
  • [41] Elasticity of healthy Achilles tendon decreases with the increase of age as determined by acoustic radiation force impulse imaging
    Ruan, Zhengmin
    Zhao, Bin
    Qi, Hengtao
    Zhang, Yuan
    Zhang, Feixue
    Wu, Mei
    Shao, Guangrui
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (01): : 1043 - 1050
  • [42] ACOUSTIC RADIATION FORCE IMPULSE IMAGING (ARFI) AS A TOOL TO QUANTIFY TISSUE ELASTICITY IN RENAL ALLOGRAFT REJECTION
    Stavropoulou, M.
    Braesen, J. H.
    Tyzcynski, B.
    Kribben, A.
    Eisenberger, U.
    TRANSPLANT INTERNATIONAL, 2014, 27 : 28 - 29
  • [43] Elastography acoustic radiation force impulse (ARFI): an investigation of ocular elasticity in dogs with chronic secondary glaucoma
    Madruga, Gabriela Morais
    Rossi Feliciano, Marcus Antonio
    Maronezi, Marjury Cristina
    Filezio, Marcella Rosa
    Morato Abreu, Thais Guimaraes
    Ribeiro, Alexandre Pinto
    Fachini, Francimery Aparecida
    Kobashigawa, Kamachi
    Kniphoff da Cruz, Igor Cezar
    Ramirez Uscategui, Ricardo Andres
    Laus, Jose Luiz
    TURKISH JOURNAL OF VETERINARY & ANIMAL SCIENCES, 2021, 45 (03): : 404 - 410
  • [44] In Vitro Monitoring of the Dynamic Elasticity Changes During Radiofrequency Ablation with Acoustic Radiation Force Impulse Imaging
    Eyerly, Stephanie A.
    Dumont, Douglas M.
    Trahey, Gregg E.
    Wolf, Patrick D.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (04) : 472 - 473
  • [45] ACOUSTIC RADIATION FORCE IMPULSE ELASTOGRAPHY BASED ON ELASTICITY IMAGING FOR EARLY DETECTION OF IDIOPATHIC PULMONARY FIBROSIS
    Minthara, B.
    Susanti, I.
    Astuti, L.
    RESPIROLOGY, 2016, 21 : 191 - 191
  • [46] Comparison of the Delineation of Ablation Regions with Electrode Displacement Elastography and Acoustic Radiation Force Impulse Imaging
    Yang, W.
    Varghese, T.
    Ziemlewicz, T.
    Alexander, M.
    Lubner, M.
    Hinshaw, J.
    Wells, S.
    Lee, F., Jr.
    MEDICAL PHYSICS, 2017, 44 (06) : 3238 - 3238
  • [47] Evaluation of Renal Parenchyma Elasticity With Acoustic Radiation Force Impulse Quantification in Nutcracker Syndrome and Comparisons With Grayscale Doppler Sonography and Laboratory Findings
    Bekci, Tumay
    Bilgici, Meltem Ceyhan
    Genc, Gurkan
    Tekcan, Demet
    Polat, Ahmet Veysel
    Tomak, Leman
    JOURNAL OF ULTRASOUND IN MEDICINE, 2017, 36 (01) : 61 - 67
  • [48] EVALUATION OF THE ELASTICITY OF THE PANCREAS USING ACOUSTIC RADIATION FORCE IMPULSE ELASTOGRAPHY IN PATIENTS WITH ACUTE PANCREATITIS: A SYSTEMATIC REVIEW AND META-ANALYSIS
    Rifu, Kazuma
    Watanabe, Jun
    Sasanuma, Hideki
    Taniguchi, Nobuyuki
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2022, 48 (03): : 406 - 413
  • [49] An abdominal and thyroid status with Acoustic Radiation Force Impulse Elastometry - A feasibility study Acoustic Radiation Force Impulse Elastometry of human organs
    Goertz, R. S.
    Amann, K.
    Heide, R.
    Bernatik, T.
    Neurath, M. F.
    Strobel, D.
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 80 (03) : E226 - E230
  • [50] Acoustic Radiation Force Impulse (ARFI) Imaging: A Review
    Nightingale, Kathy
    CURRENT MEDICAL IMAGING, 2011, 7 (04) : 328 - 339